Validation of an artificial intelligence system for holter ECG analysis
View through CrossRef
Abstract
Introduction
Holter ECG is a commonly used tool to identify heart rhythm disorders. As physician-based analysis of such Holter ECG is time consuming, various approaches using computerized interpretation have been developed. However, accuracy of these tools remains inferior to physician-based analysis. Thus, this study assesses the diagnostic performance of an artificial intelligence (AI)-powered Holter ECG service in comparison to results from reference databases as well as with physician-based evaluation.
Methods
231 ECG recordings, covering 2,261 hours of ECG with annotations for 10,024,315 heartbeats from the following reference databases were used: MIT-BIH Arrhythmia (MITDB), American Heart Association ECG (AHDB), MIT-BIH Atrial Fibrillation (AFDB), Long Term Atrial Fibrillation (LTAFDB). Those recordings were originally collected from a mixed population of inpatients and outpatients between 1975 and 2008. Analysis of these recordings regarding the detection of QRS complexes, QRS classification and detection of Atrial Fibrillation (AF) was performed using a novel service for AI-based ECG analysis. Results were compared to the diagnosis stated within the database.
QRS complex detection accuracy was performed on recordings from MITDB and AHADB by checking if time positions are within 150 ms to reference annotations. The QRS classification in three classes (normal, supraventricular, ventricular) was then compared for each detected heartbeat. Mismatches were adjudicated manually by three experts. Episodes of Atrial Fibrillation were detected in recordings from MITDB, AFDB and LTAFDB and compared to the reference episodes from the databases. The accuracy was evaluated based on the overlap with the reference annotations in the time-domain.
Results
Sensitivity and Precision of heartbeat detection compared to the reference databases exceeded 99% Figure 1 displays the results of heartbeat classification. Normal heartbeats were classified with a sensitivity of 99.6%, precision of 99.9% and specificity of 99.6%. For ventricular heartbeats, a sensitivity of 99.58 %, precision of 99.88 and specificity of 99.99% were achieved. Premature supraventricular contractions were recognized with a sensitivity of 99.58 %, precision of 100% and specificity of 100%.
When comparing the 248,187 QRS complex classifications of MITDB and AHADB to the AI-powered analysis, 128 discrepancies were revealed. 43 were due to a discrepancy of convention as these beats were in sections that were also marked as AF episodes by the databases and the AI-powered analysis marks all beats in AF episodes as normal. Of the 85 remaining discrepancies, more than 87% were adjudicated in favor of the AI-powered analysis by the experts.
Figure 2 shows the comparison of AF detection results showing Sensitivity and Precision above 99%.
Conclusion
AI-powered Holter ECG analysis compared to reference ECG stripes exceed 99% of sensitivity, specificity and precision.QRS complex detectionAF/ AFib detection
Title: Validation of an artificial intelligence system for holter ECG analysis
Description:
Abstract
Introduction
Holter ECG is a commonly used tool to identify heart rhythm disorders.
As physician-based analysis of such Holter ECG is time consuming, various approaches using computerized interpretation have been developed.
However, accuracy of these tools remains inferior to physician-based analysis.
Thus, this study assesses the diagnostic performance of an artificial intelligence (AI)-powered Holter ECG service in comparison to results from reference databases as well as with physician-based evaluation.
Methods
231 ECG recordings, covering 2,261 hours of ECG with annotations for 10,024,315 heartbeats from the following reference databases were used: MIT-BIH Arrhythmia (MITDB), American Heart Association ECG (AHDB), MIT-BIH Atrial Fibrillation (AFDB), Long Term Atrial Fibrillation (LTAFDB).
Those recordings were originally collected from a mixed population of inpatients and outpatients between 1975 and 2008.
Analysis of these recordings regarding the detection of QRS complexes, QRS classification and detection of Atrial Fibrillation (AF) was performed using a novel service for AI-based ECG analysis.
Results were compared to the diagnosis stated within the database.
QRS complex detection accuracy was performed on recordings from MITDB and AHADB by checking if time positions are within 150 ms to reference annotations.
The QRS classification in three classes (normal, supraventricular, ventricular) was then compared for each detected heartbeat.
