Preventing Damage to Arthroscopic Lens During Surgery
View through CrossRef
Purpose
To evaluate the relation between the relative diameter of instrument tips and the distance between the arthroscopic lens and the tips of the instruments in terms of preventing lens damage during arthroscopy.
Methods
By use of a custom‐designed station device with 2 perpendicular portals, images of multiple‐sized shaver tips (2.5, 3.5, 4.0, 5.0, and 5.5 mm) were obtained from 4 different arthroscopes (4 mm 30°, 4 mm 70°, 2.7 mm 30°, and 2.7 mm 70°) at varying distances of 0 to 10 mm at 1‐mm increments. The relative diameter of the visualized instrument tips and the distance between the arthroscopic lens and the tips of the instruments were evaluated.
Results
The relative diameter of the shaver tips measured by 2 investigators showed excellent intraobserver and interobserver reliability. By use of 2 mm as the safe distance from the arthroscopic lens to the tip of the shaver, 13 of 14 arthroscopic lens–shaver combinations were considered safe when the visualized shaver tip was smaller than one‐half of the arthroscopic visual field. Six of 14 combinations were considered unsafe when the visualized shaver tip was larger than three‐fourths of the visual field.
Conclusions
In this experimental study, the safe distance of 2 mm could be maintained when arthroscopic instruments used during surgery (e.g., shavers and burrs) were observed to be less than one‐half of the arthroscopic visual field. The relative diameter of the visualized instrument tip was a reliable guide to prevent arthroscopic lens damage.
Clinical Relevance
To avoid damaging the arthroscope lens during surgery, arthroscopists can maintain a safe distance by keeping the relative diameter of the instrument tip (e.g., arthroscopic burr) to less than one‐half of the arthroscopic visual field.
Title: Preventing Damage to Arthroscopic Lens During Surgery
Description:
Purpose
To evaluate the relation between the relative diameter of instrument tips and the distance between the arthroscopic lens and the tips of the instruments in terms of preventing lens damage during arthroscopy.
Methods
By use of a custom‐designed station device with 2 perpendicular portals, images of multiple‐sized shaver tips (2.
5, 3.
5, 4.
0, 5.
0, and 5.
5 mm) were obtained from 4 different arthroscopes (4 mm 30°, 4 mm 70°, 2.
7 mm 30°, and 2.
7 mm 70°) at varying distances of 0 to 10 mm at 1‐mm increments.
The relative diameter of the visualized instrument tips and the distance between the arthroscopic lens and the tips of the instruments were evaluated.
Results
The relative diameter of the shaver tips measured by 2 investigators showed excellent intraobserver and interobserver reliability.
By use of 2 mm as the safe distance from the arthroscopic lens to the tip of the shaver, 13 of 14 arthroscopic lens–shaver combinations were considered safe when the visualized shaver tip was smaller than one‐half of the arthroscopic visual field.
Six of 14 combinations were considered unsafe when the visualized shaver tip was larger than three‐fourths of the visual field.
Conclusions
In this experimental study, the safe distance of 2 mm could be maintained when arthroscopic instruments used during surgery (e.
g.
, shavers and burrs) were observed to be less than one‐half of the arthroscopic visual field.
The relative diameter of the visualized instrument tip was a reliable guide to prevent arthroscopic lens damage.
Clinical Relevance
To avoid damaging the arthroscope lens during surgery, arthroscopists can maintain a safe distance by keeping the relative diameter of the instrument tip (e.
g.
, arthroscopic burr) to less than one‐half of the arthroscopic visual field.
Related Results
Outcome Of Arthroscopic Vs Mini Open Subacromial Decompression In Treating Shoulder Impingement Syndrome
Outcome Of Arthroscopic Vs Mini Open Subacromial Decompression In Treating Shoulder Impingement Syndrome
Introduction: Shoulder impingement syndrome is a common diagnosis leading to shoulder disabilities. Subacromial decompression surgery either arthroscopic or mini open is indicated ...
Arthroscopy-assisted reduction for Jacob type II pediatric humeral lateral condyle fractures: a clinical efficacy study
Arthroscopy-assisted reduction for Jacob type II pediatric humeral lateral condyle fractures: a clinical efficacy study
Background
Minimally invasive approaches are being increasingly employed in pediatric orthopedic trauma surgery. Clinical practice has seen applications of mini...
Predictors of Performance on the Arthrobox Arthroscopy Simulator for Medical Students
Predictors of Performance on the Arthrobox Arthroscopy Simulator for Medical Students
Purpose
The purpose of this study was to analyze the effects of past participation in athletics, the playing of musical instruments and video games and other va...
Arthroscopic ankle surgery in people with haemophilia
Arthroscopic ankle surgery in people with haemophilia
AbstractIntroductionPeople with haemophilia (PWH) not administered primary haematological prophylaxis since childhood, that is, those treated haematologically on demand or not trea...
Low-Cost, Low-Fidelity, Self-Made Arthroscopic Surgical Simulators: A Systematic Review
Low-Cost, Low-Fidelity, Self-Made Arthroscopic Surgical Simulators: A Systematic Review
Purpose: To systematically review the literature regarding low-cost, low-fidelity, self-made arthroscopic surgical simulators and provide an overview of their use in the teaching o...
Trends in hip preservation surgery in Japan from 2014 to 2019 with a focus on hip arthroscopic surgery
Trends in hip preservation surgery in Japan from 2014 to 2019 with a focus on hip arthroscopic surgery
ABSTRACT
Secondary hip osteoarthritis due to hip dysplasia is common among Japanese populations. This study aimed to investigate the number of hip preservation surge...
Morphometric OCT parameters of the lens under accommodative stimulus. Report 1. Assessment of age-related changes
Morphometric OCT parameters of the lens under accommodative stimulus. Report 1. Assessment of age-related changes
Introduction. The structural approach to studying the mechanism of accommodation and its age-related changes focuses on analyzing morphometric parameters (size, shape, and position...
Posterior trans‐septal portal for arthroscopic surgery of the knee joint
Posterior trans‐septal portal for arthroscopic surgery of the knee joint
Summary:
The authors introduce a safe arthroscopic technique of making a portal through the posterior septum of the knee joint, posterior to the posterior cruciat...