Effect of low body mass index in outcome of Micra leadless pacemaker implantation
View through CrossRef
Background
Implantation of Micra transcatheter pacing system (TPS) requires introduction of a large caliber delivery catheter into patient’s venous system and right heart. While the investigational device exemption study population had a mean body mass index (BMI) of 27.6kg/m2, implantation outcome in patients with small body size requires further investigation. This study sort to evaluate the effect of low BMI in outcome of Micra TPS implantation Method
Consecutive patients undergoing Micra TPS implantation between 19 September 2015 and 24 May 2018 in a single tertiary referral center were studied. Procedure efficacy outcome was defined as successful implantation with threshold being low (≤2.0V/0.24msec) and stable (increase in threshold of ≤1.5V / 0.24 msec) tested at implantation and all clinical follow up in the first year. Procedure safety outcome was defined as absence of major complications in the first year after implantation. A primary analysis was planned to identify factors affecting the composite procedure efficacy and safety outcome, with the hypothesis that low BMI would be associated with poor composite outcome. A secondary analysis was planned to study the effect of BMI on final implant position, number of device deployment and procedure time.
Results
147 patients were included in the study. The mean BMI of the cohort was significantly lower than the investigational device exemption study (23.7±3.7kg/m2 vs 27.6±5.3 kg/m2, p
Micra leadless pacemaker implantation in patients with low BMI was associated with an unfavorable composite efficacy and safety outcomes. Implantation in patients with low BMI was associated with a higher right ventricular implantation position, requirg more deployment attempts and longer procedure time. More data is needed to determine the optimal pacing approach for this high-risk group of patients.
Hong Kong College of Cardiology
Title: Effect of low body mass index in outcome of Micra leadless pacemaker implantation
Description:
Background
Implantation of Micra transcatheter pacing system (TPS) requires introduction of a large caliber delivery catheter into patient’s venous system and right heart.
While the investigational device exemption study population had a mean body mass index (BMI) of 27.
6kg/m2, implantation outcome in patients with small body size requires further investigation.
This study sort to evaluate the effect of low BMI in outcome of Micra TPS implantation Method
Consecutive patients undergoing Micra TPS implantation between 19 September 2015 and 24 May 2018 in a single tertiary referral center were studied.
Procedure efficacy outcome was defined as successful implantation with threshold being low (≤2.
0V/0.
24msec) and stable (increase in threshold of ≤1.
5V / 0.
24 msec) tested at implantation and all clinical follow up in the first year.
Procedure safety outcome was defined as absence of major complications in the first year after implantation.
A primary analysis was planned to identify factors affecting the composite procedure efficacy and safety outcome, with the hypothesis that low BMI would be associated with poor composite outcome.
A secondary analysis was planned to study the effect of BMI on final implant position, number of device deployment and procedure time.
Results
147 patients were included in the study.
The mean BMI of the cohort was significantly lower than the investigational device exemption study (23.
7±3.
7kg/m2 vs 27.
6±5.
3 kg/m2, p
Micra leadless pacemaker implantation in patients with low BMI was associated with an unfavorable composite efficacy and safety outcomes.
Implantation in patients with low BMI was associated with a higher right ventricular implantation position, requirg more deployment attempts and longer procedure time.
More data is needed to determine the optimal pacing approach for this high-risk group of patients.
Related Results
P185 FOLLOW–UP OF PATIENTS UNDERGOING POST–EXTRACTION LEADLESS PACEMAKER IMPLANTATION
P185 FOLLOW–UP OF PATIENTS UNDERGOING POST–EXTRACTION LEADLESS PACEMAKER IMPLANTATION
Abstract
Introduction
The implantation of traditional stimulation devices involves numerous complications, in the short and long...
Aveir Leadless Pacemaker implantation in pediatric population: a case series
Aveir Leadless Pacemaker implantation in pediatric population: a case series
Abstract
Background
While the Medtronic Micra pacemaker provided a small device for leadless pacemaker implantation, the Aveir d...
Leadless Pacemaker tine fracture and dislocation: two case reports
Leadless Pacemaker tine fracture and dislocation: two case reports
Abstract
The rate of leadless pacemaker implantation is increasing worldwide. To date, there have been few reports of leadless pacemaker dislocation and extraction. This ar...
Tijelo u opusu Janka Polića Kamova
Tijelo u opusu Janka Polića Kamova
The doctoral disertation is dedicated to the concept of the body in the works of Janko Polić Kamov. The body is approached as a signifier system on the basis of which numerous and ...
P425Single procedure pace and ablate. evaluation of efficacy and safety comparing three different vascular routes
P425Single procedure pace and ablate. evaluation of efficacy and safety comparing three different vascular routes
Abstract
Funding Acknowledgements
No financial support for this study
Introduc...
The angle of the tines before the pull and hold test predicts engagement of the tines in Micra leadless pacemaker implantation
The angle of the tines before the pull and hold test predicts engagement of the tines in Micra leadless pacemaker implantation
AbstractBackgroundMicra leadless pacemaker is secured to the myocardium by engagement of at least 2/4 tines confirmed with pull and hold test. However, the pull and hold test is so...
Abstract 4367732: Leadless Pacemaker Implantation in the Region of Bachmann’s Bundle: Initial Results
Abstract 4367732: Leadless Pacemaker Implantation in the Region of Bachmann’s Bundle: Initial Results
Background:
Leadless pacemakers are an emerging technology offering a less invasive alternative to traditional transvenous systems. Bachmann bundle (BB) pacing ...
The fixation tines of the Micra™ leadless pacemaker are atraumatic to the tricuspid valve
The fixation tines of the Micra™ leadless pacemaker are atraumatic to the tricuspid valve
AbstractBackgroundToday, there is no manufacturer‐supplied retrieval tool for the Micra™ pacemaker (Medtronic, Minneapolis, MN, USA); therefore, off‐the‐shelf catheters have been e...