Biomechanical Comparison of 3 Adjustable-Loop Suspensory Devices for All-Inside ACL Reconstruction: A Time-Zero Full-Construct Model

View through CrossRef
Background: Little is known about the stability of adjustable-loop devices (ALDs) for anterior cruciate ligament (ACL) reconstruction (ACLR). Purpose: To evaluate the stabilization behavior of 3 different types of ALDs for all-inside ACLR in a full-construct surgical technique-based manner. Study Design: Controlled laboratory study. Methods: The femoral and tibial devices of Ultrabutton (Smith & Nephew), Infinity (Conmed), and TightRope II (Arthrex) were applied to quadrupled bovine tendon grafts (n = 8 each) with tibial-sided traction applied (350 N) for graft tensioning in a simulated fully extended knee. Knotless femoral graft fixation was based on either a suture-locking device (SLD; Ultrabutton), button-locking device (BLD; Infinity), or dual-locking device (DLD; TightRope II). All constructs were progressively loaded (50 N/500 cycles) from 50 to 300 N for 3000 cycles (0.75 Hz), including complete unloading situations and pull to failure (50 mm/min). Construct elongation, stiffness, and ultimate load were analyzed. Results: BLD showed significantly greater initial elongation (-2.69 ± 0.15 mm) than DLD (-3.19 ± 0.21 mm; P < .001) but behaved similarly to SLD (-2.93 ± 0.23 mm). While DLD and SLD had the smallest initial elongation at the same significance level, they behaved opposite to each other with gradually increasing peak loading. At the end of testing, DLD had the lowest (-0.64 ± 0.32 mm) and SLD the highest (3.41 ± 1.01 mm) total elongation ( P < .003 for both). SLD displayed significantly higher dynamic elongation (6.34 ± 0.23 mm) than BLD (3.21 ± 0.61 mm) and DLD (2.56 ± 0.31 mm) ( P < .001 for both). The failure load of BLD (865.0 ± 183.8 N) was significantly lower ( P < .026) compared with SLD and DLD (>1000 N). The predominant failure mode was suture rupture and tibial bone breakage with button subsidence (SLD, n = 4). No significant difference in stiffness between constructs was found. Conclusion: While DLD successfully restricted critical construct elongation, BLD partially and SLD completely exceeded the clinical failure threshold (>3 mm) of plastic elongation with loop lengthening during increasing cyclic peak loading with complete unloading. Higher failure loads of SLD and DLD implants (>1000 N) were achieved at similar construct stiffness to BLD. Clinical Relevance: A detailed biomechanical understanding of the stabilization potential is pertinent to the continued evolution of ALDs to improve clinical outcomes.
Title: Biomechanical Comparison of 3 Adjustable-Loop Suspensory Devices for All-Inside ACL Reconstruction: A Time-Zero Full-Construct Model
Description:
Background: Little is known about the stability of adjustable-loop devices (ALDs) for anterior cruciate ligament (ACL) reconstruction (ACLR).
Purpose: To evaluate the stabilization behavior of 3 different types of ALDs for all-inside ACLR in a full-construct surgical technique-based manner.
Study Design: Controlled laboratory study.
Methods: The femoral and tibial devices of Ultrabutton (Smith & Nephew), Infinity (Conmed), and TightRope II (Arthrex) were applied to quadrupled bovine tendon grafts (n = 8 each) with tibial-sided traction applied (350 N) for graft tensioning in a simulated fully extended knee.
Knotless femoral graft fixation was based on either a suture-locking device (SLD; Ultrabutton), button-locking device (BLD; Infinity), or dual-locking device (DLD; TightRope II).
All constructs were progressively loaded (50 N/500 cycles) from 50 to 300 N for 3000 cycles (0.
75 Hz), including complete unloading situations and pull to failure (50 mm/min).
Construct elongation, stiffness, and ultimate load were analyzed.
Results: BLD showed significantly greater initial elongation (-2.
69 ± 0.
15 mm) than DLD (-3.
19 ± 0.
21 mm; P < .
001) but behaved similarly to SLD (-2.
93 ± 0.
23 mm).
While DLD and SLD had the smallest initial elongation at the same significance level, they behaved opposite to each other with gradually increasing peak loading.
At the end of testing, DLD had the lowest (-0.
64 ± 0.
32 mm) and SLD the highest (3.
41 ± 1.
01 mm) total elongation ( P < .
003 for both).
SLD displayed significantly higher dynamic elongation (6.
34 ± 0.
23 mm) than BLD (3.
21 ± 0.
61 mm) and DLD (2.
56 ± 0.
31 mm) ( P < .
001 for both).
The failure load of BLD (865.
0 ± 183.
8 N) was significantly lower ( P < .
026) compared with SLD and DLD (>1000 N).
The predominant failure mode was suture rupture and tibial bone breakage with button subsidence (SLD, n = 4).
No significant difference in stiffness between constructs was found.
Conclusion: While DLD successfully restricted critical construct elongation, BLD partially and SLD completely exceeded the clinical failure threshold (>3 mm) of plastic elongation with loop lengthening during increasing cyclic peak loading with complete unloading.
Higher failure loads of SLD and DLD implants (>1000 N) were achieved at similar construct stiffness to BLD.
Clinical Relevance: A detailed biomechanical understanding of the stabilization potential is pertinent to the continued evolution of ALDs to improve clinical outcomes.

Related Results

Poster 247: Muscle ERRγ Overexpression Mitigates the Muscle Atrophy after ACL injury
Poster 247: Muscle ERRγ Overexpression Mitigates the Muscle Atrophy after ACL injury
Objectives: Anterior cruciate ligament (ACL) reconstruction is the 6th most common orthopedic procedure performed in the United States (1,2). There is substantial evidence to sugge...
Effect of cervical suspensory traction in the treatment of severe cervical kyphotic deformity
Effect of cervical suspensory traction in the treatment of severe cervical kyphotic deformity
ObjectiveThis study aimed to investigate a new noninvasive traction method on the treatment of severe cervical kyphotic deformity.MethodsThe clinical data of patients with severe c...
Subjective knee apprehension is not associated to physical parameters 6–12 months after anterior cruciate ligament reconstruction
Subjective knee apprehension is not associated to physical parameters 6–12 months after anterior cruciate ligament reconstruction
AbstractPurposeAnterior cruciate ligament (ACL) rupture is a common injury and psychological parameters measured at 6–8 months are said to be almost more predictive for return to s...
Artelon Graft Augmentation in Revision Anterior Cruciate Ligament Reconstruction: A Case Report
Artelon Graft Augmentation in Revision Anterior Cruciate Ligament Reconstruction: A Case Report
Introduction Anterior cruciate ligament (ACL) injuries are becoming more common in the United States. These commonly present following a non-contact twisting injury to the knee wit...
ANTERIOR CRUCIATE LIGAMENT, DESCRIPTION, RUPTURE AND TREATMENT
ANTERIOR CRUCIATE LIGAMENT, DESCRIPTION, RUPTURE AND TREATMENT
Introduction: the anterior cruciate ligament (ACL) along with the posterior cruciate ligament are the central stabilizers of the knee. Rupture of the ACL usually occurs in active a...
Inside-Out Repair of Medial Meniscal Ramp Lesions in Patients Undergoing Anterior Cruciate Ligament Reconstruction
Inside-Out Repair of Medial Meniscal Ramp Lesions in Patients Undergoing Anterior Cruciate Ligament Reconstruction
Background: Medial meniscal ramp lesions are disruptions at the meniscocapsular junction and/or meniscotibial attachment of the posterior horn of the medial meniscus, a...

Back to Top