A Novel Semi‐Cannulated Screw Enhanced Bone Cement Augmentation and Pullout Strength in Posterior Cervical Lateral Mass Screw Fixations: An In Vitro Biomechanical and Morphological Study
View through CrossRef
ObjectiveTo develop a novel semi‐cannulated lateral mass screw (SC‐LMS) for cervical posterior fixations and compare the fixation stability and safety of SC‐LMS with regular solid lateral mass screw (S‐LMS) in bone cement augmentation and pullout strength using fresh cadaveric cervical vertebrae.MethodsThe conventional multiaxial screw for cervical lateral mass fixation was modified to a cannulated screw with two lateral holes, used for bone cement injection in situ. Eight fresh human cervical vertebrae (C3, C4, and C5) were collected and used. μCT scan was performed to evaluate the bone quality of the lateral masses, including bone mineral density (BMD), trabecular thickness (Tb.Th), and trabecular separation (Tb.Sp). SCLMS or S‐LMS were randomly inserted into the paired cervical vertebrae and pulled out as a screw loosening model. These screws were reinserted in with bone cement augmentation, scanned by μCT to obtain the bone cement distribution along the screws, and pulled out to test the screw purchase strength.ResultsFmax values exhibited strong positive correlations with the local BMD (???? = 0.8640, p < 0.0001) and Tb.Th (???? = 0.6795, p = 0.0038), whereas a negative correlation with Tb.Sp (???? = −0.5567, p = 0.0251). A significant difference was observed between the Fmax before and after PMMA injection on the SC‐LMS side (p = 0.019). The SC‐LMS exhibited lower risk of cement leakage than S‐LMS after PMMA injection, and a positive correlation was observed between ????max and the distribution volumes on the SC‐LMS side.ConclusionThe novel SC‐LMS provides more robust fixation stability and is safer than the S‐LMS for PMMA augmentation, which may be related to the cement‐screw‐cement‐bone complex formation.
Title: A Novel Semi‐Cannulated Screw Enhanced Bone Cement Augmentation and Pullout Strength in Posterior Cervical Lateral Mass Screw Fixations: An In Vitro Biomechanical and Morphological Study
Description:
ObjectiveTo develop a novel semi‐cannulated lateral mass screw (SC‐LMS) for cervical posterior fixations and compare the fixation stability and safety of SC‐LMS with regular solid lateral mass screw (S‐LMS) in bone cement augmentation and pullout strength using fresh cadaveric cervical vertebrae.
MethodsThe conventional multiaxial screw for cervical lateral mass fixation was modified to a cannulated screw with two lateral holes, used for bone cement injection in situ.
Eight fresh human cervical vertebrae (C3, C4, and C5) were collected and used.
μCT scan was performed to evaluate the bone quality of the lateral masses, including bone mineral density (BMD), trabecular thickness (Tb.
Th), and trabecular separation (Tb.
Sp).
SCLMS or S‐LMS were randomly inserted into the paired cervical vertebrae and pulled out as a screw loosening model.
These screws were reinserted in with bone cement augmentation, scanned by μCT to obtain the bone cement distribution along the screws, and pulled out to test the screw purchase strength.
ResultsFmax values exhibited strong positive correlations with the local BMD (???? = 0.
8640, p < 0.
0001) and Tb.
Th (???? = 0.
6795, p = 0.
0038), whereas a negative correlation with Tb.
Sp (???? = −0.
5567, p = 0.
0251).
A significant difference was observed between the Fmax before and after PMMA injection on the SC‐LMS side (p = 0.
019).
The SC‐LMS exhibited lower risk of cement leakage than S‐LMS after PMMA injection, and a positive correlation was observed between ????max and the distribution volumes on the SC‐LMS side.
ConclusionThe novel SC‐LMS provides more robust fixation stability and is safer than the S‐LMS for PMMA augmentation, which may be related to the cement‐screw‐cement‐bone complex formation.
Related Results
Frequency of Common Chromosomal Abnormalities in Patients with Idiopathic Acquired Aplastic Anemia
Frequency of Common Chromosomal Abnormalities in Patients with Idiopathic Acquired Aplastic Anemia
Objective: To determine the frequency of common chromosomal aberrations in local population idiopathic determine the frequency of common chromosomal aberrations in local population...
Lateral Mass Fixation in Subaxial Cervical Spine: Anatomic Review
Lateral Mass Fixation in Subaxial Cervical Spine: Anatomic Review
Introduction
The cervical spine is a highly mobile segment of the spinal column, liable to a variety of diseases and susceptible to trauma. It is a complex regi...
Bone Allograft Pedicle Screw Augmentation
Bone Allograft Pedicle Screw Augmentation
Study Design:
We performed a comprehensive cadaveric biomechanical study to compare the fixation strength of primary screws, screws augmented with bone allograft, and s...
The cement-bone bond is weaker than cement-cement bond in cement-in-cement revision arthroplasty. A comparative biomechanical study
The cement-bone bond is weaker than cement-cement bond in cement-in-cement revision arthroplasty. A comparative biomechanical study
This study compares the strength of the native bone-cement bond and the old-new cement bond under cyclic loading, using third generation cementing technique, rasping and contaminat...
Risk factors for bone cement displacement after percutaneous vertebral augmentation for osteoporotic vertebral compression fractures
Risk factors for bone cement displacement after percutaneous vertebral augmentation for osteoporotic vertebral compression fractures
PurposeTo explore the risk factors of bone cement displacement after percutaneous vertebral augmentation (PVA) in patients with osteoporotic vertebral compression fracture (OVCF).M...
Biomechanical Study: Determining the Optimum Insertion Angle for Screw‐In Suture Anchors—Is Deadman's Angle Correct?
Biomechanical Study: Determining the Optimum Insertion Angle for Screw‐In Suture Anchors—Is Deadman's Angle Correct?
Purpose
To assess the effect of the insertion angle and the angle of applied load on the pullout strength of screw‐in suture anchors.
...
FLY ASH FOUNDATION REINFORCED BY CEMENT–SOIL MIXING PILES
FLY ASH FOUNDATION REINFORCED BY CEMENT–SOIL MIXING PILES
Cement-soil mixing piles have been commonly used to enhance the bearing capacity of fly ash stratum and mitigate the settlement damage to the surrounding environment. However, only...
The anti-loosening characteristics of cortical bone trajectory screws and conventional pedicle screws under different bone quality conditions: a comparative study
The anti-loosening characteristics of cortical bone trajectory screws and conventional pedicle screws under different bone quality conditions: a comparative study
Abstract
Purpose: We aimed to compare the characteristics of anti-loosening and the pullout force between cortical bone trajectory (CBT) and conventional trajectory (CT) sc...