Abstract TAC117: Evaluating Wearable Cuffless Devices for Nocturnal Hypertension: Accuracy and Usability
View through CrossRef
Introduction:
Nocturnal hypertension, defined as elevated blood pressure (BP) during sleep, is a stronger predictor of cardiovascular morbidity and mortality than daytime hypertension. Although ambulatory blood pressure monitoring (ABPM) is considered the diagnostic gold standard, frequent cuff inflations can disrupt sleep, reduce comfort, and hinder adherence in clinical settings. Recent advances in wearable, cuffless BP monitors offer a promising alternative by enabling continuous, less intrusive measurements, potentially enhancing patient comfort and adherence.
Hypothesis:
We hypothesized that wearable, cuffless BP monitors accurately detect nocturnal hypertension and improve patient adherence compared to ABPM.
Methods:
Following PRISMA guidelines, we systematically searched PubMed and Google Scholar for studies published through January 2025 using keywords such as “blood pressure Monitoring, Ambulatory,” “Nocturnal Hypertension,” and “Wearable Electronic Devices.” Eligible studies compared wearable or cuffless BP devices with ABPM or other validated methods and reported diagnostic accuracy metrics or usability outcomes. Extracted data included study design, sample size, device type, measurement duration, BP differences, AUC values, and patient-reported outcomes.
Results:
Eight studies examined wearable, cuff-free BP monitors and found that their readings closely matched those from standard ABPM. Mean systolic BP differences ranged from −1.8 to 3.9 mmHg, and diastolic BP from −2.3 to 1.16 mmHg. Nighttime biases remained within clinical limits, with AUC values up to 1.0, indicating excellent accuracy. Technologies used included photoplethysmography, piezoelectric sensors, and ballistocardiography, which estimates BP through parameters like pulse transit time and pulse wave velocity. A pooled analysis of 16 studies (n=974) revealed small mean differences between wearable and reference devices, with average systolic and diastolic differences of 3.42 mmHg and 1.16 mmHg, respectively, supporting their clinical reliability. Usability studies reported improved comfort, fewer sleep disturbances, and higher adherence.
Conclusion:
Wearable, cuffless BP monitors offer high diagnostic accuracy and improved usability for detecting nocturnal hypertension. These devices show strong potential as alternatives to ABPM for routine clinical use and home monitoring. Further longitudinal studies are needed to evaluate their long-term impact on cardiovascular outcomes.
Ovid Technologies (Wolters Kluwer Health)
Title: Abstract TAC117: Evaluating Wearable Cuffless Devices for Nocturnal Hypertension: Accuracy and Usability
Description:
Introduction:
Nocturnal hypertension, defined as elevated blood pressure (BP) during sleep, is a stronger predictor of cardiovascular morbidity and mortality than daytime hypertension.
Although ambulatory blood pressure monitoring (ABPM) is considered the diagnostic gold standard, frequent cuff inflations can disrupt sleep, reduce comfort, and hinder adherence in clinical settings.
Recent advances in wearable, cuffless BP monitors offer a promising alternative by enabling continuous, less intrusive measurements, potentially enhancing patient comfort and adherence.
Hypothesis:
We hypothesized that wearable, cuffless BP monitors accurately detect nocturnal hypertension and improve patient adherence compared to ABPM.
Methods:
Following PRISMA guidelines, we systematically searched PubMed and Google Scholar for studies published through January 2025 using keywords such as “blood pressure Monitoring, Ambulatory,” “Nocturnal Hypertension,” and “Wearable Electronic Devices.
” Eligible studies compared wearable or cuffless BP devices with ABPM or other validated methods and reported diagnostic accuracy metrics or usability outcomes.
Extracted data included study design, sample size, device type, measurement duration, BP differences, AUC values, and patient-reported outcomes.
Results:
Eight studies examined wearable, cuff-free BP monitors and found that their readings closely matched those from standard ABPM.
Mean systolic BP differences ranged from −1.
8 to 3.
9 mmHg, and diastolic BP from −2.
3 to 1.
16 mmHg.
Nighttime biases remained within clinical limits, with AUC values up to 1.
0, indicating excellent accuracy.
Technologies used included photoplethysmography, piezoelectric sensors, and ballistocardiography, which estimates BP through parameters like pulse transit time and pulse wave velocity.
A pooled analysis of 16 studies (n=974) revealed small mean differences between wearable and reference devices, with average systolic and diastolic differences of 3.
42 mmHg and 1.
16 mmHg, respectively, supporting their clinical reliability.
Usability studies reported improved comfort, fewer sleep disturbances, and higher adherence.
Conclusion:
Wearable, cuffless BP monitors offer high diagnostic accuracy and improved usability for detecting nocturnal hypertension.
These devices show strong potential as alternatives to ABPM for routine clinical use and home monitoring.
Further longitudinal studies are needed to evaluate their long-term impact on cardiovascular outcomes.
Related Results
Applying a user-centered approach to evaluate the usability of a mobile application for health professionals in home care services (Preprint)
Applying a user-centered approach to evaluate the usability of a mobile application for health professionals in home care services (Preprint)
BACKGROUND
Mobile health (mHealth), or the use of mobile devices in medicine and health, is a sub-category of e-health. mHealth interventions are designed t...
Usability Quality Model: An Enhancement of Dromey's Model
Usability Quality Model: An Enhancement of Dromey's Model
ABSTRACT
Usability is a fundamental software quality attribute that strongly influences user performance, acceptance, and the overall success of software systems....
Effects of reduced blood pressure dipping on the progression of chronic kidney disease in children
Effects of reduced blood pressure dipping on the progression of chronic kidney disease in children
Abstract
Background: Hypertension is a common complication of chronic kidney disease (CKD) in children. It is related to the progression of CKD. However, current guidelines...
GW24-e3137 The etiology investigate of hypertension incidence in children
GW24-e3137 The etiology investigate of hypertension incidence in children
Objectives
Through retrospective analysis of related factors of children’s hypertension what is composed by primary hypertension and secondary hypertension diseas...
Review on hypertension diagnosis using expert system and wearable devices
Review on hypertension diagnosis using expert system and wearable devices
<span>The popularity of smartphones and wearable devices is increasing in the global market. These devices track physical exercise records, heartbeat, medicines, and self-hea...
P275 Uncontrolled post-clinic blood pressure readings can identify nocturnal non-dipping blood pressure pattern
P275 Uncontrolled post-clinic blood pressure readings can identify nocturnal non-dipping blood pressure pattern
Abstract
Funding Acknowledgements
Aga Khan University Faculty of Health Sciences Research Committee (#80096)
...
Contribution of Transcutaneous P
CO
2
in Obesity Hypoventilation Syndrome
Contribution of Transcutaneous P
CO
2
in Obesity Hypoventilation Syndrome
BACKGROUND:
Increasing numbers of patients have obesity-hypoventilation syndrome (OHS) because of the increasing prevalence of obesity. T...
PS-C19-4: NOCTURNAL HYPERTENSION AMONG TREATED HYPERTENSIVE PATIENTS
PS-C19-4: NOCTURNAL HYPERTENSION AMONG TREATED HYPERTENSIVE PATIENTS
Nocturnal hypertension among treated hypertensive patients Pagmadulam sukhbaatar t, Burmaa Badrakh 2, Naranchimeg Sodovsuren, 2Tsolmon Unurjargal2 t Mongolia-Japan University Hospi...