Recirculation in single lumen cannula venovenous extracorporeal membrane oxygenation: A non-randomized bi-centric trial

View through CrossRef
Background Recirculation is a common problem in venovenous (VV) extracorporeal membrane oxygenation (ECMO). The aims of this study were to compare recirculation fraction (R f ) between femoro-jugular and jugulo-femoral VV ECMO configurations, to identify risk factors for recirculation and to assess the impact on hemolysis. Methods Patients in the medical intensive care unit (ICU) at the University Medical Center Regensburg, Germany receiving VV ECMO with femoro-jugular, and jugulo-femoral configuration at the ECMO Center Karolinska, Sweden, were included in this non-randomized prospective study. Total ECMO flow ( Q EC ), recirculated flow (Q REC ), and recirculation fraction R f = Q REC /Q EC were determined using ultrasound dilution technology. Effective ECMO flow (Q EFF ) was defined as Q EFF = Q EC * (1–R f ). Demographics, cannula specifics, and markers of hemolysis were assessed. Survival was evaluated at discharge from ICU. Results Thirty-seven patients with femoro-jugular configuration underwent 595 single-point measurements and 18 patients with jugulo-femoral configuration 231 measurements. R f was lower with femoro-jugular compared to jugulo-femoral configuration [5 (0, 11) vs. 19 (13, 28) %, respectively ( p < 0.001)], resulting in similar Q EFF [2.80 (2.21, 3.39) vs. 2.79 (2.39, 3.08) L/min ( p = 0.225)] despite lower Q EC with femoro-jugular configuration compared to jugulo-femoral [3.01 (2.40, 3.70) vs. 3.57 (3.05, 4.06) L/min, respectively ( p < 0.001)]. In multivariate regression analysis, the type of configuration, distance between the two cannula tips, ECMO flow, and heart rate were significantly associated with Rf [B (95% CI): 25.8 (17.6, 33.8), p < 0.001; 960.4 (960.7, 960.1), p = 0.009; 4.2 (2.5, 5.9), p < 0.001; 960.1 (960.2, 0.0), p = 0.027]. Hemolysis was similar in subjects with Rf > 8 vs. ≤ 8%. Explorative data on survival showed comparable results in the femoro-jugular and the jugulo-femoral group (81 vs. 72%, p = 0.455). Conclusion VV ECMO with femoro-jugular configuration caused less recirculation. Further risk factors for higher R f were shorter distance between the two cannula tips, higher ECMO flow, and lower heart rate. R f did not affect hemolysis.
Title: Recirculation in single lumen cannula venovenous extracorporeal membrane oxygenation: A non-randomized bi-centric trial
Description:
Background Recirculation is a common problem in venovenous (VV) extracorporeal membrane oxygenation (ECMO).
The aims of this study were to compare recirculation fraction (R f ) between femoro-jugular and jugulo-femoral VV ECMO configurations, to identify risk factors for recirculation and to assess the impact on hemolysis.
Methods Patients in the medical intensive care unit (ICU) at the University Medical Center Regensburg, Germany receiving VV ECMO with femoro-jugular, and jugulo-femoral configuration at the ECMO Center Karolinska, Sweden, were included in this non-randomized prospective study.
Total ECMO flow ( Q EC ), recirculated flow (Q REC ), and recirculation fraction R f = Q REC /Q EC were determined using ultrasound dilution technology.
Effective ECMO flow (Q EFF ) was defined as Q EFF = Q EC * (1–R f ).
Demographics, cannula specifics, and markers of hemolysis were assessed.
Survival was evaluated at discharge from ICU.
Results Thirty-seven patients with femoro-jugular configuration underwent 595 single-point measurements and 18 patients with jugulo-femoral configuration 231 measurements.
R f was lower with femoro-jugular compared to jugulo-femoral configuration [5 (0, 11) vs.
19 (13, 28) %, respectively ( p < 0.
001)], resulting in similar Q EFF [2.
80 (2.
21, 3.
39) vs.
2.
79 (2.
39, 3.
08) L/min ( p = 0.
225)] despite lower Q EC with femoro-jugular configuration compared to jugulo-femoral [3.
01 (2.
40, 3.
70) vs.
3.
57 (3.
05, 4.
06) L/min, respectively ( p < 0.
001)].
In multivariate regression analysis, the type of configuration, distance between the two cannula tips, ECMO flow, and heart rate were significantly associated with Rf [B (95% CI): 25.
8 (17.
6, 33.
8), p < 0.
001; 960.
4 (960.
7, 960.
1), p = 0.
009; 4.
2 (2.
5, 5.
9), p < 0.
001; 960.
1 (960.
2, 0.
0), p = 0.
027].
Hemolysis was similar in subjects with Rf > 8 vs.
≤ 8%.
Explorative data on survival showed comparable results in the femoro-jugular and the jugulo-femoral group (81 vs.
72%, p = 0.
455).
Conclusion VV ECMO with femoro-jugular configuration caused less recirculation.
Further risk factors for higher R f were shorter distance between the two cannula tips, higher ECMO flow, and lower heart rate.
R f did not affect hemolysis.

