Characterisation of blink dynamics using a high‐speed infrared imaging system
View through CrossRef
Abstract
Purpose
Development and validation of a non‐invasive imaging system to capture spontaneous blinking and characterise blink dynamics using a custom image analysis algorithm.
Methods
A pilot study investigated the influence of illumination type on blink dynamics. Spontaneous blinking was recorded in eight subjects using a high‐speed camera under two illumination conditions: white light and infrared. Custom semi‐automated software assessed palpebral aperture height (PAH), blink rate, blink speed, blink completeness and blink duration. The main clinical study compared two different image analysis techniques. Spontaneous blinking was recorded in 20 subjects using a high‐speed infrared camera. Blink speed and duration were determined using two techniques: manual tracking and semi‐automated tracking. Agreement between the two techniques was assessed using Bland‐Altman analysis. Coefficients of repeatability (COR) were calculated for the semi‐automated technique.
Results
There were significant differences between white light and infrared illumination for PAH (
p
< 0.0001), blink rate (
p
= 0.04), closing speed (
p
= 0.009), opening speed (
p
< 0.0001) and blink duration (
p
= 0.0003). The mean differences (95% limits of agreement) between the two techniques were 0.6 mm s
−1
(−15.9 to 17.0) closing speed, 1.5 mm s
−1
(−6.8 to 9.8) opening speed, 2.4 ms (−6.8 to 11.5) closed phase duration and 5.0 ms (−19.4 to 29.3) total blink duration. COR values were 10.1 mm s
−1
closing speed, 3.7 mm s
−1
opening speed, 6.7 ms closed phase duration and 11.2 ms total blink duration.
Conclusions
This study has shown that spontaneous blinking can be characterised using a non‐invasive imaging system. The semi‐automated analysis provides a rapid characterisation of blink dynamics, allowing its application in large‐scale trials in a number of clinical areas.
Springer Science and Business Media LLC
Title: Characterisation of blink dynamics using a high‐speed infrared imaging system
Description:
Abstract
Purpose
Development and validation of a non‐invasive imaging system to capture spontaneous blinking and characterise blink dynamics using a custom image analysis algorithm.
Methods
A pilot study investigated the influence of illumination type on blink dynamics.
Spontaneous blinking was recorded in eight subjects using a high‐speed camera under two illumination conditions: white light and infrared.
Custom semi‐automated software assessed palpebral aperture height (PAH), blink rate, blink speed, blink completeness and blink duration.
The main clinical study compared two different image analysis techniques.
Spontaneous blinking was recorded in 20 subjects using a high‐speed infrared camera.
Blink speed and duration were determined using two techniques: manual tracking and semi‐automated tracking.
Agreement between the two techniques was assessed using Bland‐Altman analysis.
Coefficients of repeatability (COR) were calculated for the semi‐automated technique.
Results
There were significant differences between white light and infrared illumination for PAH (
p
< 0.
0001), blink rate (
p
= 0.
04), closing speed (
p
= 0.
009), opening speed (
p
< 0.
0001) and blink duration (
p
= 0.
0003).
The mean differences (95% limits of agreement) between the two techniques were 0.
6 mm s
−1
(−15.
9 to 17.
0) closing speed, 1.
5 mm s
−1
(−6.
8 to 9.
8) opening speed, 2.
4 ms (−6.
8 to 11.
5) closed phase duration and 5.
0 ms (−19.
4 to 29.
3) total blink duration.
COR values were 10.
1 mm s
−1
closing speed, 3.
7 mm s
−1
opening speed, 6.
7 ms closed phase duration and 11.
2 ms total blink duration.
Conclusions
This study has shown that spontaneous blinking can be characterised using a non‐invasive imaging system.
The semi‐automated analysis provides a rapid characterisation of blink dynamics, allowing its application in large‐scale trials in a number of clinical areas.
Related Results
Dry Eye and Blink Rate Simulation with a Pig Eye Model
Dry Eye and Blink Rate Simulation with a Pig Eye Model
ABSTRACT
Purpose.
To simulate medium level “dry eye” and investigate the effect of “blink” rates in “dry eye” condition u...
Quantitative EEG Features and Machine Learning Classifiers for Eye-Blink Artifact Detection: A Comparative Study
Quantitative EEG Features and Machine Learning Classifiers for Eye-Blink Artifact Detection: A Comparative Study
Abstract
Ocular artifact, namely eye-blink artifact, is an unavoidable and one of the most destructive noises in EEG signals. Many solutions were proposed regarding the det...
Quantitative EEG Features and Machine Learning Classifiers for Eye-Blink Artifact Detection: A Comparative Study
Quantitative EEG Features and Machine Learning Classifiers for Eye-Blink Artifact Detection: A Comparative Study
Abstract
Ocular artifact, namely eye-blink artifact, is an inevitable and one of the most destructive noises of EEG signals. Many solutions of detecting the eye-blink artif...
Feature Extraction from Regenerated EEG – A Better Approach for ICA Based Eye Blink Artifact Detection
Feature Extraction from Regenerated EEG – A Better Approach for ICA Based Eye Blink Artifact Detection
Purpose: Independent Component Analysis (ICA) decomposition is a commonly used technique for eye blink artifact detection from Electroencephalogram (EEG) signals. Feature extractio...
A role for attentional bias in cognitive deficits in chronic pain?
A role for attentional bias in cognitive deficits in chronic pain?
<p>This thesis extends current understanding of cognitive deficits in people with chronic pain, specifically those related to attention. Researchers have proposed that attent...
Effects of addiction-associated and affective stimuli on the attentional blink in a sample of abstinent opiate dependent patients
Effects of addiction-associated and affective stimuli on the attentional blink in a sample of abstinent opiate dependent patients
The attentional blink reveals the limits of the brain's ability in information processing. It has been extensively studied in people with neurological and psychiatric disturbances ...
Rancang Bangun Sistem Terapi Infrared Otomatis Untuk Terapi Far-Infrared Pada Spinal Cord
Rancang Bangun Sistem Terapi Infrared Otomatis Untuk Terapi Far-Infrared Pada Spinal Cord
Terapi infrared adalah terapi yang banyak dilakukan dengan memanfaatkan efek panas yang ditimbulkan pada tubuh manusia terutama pada permukaan kulit yang terpapar secara langsung. ...
Design of high transparent infrared stealth thin films based on FTO/Ag/FTO structure
Design of high transparent infrared stealth thin films based on FTO/Ag/FTO structure
Multi-spectral compatible stealth materials have become an imperative development trend, especially visible and infrared compatible stealth materials have become the most important...