The Application of Elliptic Cylindrical Phantom in Brachytherapy Dosimetric Study of HDR 192Ir Source
View through CrossRef
The purpose of this study is to determine the radial dose function of HDR 192Ir source based on Monte Carlo simulation using elliptic cylindrical phantom, similar to realistic shape of pelvis, in brachytherapy dosimetric study. The elliptic phantom size and shape was determined by analysis of dimensions of pelvis on CT images of 20 patients treated with brachytherapy for cervical cancer. The radial dose function obtained using the elliptic cylindrical water phantom was compared with radial dose functions for different spherical phantom sizes, including the Williamsion’s data loaded into conventional planning system. The differences in the radial dose function for the different spherical water phantoms increase with radial distance, r, and the largest differences in the radial dose function appear for the smallest phantom size. The radial dose function of the elliptic cylindrical phantom significantly decreased with radial distance in the vertical direction due to different scatter condition in comparison with the Williamson’s data. Considering doses to ICRU rectum and bladder points, doses to reference points can be underestimated up to 1-2% at the distance from 3 to 6 cm. The radial dose function in this study could be used as realistic data for calculating the brachytherapy dosimetry for cervical cancer.
Title: The Application of Elliptic Cylindrical Phantom in Brachytherapy Dosimetric Study of HDR 192Ir Source
Description:
The purpose of this study is to determine the radial dose function of HDR 192Ir source based on Monte Carlo simulation using elliptic cylindrical phantom, similar to realistic shape of pelvis, in brachytherapy dosimetric study.
The elliptic phantom size and shape was determined by analysis of dimensions of pelvis on CT images of 20 patients treated with brachytherapy for cervical cancer.
The radial dose function obtained using the elliptic cylindrical water phantom was compared with radial dose functions for different spherical phantom sizes, including the Williamsion’s data loaded into conventional planning system.
The differences in the radial dose function for the different spherical water phantoms increase with radial distance, r, and the largest differences in the radial dose function appear for the smallest phantom size.
The radial dose function of the elliptic cylindrical phantom significantly decreased with radial distance in the vertical direction due to different scatter condition in comparison with the Williamson’s data.
Considering doses to ICRU rectum and bladder points, doses to reference points can be underestimated up to 1-2% at the distance from 3 to 6 cm.
The radial dose function in this study could be used as realistic data for calculating the brachytherapy dosimetry for cervical cancer.
Related Results
Nasopharyngeal intracavitary brachytherapy
Nasopharyngeal intracavitary brachytherapy
AbstractBACKGROUNDLocally persistent nasopharyngeal carcinoma (NPC) carries an increased risk of local failure if additional treatment is not given. It has been shown that intracav...
SU‐GG‐T‐03: Dosimetric Study of HDR 192Ir and LDR 125I Brachytherapy Sources Using the Penelope Monte Carlo Code
SU‐GG‐T‐03: Dosimetric Study of HDR 192Ir and LDR 125I Brachytherapy Sources Using the Penelope Monte Carlo Code
Purpose: To determine the dosimetric parameters of two HDR models 192Ir (microSelectron, and M‐19), and LDR 125I brachytherapy sources. Method and Materials: In this study we used ...
Body image and perception among adults with and without phantom limb pain
Body image and perception among adults with and without phantom limb pain
AbstractBackgroundFollowing lower‐limb amputation, phantom limb pain (i.e., pain perceived as coming from the amputated portion of the limb) is common. Phantom limb pain may be ass...
TH‐SAM‐BRB‐01: HDR and LDR Brachytherapy: Everything You Need to Know
TH‐SAM‐BRB‐01: HDR and LDR Brachytherapy: Everything You Need to Know
The development of Brachytherapy began about 1899 following the purification of radium by Pierre and Marie Curie and has been developing rapidly and continually since. The focus of...
Dosimetric accuracy of a deterministic radiation transport based brachytherapy treatment planning system. Part I: Single sources and bounded homogeneous geometries
Dosimetric accuracy of a deterministic radiation transport based brachytherapy treatment planning system. Part I: Single sources and bounded homogeneous geometries
Purpose:The aim of this work is to validate a deterministic radiation transport based treatment planning system (TPS) for single brachytherapy source dosimetry in homogeneous water...
A dosimetric comparison of versus for HDR prostate brachytherapy
A dosimetric comparison of versus for HDR prostate brachytherapy
For the purpose of evaluating the use of for prostate High Dose Rate brachytherapy (HDR), a hypothetical source is assumed with the exact same design of the new microSelectron so...
Abstract CT159: Olaparib combinations (OLAPCO)for homology-directed DNA repairdefects (HDR)
Abstract CT159: Olaparib combinations (OLAPCO)for homology-directed DNA repairdefects (HDR)
Abstract
INTRODUCTION: Genetic profiling of tumors used to identify treatment choices for cancer patients focus on kinase mutations and well-defined translocations; ...
COMPARATIVE STUDY OF CONFORMAL HDR BRACHYTHERAPY WITH ICRU POINTS DOSES AND TOXICITY ASSESSMENT
COMPARATIVE STUDY OF CONFORMAL HDR BRACHYTHERAPY WITH ICRU POINTS DOSES AND TOXICITY ASSESSMENT
Background: Toxicity assessment is done by comparing HDR brachytherapy with ICRU point doses. D2cc doses of OAR in HDR brachytherapy is slightly more than ICRU point doses. Object...