Validation of Sleep Measurements of an Actigraphy Watch: Instrument Validation Study (Preprint)
View through CrossRef
BACKGROUND
The iAide2 (Tokai) physical activity monitoring system includes diverse measurements and wireless features useful to researchers. The iAide2’s sleep measurement capabilities have not been compared to validated sleep measurement standards in any published work.
OBJECTIVE
We aimed to assess the iAide2’s sleep duration and total sleep time (TST) measurement performance and perform calibration if needed.
METHODS
We performed free-living sleep monitoring in 6 convenience-sampled participants without known sleep disorders recruited from within the Waki DTx Laboratory at the Graduate School of Medicine, University of Tokyo. To assess free-living sleep, we validated the iAide2 against a second actigraph that was previously validated against polysomnography, the MotionWatch 8 (MW8; CamNtech Ltd). The participants wore both devices on the nondominant arm, with the MW8 closest to the hand, all day except when bathing. The MW8 and iAide2 assessments both used the MW8 EVENT-marker button to record bedtime and risetime. For the MW8, MotionWare Software (version 1.4.20; CamNtech Ltd) provided TST, and we calculated sleep duration from the sleep onset and sleep offset provided by the software. We used a similar process with the iAide2, using iAide2 software (version 7.0). We analyzed 64 nights and evaluated the agreement between the iAide2 and the MW8 for sleep duration and TST based on intraclass correlation coefficients (ICCs).
RESULTS
The absolute ICCs (2-way mixed effects, absolute agreement, single measurement) for sleep duration (0.69, 95% CI –0.07 to 0.91) and TST (0.56, 95% CI –0.07 to 0.82) were moderate. The consistency ICC (2-way mixed effects, consistency, single measurement) was excellent for sleep duration (0.91, 95% CI 0.86-0.95) and moderate for TST (0.78, 95% CI 0.67-0.86). We determined a simple calibration approach. After calibration, the ICCs improved to 0.96 (95% CI 0.94-0.98) for sleep duration and 0.82 (95% CI 0.71-0.88) for TST. The results were not sensitive to the specific participants included, with an ICC range of 0.96-0.97 for sleep duration and 0.79-0.87 for TST when applying our calibration equation to data removing one participant at a time and 0.96-0.97 for sleep duration and 0.79-0.86 for TST when recalibrating while removing one participant at a time.
CONCLUSIONS
The measurement errors of the uncalibrated iAide2 for both sleep duration and TST seem too large for them to be useful as absolute measurements, though they could be useful as relative measurements. The measurement errors after calibration are low, and the calibration approach is general and robust, validating the use of iAide2’s sleep measurement functions alongside its other features in physical activity research.
JMIR Publications Inc.
Title: Validation of Sleep Measurements of an Actigraphy Watch: Instrument Validation Study (Preprint)
Description:
BACKGROUND
The iAide2 (Tokai) physical activity monitoring system includes diverse measurements and wireless features useful to researchers.
The iAide2’s sleep measurement capabilities have not been compared to validated sleep measurement standards in any published work.
OBJECTIVE
We aimed to assess the iAide2’s sleep duration and total sleep time (TST) measurement performance and perform calibration if needed.
METHODS
We performed free-living sleep monitoring in 6 convenience-sampled participants without known sleep disorders recruited from within the Waki DTx Laboratory at the Graduate School of Medicine, University of Tokyo.
To assess free-living sleep, we validated the iAide2 against a second actigraph that was previously validated against polysomnography, the MotionWatch 8 (MW8; CamNtech Ltd).
The participants wore both devices on the nondominant arm, with the MW8 closest to the hand, all day except when bathing.
The MW8 and iAide2 assessments both used the MW8 EVENT-marker button to record bedtime and risetime.
For the MW8, MotionWare Software (version 1.
4.
20; CamNtech Ltd) provided TST, and we calculated sleep duration from the sleep onset and sleep offset provided by the software.
We used a similar process with the iAide2, using iAide2 software (version 7.
0).
We analyzed 64 nights and evaluated the agreement between the iAide2 and the MW8 for sleep duration and TST based on intraclass correlation coefficients (ICCs).
