Ultrasound Biomicroscopy for Determining Visian Implantable Contact Lens Length in Phakic IOL Implantation
View through CrossRef
ABSTRACT
PURPOSE: To assess the efficacy of the ultrasound biomicroscopic (UBM) method in estimating the sulcus-to-sulcus horizontal diameter for Visian Implantable Contact Lens (ICL, model V4) length determination to obtain optimal ICL vault.
METHODS: The results of postoperative ICL vaults in 30 eyes of 18 patients were retrospectively analyzed. In 17 eyes, ICL length was determined using the conventional method, and in 13 eyes, ICL length was determined using the UBM method. The UBM method was carried out by measuring the sulcus to limbus distance on each side by 50 MHz UBM and adding the white-to-white diameter by caliper or Orbscan. The ICL vaults were measured using the UBM method at 1 and 6 months postoperatively and the results were compared between the two groups. Ideal ICL vault was defined as vault between 250 and 750 µm. The relation between the ICL vault, footplate location, and ICL power was also investigated.
RESULTS: In the UBM method group, ICL vault was within the ideal range in all 13 (100%) eyes at 1 and 6 months postoperatively, whereas in the conventional method group, 10 (58.8%) eyes showed ideal vault at 1 month postoperatively (P=.01) and 9 (52.9%) eyes showed ideal vault at 6 months postoperatively (P<.01). The ideal ICL footplate location was achieved in the ciliary sulcus in 11 (84.6%) eyes of the UBM method group and 10 (64.7%) eyes of the conventional method group. However, the differences between the two groups were not statistically significant. The ICL vault was not significantly affected by the ICL power.
CONCLUSIONS: Implantable Contact Lens length determined by the UBM method achieved significantly more ideal ICL vault than that of the conventional white-to-white method. The UBM method is superior to the conventional method in terms of predicting the sulcus-to-sulcus horizontal diameter for ICL length determination. [J Refract Surg. 2007;23:362-367.]
Title: Ultrasound Biomicroscopy for Determining Visian Implantable Contact Lens Length in Phakic IOL Implantation
Description:
ABSTRACT
PURPOSE: To assess the efficacy of the ultrasound biomicroscopic (UBM) method in estimating the sulcus-to-sulcus horizontal diameter for Visian Implantable Contact Lens (ICL, model V4) length determination to obtain optimal ICL vault.
METHODS: The results of postoperative ICL vaults in 30 eyes of 18 patients were retrospectively analyzed.
In 17 eyes, ICL length was determined using the conventional method, and in 13 eyes, ICL length was determined using the UBM method.
The UBM method was carried out by measuring the sulcus to limbus distance on each side by 50 MHz UBM and adding the white-to-white diameter by caliper or Orbscan.
The ICL vaults were measured using the UBM method at 1 and 6 months postoperatively and the results were compared between the two groups.
Ideal ICL vault was defined as vault between 250 and 750 µm.
The relation between the ICL vault, footplate location, and ICL power was also investigated.
RESULTS: In the UBM method group, ICL vault was within the ideal range in all 13 (100%) eyes at 1 and 6 months postoperatively, whereas in the conventional method group, 10 (58.
8%) eyes showed ideal vault at 1 month postoperatively (P=.
01) and 9 (52.
9%) eyes showed ideal vault at 6 months postoperatively (P<.
01).
The ideal ICL footplate location was achieved in the ciliary sulcus in 11 (84.
6%) eyes of the UBM method group and 10 (64.
7%) eyes of the conventional method group.
However, the differences between the two groups were not statistically significant.
The ICL vault was not significantly affected by the ICL power.
CONCLUSIONS: Implantable Contact Lens length determined by the UBM method achieved significantly more ideal ICL vault than that of the conventional white-to-white method.
The UBM method is superior to the conventional method in terms of predicting the sulcus-to-sulcus horizontal diameter for ICL length determination.
[J Refract Surg.
2007;23:362-367.
].
Related Results
Biometry and Intraocular Lens Power Calculation by Combined Scheimpflug-Placido Disc versus Optical Interferometry Devices
Biometry and Intraocular Lens Power Calculation by Combined Scheimpflug-Placido Disc versus Optical Interferometry Devices
Purpose: To compare the results of the current gold standard, laser interferometry, and keratometry by the IOL-Master, with a newly developed Galilei G6 using raytracing software O...
Calculation of ocular magnification in phakic and pseudophakic eyes based on anterior segment OCT data
Calculation of ocular magnification in phakic and pseudophakic eyes based on anterior segment OCT data
AbstractPurposeThe purpose of this study is to develop a straightforward mathematical concept for determination of object to image magnification in both phakic and pseudophakic eye...
IOL calculation using paraxial matrix optics
IOL calculation using paraxial matrix optics
AbstractMatrix methods have a long tradition in paraxial physiological optics. They are especially suited to describe and handle optical systems in a simple and intuitive manner. W...
Impact and Correction of an Anterior Phakic Intraocular Lens on Swept-Source Optical Coherence Tomography Biometry
Impact and Correction of an Anterior Phakic Intraocular Lens on Swept-Source Optical Coherence Tomography Biometry
Purpose:
To evaluate the impact of anterior chamber phakic intraocular lens (pIOL) on swept-source optical coherence tomography (SS-OCT) biometric measurements and IOL ...
Incision size changes after cataract surgery with intraocular lens implantation: comparison of 2 preloaded IOL implantation injectors
Incision size changes after cataract surgery with intraocular lens implantation: comparison of 2 preloaded IOL implantation injectors
Purpose:
To compare incision size enlargement at different times of cataract surgery and visual outcomes after surgery in 2 different intraocular lens (IOL) delivery sy...
Clinical Applications of Anterior Segment Optical Coherence Tomography in Managing Phakic and Secondary Intraocular Lens Implants: A Comprehensive Review
Clinical Applications of Anterior Segment Optical Coherence Tomography in Managing Phakic and Secondary Intraocular Lens Implants: A Comprehensive Review
Anterior segment optical coherence tomography (AS-OCT) has emerged as a crucial imaging technique in ophthalmology, particularly for evaluating intraocular structures and the behav...
Factors Affecting Near Vision After Monofocal Intraocular Lens Implantation
Factors Affecting Near Vision After Monofocal Intraocular Lens Implantation
PURPOSE:
To identify factors that influence near vision after monofocal intraocular lens (IOL) implantation for distance vision.
METHO...
Twenty-Seven-Gauge Needle-Assisted Technique for Repositioning of Late Luxated or Subluxated Capsular Bag-Intraocular Lens Complex: Inverted Scleral Tunnel for Flanged Polypropylene Suture Technique
Twenty-Seven-Gauge Needle-Assisted Technique for Repositioning of Late Luxated or Subluxated Capsular Bag-Intraocular Lens Complex: Inverted Scleral Tunnel for Flanged Polypropylene Suture Technique
Abstract
Purpose: To describe a minimally invasive and atraumatic technique for managing a polypropylene suture-assisted scleral fixation intraocular lens-capsular bag comp...