Mismatches were adjudicated manually by three experts.
Episodes of Atrial Fibrillation were detected in recordings from MITDB, AFDB and LTAFDB and compared to the reference episodes from the databases.
The accuracy was evaluated based on the overlap with the reference annotations in the time-domain.
Results
Sensitivity and Precision of heartbeat detection compared to the reference databases exceeded 99% Figure 1 displays the results of heartbeat classification.
Normal heartbeats were classified with a sensitivity of 99.
6%, precision of 99.
9% and specificity of 99.
6%.
For ventricular heartbeats, a sensitivity of 99.
58 %, precision of 99.
88 and specificity of 99.
99% were achieved.
Premature supraventricular contractions were recognized with a sensitivity of 99.
58 %, precision of 100% and specificity of 100%.
When comparing the 248,187 QRS complex classifications of MITDB and AHADB to the AI-powered analysis, 128 discrepancies were revealed.
43 were due to a discrepancy of convention as these beats were in sections that were also marked as AF episodes by the databases and the AI-powered analysis marks all beats in AF episodes as normal.
Of the 85 remaining discrepancies, more than 87% were adjudicated in favor of the AI-powered analysis by the experts.
Figure 2 shows the comparison of AF detection results showing Sensitivity and Precision above 99%.
Conclusion
AI-powered Holter ECG analysis compared to reference ECG stripes exceed 99% of sensitivity, specificity and precision.
QRS complex detectionAF/ AFib detection.
Related Results
Inter-beat and intra-beat ECG interval analysis based on state space and hidden markov models
Inter-beat and intra-beat ECG interval analysis based on state space and hidden markov models
Analyse des intervalles ECG inter- et intra-battement sur des modèles d'espace d'état et de Markov cachés
Les maladies cardiovasculaires sont l'une des principales ...
Complex Deep Learning Models for Denoising of Human Heart ECG signals
Complex Deep Learning Models for Denoising of Human Heart ECG signals
Effective and powerful methods for denoising real electrocardiogram (ECG) signals are important for wearable sensors and devices. Deep Learning (DL) models have been used extensive...
Self-Supervised Electrocardiograph De-noising
Self-Supervised Electrocardiograph De-noising
Abstract
The electrocardiogram (ECG) records heart-beats and is potentially life-saving. However, the ECG signals (e.g., recorded by the standard ECG monitoring sys...
Development of AI-based method to detect the subtle ECG deviations from the population ECG norm
Development of AI-based method to detect the subtle ECG deviations from the population ECG norm
Abstract
Funding Acknowledgements
Type of funding sources: Public grant(s) – National budget only. Main funding source(s): Natio...
How Can We Identify the Best Implantation Site for an ECG Event Recorder?
How Can We Identify the Best Implantation Site for an ECG Event Recorder?
ZELLERHOFF, C., et al.: How Can We Identify the Best Implantation Site for an ECG Event Recorder? The aim of this study was to show how to find the preferable implantation site for...
A novel fetal ecg signal extraction from maternal ecg signal using conditional generative adversarial networks (CGAN)
A novel fetal ecg signal extraction from maternal ecg signal using conditional generative adversarial networks (CGAN)
Fetal Electrocardiogram (ECG) signal extraction from non-invasive abdominal ECG signal is one of the important clinical practices followed to observe the fetal health state. Inform...
Validation in Doctoral Education: Exploring PhD Students’ Perceptions of Belonging to Scaffold Doctoral Identity Work
Validation in Doctoral Education: Exploring PhD Students’ Perceptions of Belonging to Scaffold Doctoral Identity Work
Aim/Purpose: The aim of this article is to make a case of the role of validation in doctoral education. The purpose is to detail findings from three studies which explore PhD stude...
La luz: de herramienta a lenguaje. Una nueva metodología de iluminación artificial en el proyecto arquitectónico.
La luz: de herramienta a lenguaje. Una nueva metodología de iluminación artificial en el proyecto arquitectónico.
The constant development of artificial lighting throughout the twentieth century helped to
develop architecture to the current situation in which a new methodology is needed for
...