Flexible Bronchoscope Damage and Repair Costs in a Bronchoscopy Teaching Unit

View through CrossRef
<i>Background:</i> A significant part of the budget of our bronchoscopy unit represents repair costs for damaged bronchoscopes. <i>Objectives:</i> The purpose of this study was to determine the frequency, type and cause of damage to the bronchoscope as well as the repair costs. <i>Methods:</i> Frequency, type and cause of bronchoscope damage and repair costs of 13 new bronchoscopes that were used between August 1, 2001, and December 31, 2006, were retrospectively studied. <i>Results:</i> We recorded 47 instances of bronchoscope damage during the study, which is 1 instance of damage/141.6 procedures. Six instances of damage (12.7%) were potentially preventable. The most frequent wear and tear damage was to the rubber sheath on the distal bending portion of flexible bronchoscopes, and the most frequently preventable damage was that of the suction channel of the bronchoscope. The repair costs totaled 34,950.00 EUR or 5.25 EUR/procedure. 17,781.00 EUR (50.9%) can be attributed to preventable damage. The use of bronchoscopes for educational purposes was not associated with a higher rate of bronchoscope damage at our institution. <i>Conclusions:</i> Only a small number of occurrences of bronchoscope damage in our unit are potentially preventable, but they still represent an important expense. The relatively low occurrence of preventable damage is a result of the successful bronchoscopy training program.
Title: Flexible Bronchoscope Damage and Repair Costs in a Bronchoscopy Teaching Unit
Description:
<i>Background:</i> A significant part of the budget of our bronchoscopy unit represents repair costs for damaged bronchoscopes.
<i>Objectives:</i> The purpose of this study was to determine the frequency, type and cause of damage to the bronchoscope as well as the repair costs.
<i>Methods:</i> Frequency, type and cause of bronchoscope damage and repair costs of 13 new bronchoscopes that were used between August 1, 2001, and December 31, 2006, were retrospectively studied.
<i>Results:</i> We recorded 47 instances of bronchoscope damage during the study, which is 1 instance of damage/141.
6 procedures.
Six instances of damage (12.
7%) were potentially preventable.
The most frequent wear and tear damage was to the rubber sheath on the distal bending portion of flexible bronchoscopes, and the most frequently preventable damage was that of the suction channel of the bronchoscope.
The repair costs totaled 34,950.
00 EUR or 5.
25 EUR/procedure.
17,781.
00 EUR (50.
9%) can be attributed to preventable damage.
The use of bronchoscopes for educational purposes was not associated with a higher rate of bronchoscope damage at our institution.
<i>Conclusions:</i> Only a small number of occurrences of bronchoscope damage in our unit are potentially preventable, but they still represent an important expense.
The relatively low occurrence of preventable damage is a result of the successful bronchoscopy training program.

Related Results

Study of Clinicoradiological Profile of Patients Undergoing Fiberoptic Bronchoscopy
Study of Clinicoradiological Profile of Patients Undergoing Fiberoptic Bronchoscopy
Introduction: Bronchoscopy is a procedure to visualize the tracheobronchial tree. There are three types of Bronchoscopy, rigid, flexible, and virtual Bronchoscopy. Rigid bronchosco...
Hemorrhage Risk of Patients Receiving Bronchoscope Examination or Treatment
Hemorrhage Risk of Patients Receiving Bronchoscope Examination or Treatment
Abstract Background: The type of procedures has changed and become more complex and sophisticated with the development and extensive use of bronchoscopy. This study aimed t...
The factors influencing bronchoscope passing through the glottis
The factors influencing bronchoscope passing through the glottis
Abstract Background Flexible bronchoscopy is an important procedure in respiratory diseases. It is crucial whether bronchoscope can successfully pass through the glottis. T...
Concomitant Medial Meniscal Root Repair with Extrusion Repair (Centralization Technique)
Concomitant Medial Meniscal Root Repair with Extrusion Repair (Centralization Technique)
Background: Meniscal extrusion is a phenomenon in which a degenerative posterior horn tear, radial tear, or root tear results in displacement of the body of the meniscu...
A systematic review and cost effectiveness analysis of reusable vs. single‐use flexible bronchoscopes
A systematic review and cost effectiveness analysis of reusable vs. single‐use flexible bronchoscopes
Summary The cost effectiveness of reusable vs. single‐use flexible bronchoscopy in the peri‐operative setting has yet to be determined. We therefore aimed to dete...
Clinical application of flexible bronchoscopy in neonatal respiratory diseases
Clinical application of flexible bronchoscopy in neonatal respiratory diseases
Abstract Background: Respiratory diseases are an important component of neonatal diseases, and flexible bronchoscopy is increasingly used in neonates. This study aimed to i...

Back to Top