Diagnostic accuracy of an oscillometric blood pressure monitor for atrial fibrillation screening
View through CrossRef
Objective
Atrial fibrillation is the most common arrhythmia of clinical significance and hypertension is one of its major risk factors. This study aimed to evaluate the diagnostic accuracy of an automated oscillometric blood pressure (BP) monitor with the function of atrial fibrillation detection for atrial fibrillation screening.
Materials and methods
Patients attending outpatient cardiology clinics were recruited for atrial fibrillation screening by the BP monitor with triplicate BP measurements for atrial fibrillation detection. Furthermore, a single-lead ECG was recorded simultaneously for comparison as the reference standard. The diagnostic test’s evaluation index were analyzed, including sensitivity, specificity, and receiver operator characteristic (ROC) analysis.
Results
A total of 295 participants were analyzed including 166 males and 129 females, with an average age of 72.5 ± 5.9 years. The sensitivity and specificity for atrial fibrillation detection by the device were 1.000 and 0.904, respectively, with the area under the ROC curve of 0.952 (95% confidence interval: 0.929–0.975, P < 0.001). Furthermore, the device had a Kappa-value of 0.781 (P < 0.001) with the single-lead ECG in detecting atrial fibrillation.
Conclusion
The automated oscillometric BP monitor (G.LAB MD41A0) with atrial fibrillation detection function has high sensitivity and specificity with good accuracy for atrial fibrillation screening, which could be used as a reliable screening tool for the early detection of atrial fibrillation with potential benefits.
Ovid Technologies (Wolters Kluwer Health)
Title: Diagnostic accuracy of an oscillometric blood pressure monitor for atrial fibrillation screening
Description:
Objective
Atrial fibrillation is the most common arrhythmia of clinical significance and hypertension is one of its major risk factors.
This study aimed to evaluate the diagnostic accuracy of an automated oscillometric blood pressure (BP) monitor with the function of atrial fibrillation detection for atrial fibrillation screening.
Materials and methods
Patients attending outpatient cardiology clinics were recruited for atrial fibrillation screening by the BP monitor with triplicate BP measurements for atrial fibrillation detection.
Furthermore, a single-lead ECG was recorded simultaneously for comparison as the reference standard.
The diagnostic test’s evaluation index were analyzed, including sensitivity, specificity, and receiver operator characteristic (ROC) analysis.
Results
A total of 295 participants were analyzed including 166 males and 129 females, with an average age of 72.
5 ± 5.
9 years.
The sensitivity and specificity for atrial fibrillation detection by the device were 1.
000 and 0.
904, respectively, with the area under the ROC curve of 0.
952 (95% confidence interval: 0.
929–0.
975, P < 0.
001).
Furthermore, the device had a Kappa-value of 0.
781 (P < 0.
001) with the single-lead ECG in detecting atrial fibrillation.
Conclusion
The automated oscillometric BP monitor (G.
LAB MD41A0) with atrial fibrillation detection function has high sensitivity and specificity with good accuracy for atrial fibrillation screening, which could be used as a reliable screening tool for the early detection of atrial fibrillation with potential benefits.
Related Results
[RETRACTED] Guardian Blood Balance –Feel the difference Guardian Blood Balance makes! v1
[RETRACTED] Guardian Blood Balance –Feel the difference Guardian Blood Balance makes! v1
[RETRACTED]Guardian Blood Balance Reviews (Works Or Hoax) Does Guardian Botanicals Blood Balance AU Really Works? Read Updated Report! Diabetes and Hypertension is such a health p...
Adiponectin and Lone atrial fibrillation
Adiponectin and Lone atrial fibrillation
Objective: Lone atrial fibrillation is an idiopathic arrhythmia seen in younger individuals without any secondary disease. Adiponectin is an endogenous adipocytokine that increases...
RELATIONSHIP BETWEEN ATRIAL FIBRILLATION CARDIOVERSION AND F
RELATIONSHIP BETWEEN ATRIAL FIBRILLATION CARDIOVERSION AND F
Objectives
To investigate the relationship between atrial fibrillation cardioversion and f wave in electrocardiogram, providing an ordinary and noninvasive method...
ASSA13-03-9 Decreased Expression of Small-Conductance Ca2+-Activated K+ Channels SK1, SK2, and SK3 in Patients with Persistent Atrial Fibrillation
ASSA13-03-9 Decreased Expression of Small-Conductance Ca2+-Activated K+ Channels SK1, SK2, and SK3 in Patients with Persistent Atrial Fibrillation
Background
Small-conductance Ca2+-activated K+ channels (SK channels) have been reported involved in atrial fibrillation (AF) as a new ion channel candidates, as ...
Catheter ablation as a potential treatment alternative for atrial fibrillation among patients with heart failure with preserved ejection fraction: a retrospective cohort study
Catheter ablation as a potential treatment alternative for atrial fibrillation among patients with heart failure with preserved ejection fraction: a retrospective cohort study
Abstract
Background
In patients with heart failure with reduced ejection fraction, catheter ablation for atrial fibrillation is a useful alternative to drug-induced rhythm...
Left atrial strain parameters are able to predict presence of atrial fibrillation
Left atrial strain parameters are able to predict presence of atrial fibrillation
Abstract
Background
Three-dimensional speckle-tracking echocardiography allows evaluation of left atrial function, left atrial f...
REMODELLING OF CONNEXIN 43 IN ATRIAL MYOCARDIUM OF PATIENTS WITH ATRIAL FIBRILLATION
REMODELLING OF CONNEXIN 43 IN ATRIAL MYOCARDIUM OF PATIENTS WITH ATRIAL FIBRILLATION
Objectives
Remodelling of connexins was found accompanying with atrial fibrillation. The aim of the study is to investigate whether it is the remodelling of conne...
Left atrial function assessment by speckle tracking echocardiography in low-risk atrial fibrillation patients
Left atrial function assessment by speckle tracking echocardiography in low-risk atrial fibrillation patients
Objective: To compare the left atrial function utilising speckle tracking echocardiography in patients with low-risk and high-risk non-valvular atrial fibrillation.Methods: The des...