Evaluation of Well-Type Chamber Calibration Factor for Measurement of Ir-192 Brachytherapy Source Over 10 Institutions

View through CrossRef
This paper describes about the evaluation of well-type chamber calibration factor for measurement Ir-192 brachytherapy source. Data were collected at ten institutions that have modalities of Ir-192 brachytherapy. The well-type chamber measures the Ir-192 brachytherapy sources in terms of air kerma strength. Calibration was carried out using the substitution method which adopted from a protocol published by the International Atomic Energy Agency. The well-type chamber standard used was HDR-1000 Plus. The results obtained that NAKS for well-type chamber from PTW manufacturers have a range of values from 9.063 x 105 – 9.875 x 105Gy m2 h-1 A-1 with an average value 9.522 x 105Gy m2 h-1 A-1, while the well-type chamber from Standard Imaging manufacturers have a range of values from 4.634 x 105 – 4.662 x 105Gy m2 h-1 A-1 with an average value 4.648 x 105Gy m2 h-1 A-1. The range of deviations obtained between NAKS new and NAKS old ranges from 0.8% to -1.8%, while the range of normalization results of NAKS new to the average value of NAKS new ranges from 0.952 - 1.029. There was one calibration factor that has a deviation of more than 1% over the NAKS old. Based on the calculation of uncertainty, all the calibration factor results obtained were still within the accepted range.
Title: Evaluation of Well-Type Chamber Calibration Factor for Measurement of Ir-192 Brachytherapy Source Over 10 Institutions
Description:
This paper describes about the evaluation of well-type chamber calibration factor for measurement Ir-192 brachytherapy source.
Data were collected at ten institutions that have modalities of Ir-192 brachytherapy.
The well-type chamber measures the Ir-192 brachytherapy sources in terms of air kerma strength.
Calibration was carried out using the substitution method which adopted from a protocol published by the International Atomic Energy Agency.
The well-type chamber standard used was HDR-1000 Plus.
The results obtained that NAKS for well-type chamber from PTW manufacturers have a range of values from 9.
063 x 105 – 9.
875 x 105Gy m2 h-1 A-1 with an average value 9.
522 x 105Gy m2 h-1 A-1, while the well-type chamber from Standard Imaging manufacturers have a range of values from 4.
634 x 105 – 4.
662 x 105Gy m2 h-1 A-1 with an average value 4.
648 x 105Gy m2 h-1 A-1.
The range of deviations obtained between NAKS new and NAKS old ranges from 0.
8% to -1.
8%, while the range of normalization results of NAKS new to the average value of NAKS new ranges from 0.
952 - 1.
029.
There was one calibration factor that has a deviation of more than 1% over the NAKS old.
Based on the calculation of uncertainty, all the calibration factor results obtained were still within the accepted range.

Related Results

(Invited) Strategies for Calibration Cost Reduction in Heterogeneous Chemical Sensor Arrays
(Invited) Strategies for Calibration Cost Reduction in Heterogeneous Chemical Sensor Arrays
Introduction Heterogeneous gas sensor arrays coupled with machine learning algorithms have been proposed for a wide range of applications. However, i...
Nasopharyngeal intracavitary brachytherapy
Nasopharyngeal intracavitary brachytherapy
AbstractBACKGROUNDLocally persistent nasopharyngeal carcinoma (NPC) carries an increased risk of local failure if additional treatment is not given. It has been shown that intracav...
KEDUDUKAN AHLI BAHASA DALAM PEMBUKTIAN PERKARA PENCEMARAN NAMA BAIK (STUDI PUTUSAN NOMOR: 47/PID.SUS/2019/PN. MGT)
KEDUDUKAN AHLI BAHASA DALAM PEMBUKTIAN PERKARA PENCEMARAN NAMA BAIK (STUDI PUTUSAN NOMOR: 47/PID.SUS/2019/PN. MGT)
<em><span id="page3R_mcid52" class="markedContent"><span style="left: calc(var(--scale-factor)*125.30px); top: calc(var(--scale-factor)*539.11px); font-size: calc(va...
Prostatic artery embolization as a method for prevention of infravesical obstruction after low-dose-rate brachytherapy
Prostatic artery embolization as a method for prevention of infravesical obstruction after low-dose-rate brachytherapy
BACKGROUND: Infravesical obstruction in patients with localized prostate cancer caused by concomitant benign prostatic hyperplasia substantially limits the feasibility of low-dose-...
Is Advanced Age a Limiting Factor in the Practice of High Dose Rate Brachytherapy in Patients Treated for Cervical Cancer?
Is Advanced Age a Limiting Factor in the Practice of High Dose Rate Brachytherapy in Patients Treated for Cervical Cancer?
Purpose: To assess the percentage of elderly patients not receiving brachytherapy in our practice and identify the factors influencing the decision to forgo this therapeutic modali...
TATA LAKSANA BRAKHITERAPI CO-60 TEKNIK INTRAKAVITER LENGKAP PADA KASUS KANKER SERVIKS
TATA LAKSANA BRAKHITERAPI CO-60 TEKNIK INTRAKAVITER LENGKAP PADA KASUS KANKER SERVIKS
Background: Brachytherapy is a radiation treatment by bringing radiation sources closer to the primary tumor which aims to provide additional therapeutic doses (booster) after exte...
Pain and Complications Assessment in Gynecological Cancer Brachytherapy Patients Under Spinal Anesthesia
Pain and Complications Assessment in Gynecological Cancer Brachytherapy Patients Under Spinal Anesthesia
Background: Brachytherapy is a radiation therapy that can deliver radiation to cancer by minimizing radiation exposure to adjacent organs or tissues. In its implementation, brachyt...

Back to Top