RESULTS
The absolute ICCs (2-way mixed effects, absolute agreement, single measurement) for sleep duration (0.
69, 95% CI –0.
07 to 0.
91) and TST (0.
56, 95% CI –0.
07 to 0.
82) were moderate.
The consistency ICC (2-way mixed effects, consistency, single measurement) was excellent for sleep duration (0.
91, 95% CI 0.
86-0.
95) and moderate for TST (0.
78, 95% CI 0.
67-0.
86).
We determined a simple calibration approach.
After calibration, the ICCs improved to 0.
96 (95% CI 0.
94-0.
98) for sleep duration and 0.
82 (95% CI 0.
71-0.
88) for TST.
The results were not sensitive to the specific participants included, with an ICC range of 0.
96-0.
97 for sleep duration and 0.
79-0.
87 for TST when applying our calibration equation to data removing one participant at a time and 0.
96-0.
97 for sleep duration and 0.
79-0.
86 for TST when recalibrating while removing one participant at a time.
CONCLUSIONS
The measurement errors of the uncalibrated iAide2 for both sleep duration and TST seem too large for them to be useful as absolute measurements, though they could be useful as relative measurements.
The measurement errors after calibration are low, and the calibration approach is general and robust, validating the use of iAide2’s sleep measurement functions alongside its other features in physical activity research.
Related Results
Title: CHANGES IN SUBJECTIVE AND OBJECTIVE SLEEP MEASURES DURING RECOVERY IN PEDIATRIC CONCUSSION
Title: CHANGES IN SUBJECTIVE AND OBJECTIVE SLEEP MEASURES DURING RECOVERY IN PEDIATRIC CONCUSSION
Background: Sleep problems after concussion have been associated with longer recovery times and more severe symptoms. Early research has primarily evaluated sleep via subjective me...
The history of sleep research and sleep medicine in Europe
The history of sleep research and sleep medicine in Europe
SummarySleep became a subject of scientific research in the second half of the 19th century. Since sleep, unlike other physiological functions, cannot be attributed to a specific o...
Nurse-delivered sleep restriction therapy to improve insomnia disorder in primary care: the HABIT RCT
Nurse-delivered sleep restriction therapy to improve insomnia disorder in primary care: the HABIT RCT
Background
Insomnia is a prevalent and distressing sleep disorder. Multicomponent cognitive–behavioural therapy is the recommended first-line treatment, but access remains extremel...
O016 Actigraphy: A review of practice down under
O016 Actigraphy: A review of practice down under
Abstract
Background
In 2018, the American Academy of Sleep Medicine (AASM) published practice guidelines for use of actigraphy i...
Median Preoptic Astrocytes: Role in Sleep Regulation and Potential Mediators of Sex Differences
Median Preoptic Astrocytes: Role in Sleep Regulation and Potential Mediators of Sex Differences
One in three Americans suffer from chronic sleep disorders, and women are 40% more likely than men to experience sleep disorders. This disparity emerges at puberty and is strongly ...
PERSONALITY AND SELF-REPORTED AND ACTIGRAPHY-MEASURED SLEEP HEALTH IN ADULTHOOD
PERSONALITY AND SELF-REPORTED AND ACTIGRAPHY-MEASURED SLEEP HEALTH IN ADULTHOOD
Abstract
Personality may be associated with sleep health, however, the majority of existing studies rely on self-reported measures of sleep (often focusing on sleep ...
Open-source Longitudinal Sleep Analysis From Accelerometer Data (DPSleep): Algorithm Development and Validation (Preprint)
Open-source Longitudinal Sleep Analysis From Accelerometer Data (DPSleep): Algorithm Development and Validation (Preprint)
BACKGROUND
Wearable devices are now widely available to collect continuous objective behavioral data from individuals and to measure sleep.
...
A comparison of sleep quality in Costa Rican urban and rural older adults
A comparison of sleep quality in Costa Rican urban and rural older adults
Objective Evidence regarding sleep quality in older adults residing in urban and rural areas is lacking among Latino populations living outside the United States. The purpose of th...