Long‐Term Stapedectomy Results With the McGee Stapes Prosthesis
View through CrossRef
AbstractObjective To report long‐term follow up on patients undergoing stapedectomy with the McGee stapes prosthesis.Study Design A retrospective case series.Methods Charts recording 170 stapedectomies (120 primary and 50 revision stapedectomies) from 1989 to 1999 were analyzed, and audiometric data and the findings at repeat exploration reported.Results Over a 10‐year period, 11.2% of patients had failure of their initial air‐bone gap closure with the McGee piston. Seventy‐seven percent (in 10 of 13 patients) of McGee pistons that failed were found to have the platinum ribbon displaced laterally from the incus and the piston pushed out of the stapedotomy. The failure rate was not significantly different (P = .72) from that in our patients who underwent placement of a Robinson cup prosthesis (9.5%), but the pattern of displacement was unique. Failures occurred earlier with the McGee piston (average time to failure of 2.5 y) than with the Robinson cup (average time to failure of 8.6 y).Conclusion The McGee piston was found to have a unique pattern of dislocation that is likely related to the malleable nature of the platinum ribbon and loosening of the crimped “shepherd's crook” over time.
Title: Long‐Term Stapedectomy Results With the McGee Stapes Prosthesis
Description:
AbstractObjective To report long‐term follow up on patients undergoing stapedectomy with the McGee stapes prosthesis.
Study Design A retrospective case series.
Methods Charts recording 170 stapedectomies (120 primary and 50 revision stapedectomies) from 1989 to 1999 were analyzed, and audiometric data and the findings at repeat exploration reported.
Results Over a 10‐year period, 11.
2% of patients had failure of their initial air‐bone gap closure with the McGee piston.
Seventy‐seven percent (in 10 of 13 patients) of McGee pistons that failed were found to have the platinum ribbon displaced laterally from the incus and the piston pushed out of the stapedotomy.
The failure rate was not significantly different (P = .
72) from that in our patients who underwent placement of a Robinson cup prosthesis (9.
5%), but the pattern of displacement was unique.
Failures occurred earlier with the McGee piston (average time to failure of 2.
5 y) than with the Robinson cup (average time to failure of 8.
6 y).
Conclusion The McGee piston was found to have a unique pattern of dislocation that is likely related to the malleable nature of the platinum ribbon and loosening of the crimped “shepherd's crook” over time.
Related Results
Stapedotomy or Stapedectomy: Does It Really Matter?
Stapedotomy or Stapedectomy: Does It Really Matter?
Abstract
Introduction Otosclerosis leads to stapes fixation and consequent conductive hearing loss. Surgery is the mainstay of treatment, and it can be achieved through s...
Preoperative prediction of stapes exposure during otomicrosurgery based on temporal bone computed tomography images:A cohort observational study
Preoperative prediction of stapes exposure during otomicrosurgery based on temporal bone computed tomography images:A cohort observational study
Introduction:
This prospective study aimed to evaluate the
predictive value of high-resolution computed tomography (CT) scan of the
temporal bone in predicting th...
Sound transmission by chamber prosthesis of the middle ear
Sound transmission by chamber prosthesis of the middle ear
Abstract
Tests done on specimens cut from the temporal bones show, that the stapedotomy can be more effective, if instead of the piston prosthesis, the ear chamber ...
Tympanoplasty with an Intact Stapes Superstructure in Chronic Otitis Media
Tympanoplasty with an Intact Stapes Superstructure in Chronic Otitis Media
BACKGROUND: The objective of our study was to ascertain the functional results in terms of air bone gap (ABG) closure over 4 frequencies (0.5, 1, 2, 3 kHz) in patients with chronic...
Abstract P6-14-10: A retrospective study of peri-prosthesis infection after prosthesis based breast reconstruction in a single breast center in east China and impact of an evidence-based protocol
Abstract P6-14-10: A retrospective study of peri-prosthesis infection after prosthesis based breast reconstruction in a single breast center in east China and impact of an evidence-based protocol
Abstract
Purpose
To investigate the status of surgical site infection after prosthesis-based breast reconstruction in a single center and evaluate the...
Experience with Stapedectomy in a Developing Country: A Review of 200 Cases
Experience with Stapedectomy in a Developing Country: A Review of 200 Cases
Since stapedectomy was first performed in 1956, many innovations have been described. We retrospectively analyzed 200 cases of stapedectomy that had been performed with standard ot...
Biomechanical comparison between the custom-made mandibular condyle prosthesis and total temporomandibular joint prosthesis in finite element analysis
Biomechanical comparison between the custom-made mandibular condyle prosthesis and total temporomandibular joint prosthesis in finite element analysis
Purpose: The purpose of this study was to compare the biomechanical behavior of the custom-made mandibular condyle prosthesis and total TMJ prosthesis. Methods: Three models of one...
Penile prosthesis – a viable solution for erectile dysfunction refractory to conservative therapy
Penile prosthesis – a viable solution for erectile dysfunction refractory to conservative therapy
Introduction: Erectile dysfunction(ED) is a serious condition which can affect men of all ages, with an important impact over the quality of life. When conservative therapy fails,...