Minimal Tourniquet Pressure to Maintain Arterial Closure in Upper Limb Surgery
View through CrossRef
Complications of the pneumatic tourniquet used during limb surgery result from excessive direct pressure. Traditional recommendations suggests parameters for maximum pressure and time limits rather than the minimal effective pressure to achieve a bloodless field. A clinical study was undertaken to evaluate the pneumatic tourniquet setting required for adequate haemostasis in the upper limb. The correlations between several possible influencing parameters (age, sex, arm fat thickness, extremity length, systolic, diastolic, and mean blood pressures) and the minimal pneumatic tourniquet pressure at which the peripheral pulse reappeared were studied in 50 patients undergoing surgery, using a Doppler stethoscope. The average Doppler Opening Pressure was 168.5 ± 42.7 mmHg and the only significant clinical variable was the mean blood pressure. From these results an equation was derived to predict the minimal effective tourniquet pressure. The mean calculated tourniquet pressure was 202.3 ± 34.2 mmHg, well below the 250 to 300 mmHg previously recommended. The technique consisted of inflating the tourniquet to a pressure of 300 mmHg, then reducing it to the calculated value. A bloodless field was maintained in all patients.
SAGE Publications
Title: Minimal Tourniquet Pressure to Maintain Arterial Closure in Upper Limb Surgery
Description:
Complications of the pneumatic tourniquet used during limb surgery result from excessive direct pressure.
Traditional recommendations suggests parameters for maximum pressure and time limits rather than the minimal effective pressure to achieve a bloodless field.
A clinical study was undertaken to evaluate the pneumatic tourniquet setting required for adequate haemostasis in the upper limb.
The correlations between several possible influencing parameters (age, sex, arm fat thickness, extremity length, systolic, diastolic, and mean blood pressures) and the minimal pneumatic tourniquet pressure at which the peripheral pulse reappeared were studied in 50 patients undergoing surgery, using a Doppler stethoscope.
The average Doppler Opening Pressure was 168.
5 ± 42.
7 mmHg and the only significant clinical variable was the mean blood pressure.
From these results an equation was derived to predict the minimal effective tourniquet pressure.
The mean calculated tourniquet pressure was 202.
3 ± 34.
2 mmHg, well below the 250 to 300 mmHg previously recommended.
The technique consisted of inflating the tourniquet to a pressure of 300 mmHg, then reducing it to the calculated value.
A bloodless field was maintained in all patients.
Related Results
Analysis of curative effect of low position tourniquet in calcaneal fracture surgery
Analysis of curative effect of low position tourniquet in calcaneal fracture surgery
Abstract
This study was designed to investigate the role and curative effect of low position tourniquet in calcaneal fracture surgery. We studied a total of 42 calcaneal fr...
Limb Occlusion Pressure Versus Standard Tourniquet Inflation Pressure in Minor Hand Surgery: A Randomized Controlled Trial
Limb Occlusion Pressure Versus Standard Tourniquet Inflation Pressure in Minor Hand Surgery: A Randomized Controlled Trial
Abstract
Background
In minor hand surgery, tourniquet is typically inflated to 250 mmHg. The pressure may be too high and cause unnecessary adverse effects. Limb occlusion...
Appropriate tourniquet use in surgery: A literature review
Appropriate tourniquet use in surgery: A literature review
Background: Tourniquet use is ubiquitous in the operating theatre. However, optimal tourniquet usage is unclear, including type, pressure, inflation time, or whether a tourniquet s...
The Pneumatic Ankle Tourniquet with Ankle Block Anesthesia for Foot Surgery
The Pneumatic Ankle Tourniquet with Ankle Block Anesthesia for Foot Surgery
The use of a pneumatic ankle tourniquet applied to the supramalleolar ankle region is a useful method of obtaining a bloodless field in surgery of the foot. The pneumatic ankle tou...
Phrenic Nerve Block for Management of Post-Thoracic Outlet Decompression Cough: A Case Report and Literature Review
Phrenic Nerve Block for Management of Post-Thoracic Outlet Decompression Cough: A Case Report and Literature Review
Abstract
Introduction
Thoracic outlet syndrome is a group of disorders arising from compressive forces on the neurovascular bundle in that region due to different etiologies. This...
The Use of the Pulse Oximeter in Limb Ischemia: The Pulse Study
The Use of the Pulse Oximeter in Limb Ischemia: The Pulse Study
Purpose: Does a non-invasive pulse oximeter probe accurately indicate limb ischemia with inflation of a tourniquet? How are the pulse waveform, perfusion index and oxygen saturatio...
Effects of tourniquet use on clinical outcomes and cement penetration in TKA when tranexamic acid administrated: a randomized controlled trial
Effects of tourniquet use on clinical outcomes and cement penetration in TKA when tranexamic acid administrated: a randomized controlled trial
Abstract
Background
The role of a tourniquet is still controversial for patients undergoing total knee arthroplasty (TKA). Our current study was per...
Impact of Tourniquet Use on Systemic Inflammatory Parameters, Functional Physical Recovery, and Cardiovascular Outcomes of Patients Undergoing Knee Arthroplasty: A Case-Control Study
Impact of Tourniquet Use on Systemic Inflammatory Parameters, Functional Physical Recovery, and Cardiovascular Outcomes of Patients Undergoing Knee Arthroplasty: A Case-Control Study
AbstractThe benefits of tourniquet use during orthopaedic surgery are controversial. We aim to investigate its effects on systemic inflammation, functional physical recovery, and c...