A comparison of unipolar and bipolar electrograms for cardiac pacemaker sensing.
View through CrossRef
Simultaneous unipolar and bipolar electrograms were recorded and compared from 49 pacemaker patients with bipolar endocardial electrodes. Average bipolar depolarization signal voltage equalled that of unipolar but showed greater variation. Bipolar and unipolar slew rates were equal in both mean and variance. The proximal pole voltage had little effect on the bipolar result in 8% of the cases, tended to cancel the tip voltage in 49% of the cases and augmented the tip voltage in 43% of the electrograms. The average bipolar R wave duration was 28% less, the T wave amplitude 34% less, and the ST-segment elevation 37% less than the unipolar values. By consistently attenuating the undersirable T waves and ST elevations, while leaving the depolarization signal unaffected, the bipolar electrode offered the advantage of a superior signal-to-noise ratio for sensing depolarization. In one case, however, the bipolar signal was so small as to cause a clinical sensing failure.
Title: A comparison of unipolar and bipolar electrograms for cardiac pacemaker sensing.
Description:
Simultaneous unipolar and bipolar electrograms were recorded and compared from 49 pacemaker patients with bipolar endocardial electrodes.
Average bipolar depolarization signal voltage equalled that of unipolar but showed greater variation.
Bipolar and unipolar slew rates were equal in both mean and variance.
The proximal pole voltage had little effect on the bipolar result in 8% of the cases, tended to cancel the tip voltage in 49% of the cases and augmented the tip voltage in 43% of the electrograms.
The average bipolar R wave duration was 28% less, the T wave amplitude 34% less, and the ST-segment elevation 37% less than the unipolar values.
By consistently attenuating the undersirable T waves and ST elevations, while leaving the depolarization signal unaffected, the bipolar electrode offered the advantage of a superior signal-to-noise ratio for sensing depolarization.
In one case, however, the bipolar signal was so small as to cause a clinical sensing failure.
Related Results
Relation of High‐Pass Filtered Unipolar Electrograms to Bipolar Electrograms during Ventricular Mapping
Relation of High‐Pass Filtered Unipolar Electrograms to Bipolar Electrograms during Ventricular Mapping
Background
: A filtered bipolar electrogram (EG) amplitude <1.5 mV is a robust indicator of relatively dense scar, but is influenced by the wavefront direction. Unipol...
Unipolar low-voltage in the left atrium, predictors and clinical significance in patients undergoing atrial fibrillation ablation
Unipolar low-voltage in the left atrium, predictors and clinical significance in patients undergoing atrial fibrillation ablation
Abstract
Background
Bipolar low-voltage areas (LVA) in the left atrium (LA) are well-known factors influencing the result...
The effects of aging on atrial endocardial electrograms in patients with paroxysmal atrial fibrillation
The effects of aging on atrial endocardial electrograms in patients with paroxysmal atrial fibrillation
AbstractBackground: The prevalence of atrial fibrillation (AF) has been reported to increase with advancing age. Histologic studies in AF have demonstrated that the percentage of f...
Bipolar disorders and suicidal behaviour
Bipolar disorders and suicidal behaviour
Rihmer Z, Kiss K. Bipolar disorders and suicidal behaviour.
Bipolar Disord 2002: 4(Suppl. 1): 21–25. © Blackwell Munksgaard, 2002Major depressive disorder is the leading cause of ...
Analisis Penyalahgunaan Zat sebagai Faktor Risiko Kejadian Gangguan Bipolar pada Orang Dewasa
Analisis Penyalahgunaan Zat sebagai Faktor Risiko Kejadian Gangguan Bipolar pada Orang Dewasa
Abstract. Bipolar disorder is an emotional disorder with recurrent episodes of mood swings and depression, followed by changes in activity or energy and associated with characteris...
Rod bipolar cells in the macaque monkey retina: immunoreactivity and connectivity
Rod bipolar cells in the macaque monkey retina: immunoreactivity and connectivity
Rod bipolar cells in the macaque monkey retina were labeled by three antibodies: an antibody against the alpha- and beta-subspecies of protein kinase C (PKC), a polyclonal antiseru...
Patients’ Perspective on Termination of Pacemaker Therapy—A Cross-Sectional Anonymous Survey among Patients Carrying a Pacemaker in Germany
Patients’ Perspective on Termination of Pacemaker Therapy—A Cross-Sectional Anonymous Survey among Patients Carrying a Pacemaker in Germany
Objective: To determine the opinions of patients regarding the withdrawal of pacemaker therapy. Participants and methods: A cross-sectional anonymous questionnaire was administered...
Epicardial Mapping: How to Measure Local Activation?
Epicardial Mapping: How to Measure Local Activation?
Epicardial ventricular mapping was performed in 5 dogs during sinus rhythm with a sock array containing 41 electrodes. Maps were generated with a computer‐assisted mapping system u...