Abstract Number ‐ 63: A swine model to simulate the mechanism of aspiration thrombectomy.

View through CrossRef
Introduction Aspiration thrombectomy is an effective technique to recanalize large vessel occlusions (LVO) in stroke patients and has become one of the major techniques. Advancements in technology have enabled the use of catheters with larger lumens and higher aspiration forces; however, the successful reperfusion rate of the first pass remains at 57–63%. This poses a question as to why this rate is still suboptimal when conducting aspiration thrombectomy. We established a swine model for LVO, which provides concurrent fluoroscopic and transmural visualization of real‐time vessel responses during aspiration thrombectomy, to investigate the mechanism of aspiration thrombectomy. Methods Under general anesthesia, a common carotid artery (CCA) and a superficial cervical artery (SCA) of Yorkshire swine (n = 3) were surgically exposed, and an LVO was reproduced by injecting a radiopaque clot analog via a guiding catheter. Each target artery was treated using aspiration catheters with different inner diameters (0.058, 0.068, and 0.088 inches). The CCA group (n = 7) represented large diameter vessels (4‐5mm) simulating a human ICA, and the SCA group (n = 6), with bifurcations, represented small diameter vessels (2‐3mm) simulating a human MCA. To make a consistent clot location in the CCA group, mild, non‐flow‐limiting stenosis was created using a 4‐0 Prolene suture. Fluoroscopy and a high‐resolution digital microscope camera were used simultaneously to monitor angiographic and transmural vessel behavior during the procedure. Average vessel diameter, local blood pressures proximal and distal to the occlusion site, the presence or absence of vessel collapse / reverse flow during the procedures, and pre‐and post‐angiographic findings were evaluated. Results Both fluoroscopic and transmural visualization of mechanical thrombectomy was achieved in all vessels allowing the observation of real‐time arterial wall and clot response during the procedure. The mean blood pressures distal and proximal to the occlusion site showed no significant differences between the two groups. With remote aspiration, all vessels in the SCA group (Mean Diameter: 2.34 ± 0.49mm) showed immediate vessel collapse, and regardless of the lumen size of the catheter, they all failed to recanalize the vessels. Effective clot ingestion was achieved only when direct contact aspiration was applied (5 of 6 vessels). None of the vessels in the CCA group (Mean Diameter: 5.16 ± 0.54 mm) showed vessel collapse during remote aspiration. Six out of 7 vessels (85.7%) treated with remote aspiration showed local reverse flow followed by complete aspiration when the tip of the aspiration catheter was placed 5–20mm away from the clot. One CCA vessel required direct contact aspiration to achieve clot ingestion. Conclusions This swine model to analyze vessel behavior during aspiration thrombectomy may help us understand the mechanism of aspiration thrombectomy. The model may contribute to improving the techniques and the development of new aspiration devices.
Title: Abstract Number ‐ 63: A swine model to simulate the mechanism of aspiration thrombectomy.
Description:
Introduction Aspiration thrombectomy is an effective technique to recanalize large vessel occlusions (LVO) in stroke patients and has become one of the major techniques.
Advancements in technology have enabled the use of catheters with larger lumens and higher aspiration forces; however, the successful reperfusion rate of the first pass remains at 57–63%.
This poses a question as to why this rate is still suboptimal when conducting aspiration thrombectomy.
We established a swine model for LVO, which provides concurrent fluoroscopic and transmural visualization of real‐time vessel responses during aspiration thrombectomy, to investigate the mechanism of aspiration thrombectomy.
Methods Under general anesthesia, a common carotid artery (CCA) and a superficial cervical artery (SCA) of Yorkshire swine (n = 3) were surgically exposed, and an LVO was reproduced by injecting a radiopaque clot analog via a guiding catheter.
Each target artery was treated using aspiration catheters with different inner diameters (0.
058, 0.
068, and 0.
088 inches).
The CCA group (n = 7) represented large diameter vessels (4‐5mm) simulating a human ICA, and the SCA group (n = 6), with bifurcations, represented small diameter vessels (2‐3mm) simulating a human MCA.
To make a consistent clot location in the CCA group, mild, non‐flow‐limiting stenosis was created using a 4‐0 Prolene suture.
Fluoroscopy and a high‐resolution digital microscope camera were used simultaneously to monitor angiographic and transmural vessel behavior during the procedure.
Average vessel diameter, local blood pressures proximal and distal to the occlusion site, the presence or absence of vessel collapse / reverse flow during the procedures, and pre‐and post‐angiographic findings were evaluated.
Results Both fluoroscopic and transmural visualization of mechanical thrombectomy was achieved in all vessels allowing the observation of real‐time arterial wall and clot response during the procedure.
The mean blood pressures distal and proximal to the occlusion site showed no significant differences between the two groups.
With remote aspiration, all vessels in the SCA group (Mean Diameter: 2.
34 ± 0.
49mm) showed immediate vessel collapse, and regardless of the lumen size of the catheter, they all failed to recanalize the vessels.
Effective clot ingestion was achieved only when direct contact aspiration was applied (5 of 6 vessels).
None of the vessels in the CCA group (Mean Diameter: 5.
16 ± 0.
54 mm) showed vessel collapse during remote aspiration.
Six out of 7 vessels (85.
7%) treated with remote aspiration showed local reverse flow followed by complete aspiration when the tip of the aspiration catheter was placed 5–20mm away from the clot.
One CCA vessel required direct contact aspiration to achieve clot ingestion.
Conclusions This swine model to analyze vessel behavior during aspiration thrombectomy may help us understand the mechanism of aspiration thrombectomy.
The model may contribute to improving the techniques and the development of new aspiration devices.

Related Results

THE PIG AS A BIOMEDICAL MODEL: HOW THE NATIONAL SWINE RESOURCE AND RESEARCH CENTER CAN HELP YOU
THE PIG AS A BIOMEDICAL MODEL: HOW THE NATIONAL SWINE RESOURCE AND RESEARCH CENTER CAN HELP YOU
Swine are an excellent large animal model for human health and disease because their size and physiology are similar to humans, and the progression of many diseases, such as cardio...
Knowledge and practices related to swine flu in school students of Bhavnagar, Gujarat
Knowledge and practices related to swine flu in school students of Bhavnagar, Gujarat
Context: Swine Flu caused by Influenza virus of H1N1 strain is a highly infectious respiratory disease. Knowledge and awareness among the community regarding the containment of spr...
Swine flu awareness in pregnancy
Swine flu awareness in pregnancy
Background: Swine influenza outbreaks are common in pigs year-round and infection in humans is a result of close contact with infected animals. Understanding the perception of the ...
SWINE INFLUENZA
SWINE INFLUENZA
1. It has been possible to demonstrate, in Berkefeld filtrates of infectious material from experimental cases of swine influenza, a virus which when administered intranasally to su...
E-038 Acute histologic changes observed in swine arteries treated with a mechanical thrombectomy device: Abstract E-038 Figure 1
E-038 Acute histologic changes observed in swine arteries treated with a mechanical thrombectomy device: Abstract E-038 Figure 1
Background and purposeEffectiveness of currently used mechanical thrombectomy devices has been evaluated using animal models mainly based upon the angiographical data analysis. How...
Aspiration and leadership
Aspiration and leadership
PurposeManagement needs an aspiration as to how the company will work in the future. It needs a guide for corporate priorities. Any company ‐ local or global ‐ must be driven by an...

Back to Top