Applanation Ultrasound versus Optical Biometry in the measurement of Axial Length in a Normal Eye
View through CrossRef
Objectives
Accurate measurement of the axial length (AL) of the eye is important to make sure that the patient is emmetropic postoperatively. Multiple studies conducted around the globe compare AL using optical and ultrasound biometry. However, results are still inconclusive, as some studies stated that there is no difference in both modalities, whereas other studies considered optical biometry to be more superior. We aim to compare AL measurements using applanation ultrasound and optical biometry AL Scan, in Islamabad, in order to find out which modality is more superior.
Methods
This was a prospective cross-sectional study that was conducted at the department of Ophthalmology, PAF Hospital, E9 Islamabad. The study was conducted over a period of eight months, from April to November 2023. Participants were enrolled using non-probability, consecutive sampling. The axial length of each eye was measured by both optical and ultrasound biometer.
Results
There were 80 eyes of 40 participants including both males (77.5%) and females (22.5%) aged between 16-58 years. The axial length using the optical biometer showed a greater value ranging from 22.01 to 30.18mm with a mean value of 23.89 ± 1.31mm whereas the axial length measured using applanation ultrasound was 21.62 to 30.00mm with a mean of 23.53 ± 1.29mm. The mean difference between both was 0.36 ± 0.21mm which was statistically significant (P <0.005).
Conclusion
There is a statistically significant difference between the mean readings of the AL measured by the IOL master and ultrasound biometer, so both the devices cannot be used interchangeably.
Islamabad Medical and Dental College
Title: Applanation Ultrasound versus Optical Biometry in the measurement of Axial Length in a Normal Eye
Description:
Objectives
Accurate measurement of the axial length (AL) of the eye is important to make sure that the patient is emmetropic postoperatively.
Multiple studies conducted around the globe compare AL using optical and ultrasound biometry.
However, results are still inconclusive, as some studies stated that there is no difference in both modalities, whereas other studies considered optical biometry to be more superior.
We aim to compare AL measurements using applanation ultrasound and optical biometry AL Scan, in Islamabad, in order to find out which modality is more superior.
Methods
This was a prospective cross-sectional study that was conducted at the department of Ophthalmology, PAF Hospital, E9 Islamabad.
The study was conducted over a period of eight months, from April to November 2023.
Participants were enrolled using non-probability, consecutive sampling.
The axial length of each eye was measured by both optical and ultrasound biometer.
Results
There were 80 eyes of 40 participants including both males (77.
5%) and females (22.
5%) aged between 16-58 years.
The axial length using the optical biometer showed a greater value ranging from 22.
01 to 30.
18mm with a mean value of 23.
89 ± 1.
31mm whereas the axial length measured using applanation ultrasound was 21.
62 to 30.
00mm with a mean of 23.
53 ± 1.
29mm.
The mean difference between both was 0.
36 ± 0.
21mm which was statistically significant (P <0.
005).
Conclusion
There is a statistically significant difference between the mean readings of the AL measured by the IOL master and ultrasound biometer, so both the devices cannot be used interchangeably.
Related Results
Comparison of Ultrasound Biometry with Optical Biometry for Measurement of Axial Length and Calculation of Intraocular Lens Power in Patients Undergoing Routine Cataract Surgery
Comparison of Ultrasound Biometry with Optical Biometry for Measurement of Axial Length and Calculation of Intraocular Lens Power in Patients Undergoing Routine Cataract Surgery
Objective: To compare the axial length and intraocular lens power calculated by using ultrasound biometry and optical biometry in patients with cataract
Methodology: It was prospec...
On Flores Island, do "ape-men" still exist? https://www.sapiens.org/biology/flores-island-ape-men/
On Flores Island, do "ape-men" still exist? https://www.sapiens.org/biology/flores-island-ape-men/
<span style="font-size:11pt"><span style="background:#f9f9f4"><span style="line-height:normal"><span style="font-family:Calibri,sans-serif"><b><spa...
Comparative Analysis of Refractive Outcomes Using Optical and Ultrasound Biometry in Phacoemulsification Cataract Surgery
Comparative Analysis of Refractive Outcomes Using Optical and Ultrasound Biometry in Phacoemulsification Cataract Surgery
Background: Accurate biometry is critical for optimal visual outcomes in cataract surgery. Both optical and ultrasound biometry are widely used, but their comparative effectiveness...
COMPARISON OF EYE AXIAL LENGTH AMONG YOUNG AND OLD HEALTHY VOLUNTEERS MEASURED ON TRANSORBITAL ULTRASOUND
COMPARISON OF EYE AXIAL LENGTH AMONG YOUNG AND OLD HEALTHY VOLUNTEERS MEASURED ON TRANSORBITAL ULTRASOUND
Background: Transorbital ultrasonography is a non-invasive, accessible, and effective imaging modality for ocular biometry, particularly valuable for measuring axial length—the dis...
Sleep Habits and Occurrence of Lowback Pain among Craftsmen
Sleep Habits and Occurrence of Lowback Pain among Craftsmen
<span style="color: #000000; font-family: Verdana, Arial, Helvetica, sans-serif; font-size: 10px; font-style: normal; font-variant-ligatures: normal; font-variant-caps: normal; ...
Sleep Habits and Occurrence of Lowback Pain among Craftsmen
Sleep Habits and Occurrence of Lowback Pain among Craftsmen
<span style="color: #000000; font-family: Verdana, Arial, Helvetica, sans-serif; font-size: 10px; font-style: normal; font-variant-ligatures: normal; font-variant-caps: normal; ...
Axial Length of Eyeball in Emmetropic Eye —A study in Rajshahi Medical College Hospital
Axial Length of Eyeball in Emmetropic Eye —A study in Rajshahi Medical College Hospital
Introduction: Axial length is one of the most important parameter for optometry and oculoplastic surgery. This variable is very crucial in biometry for the measurement of the power...
Visual outcome by using immersion biometry technique in senile cataract patients
Visual outcome by using immersion biometry technique in senile cataract patients
Background: Phacoemulsification with implantation of a foldable lens has evolved into a refractive surgery procedure in which optimal ocular biometry is essential for favorable vis...