Related Results

Trends in Neonatal Extracorporeal Membrane Oxygenation During a Venovenous Cannula Shortage*
Trends in Neonatal Extracorporeal Membrane Oxygenation During a Venovenous Cannula Shortage*
OBJECTIVES: To report temporal trends in venovenous extracorporeal membrane oxygenation (ECMO) use for neonatal respiratory failure in U.S. centers before and after fun...
Impact of Cannulation Strategy and Extracorporeal Blood Flow on Recirculation During Veno‐Venous Extracorporeal Membrane Oxygenation
Impact of Cannulation Strategy and Extracorporeal Blood Flow on Recirculation During Veno‐Venous Extracorporeal Membrane Oxygenation
ABSTRACTIntroductionVeno‐venous extracorporeal membrane oxygenation (V‐V ECMO) is increasingly used in the treatment of severe respiratory failure. Despite a significant increase i...
Effectiveness of Protocol Based Nursing Care of Patients with Intravenous Line Cannula in Selected Hospital at Ambala City
Effectiveness of Protocol Based Nursing Care of Patients with Intravenous Line Cannula in Selected Hospital at Ambala City
Health team members are always bound to be right for the betterment of patient, but when it becomes difficult to keep up there comes the need for a guidance to lead them on the rig...
Venoarterial extracorporeal membrane oxygenation in adult patients: predictors of mortality
Venoarterial extracorporeal membrane oxygenation in adult patients: predictors of mortality
Background: Extracorporeal membrane oxygenation is a cardiopulmonary supportive therapy. In this study, we reviewed our experience with extracorporeal membrane oxygenation support ...
Impact of Cannula Size on Clinical Outcomes in Peripheral Venoarterial Extracorporeal Membrane Oxygenation
Impact of Cannula Size on Clinical Outcomes in Peripheral Venoarterial Extracorporeal Membrane Oxygenation
Venoarterial extracorporeal membrane oxygenation (VA ECMO) is an effective mechanical circulatory support for cardiopulmonary failure. Conventionally, an arterial cannula over 15 F...
Pulsatile Right Ventricle to Pulmonary Artery Extracorporeal Membrane Oxygenation in a Pigs
Pulsatile Right Ventricle to Pulmonary Artery Extracorporeal Membrane Oxygenation in a Pigs
Background: We investigated the impact of right ventricle to pulmonary artery extracorporeal membrane oxygenation in acute respiratory dysfunction with or without pulsatile flow.  ...
Procedure for Western blot v1
Procedure for Western blot v1
Goal: This document has the objective of standardizing the protocol for Western blot. This technique allows the detection of specific proteins separated on polyacrylamide gel and t...
Recirculation in Veno-Venous Extracorporeal Membrane Oxygenation
Recirculation in Veno-Venous Extracorporeal Membrane Oxygenation
This review focuses on recirculation in the context of Veno-Venous Extracorporeal Life Support in adults. The methods employed to calculate and quantify the extent of recirculation...

Back to Top