Search engine for discovering works of Art, research articles, and books related to Art and Culture
ShareThis
Javascript must be enabled to continue!

A novel device to micromanipulate oocytes during intracytoplasmic sperm injection

View through CrossRef
Objectives: To demonstrate a novel, non-pneumatic, compliant mechanism-based micro gripper to immobilize oocytes for the Intracytoplasmic sperm injection (ICSI). Material and Methods: The micro gripper is designed intuitively based on different techniques available to design compliant mechanisms in the literature such as the Stiffness Maps technique, Kinetoelastostatic maps, and feasibility maps techniques. The gripper’s holder was made from a 2mm thick PMMA sheet; whereas, the gripper was fabricated using a hydrophilic sheet, a proprietary material of 3MTM. The gripper and holder were assembled using a biocompatible adhesive. Results: Experimental trials carried out with the gripper holder on the ICSI workbench showed that the developed gripper holder was able to hold the oocyte gently and firmly in place. A micro linear actuator was used to actuate the gripper-holder. The device was tested for its efficacy to perform ICSI by designing ICSI experiments with matured oocytes and sperm; it was found that the degeneration rate was absolutely zero percent for all the matured oocytes. Conclusions: A novel device to micro manipulate oocytes during intra-cytoplasmic sperm injection is presented in this paper and is based on the gripping principle as opposed to the conventional suction-based pipettes for holding the oocytes gently and firmly in place during ICSI. The degeneration rate was found to be zero using the gripper-based novel device.
Title: A novel device to micromanipulate oocytes during intracytoplasmic sperm injection
Description:
Objectives: To demonstrate a novel, non-pneumatic, compliant mechanism-based micro gripper to immobilize oocytes for the Intracytoplasmic sperm injection (ICSI).
Material and Methods: The micro gripper is designed intuitively based on different techniques available to design compliant mechanisms in the literature such as the Stiffness Maps technique, Kinetoelastostatic maps, and feasibility maps techniques.
The gripper’s holder was made from a 2mm thick PMMA sheet; whereas, the gripper was fabricated using a hydrophilic sheet, a proprietary material of 3MTM.
The gripper and holder were assembled using a biocompatible adhesive.
Results: Experimental trials carried out with the gripper holder on the ICSI workbench showed that the developed gripper holder was able to hold the oocyte gently and firmly in place.
A micro linear actuator was used to actuate the gripper-holder.
The device was tested for its efficacy to perform ICSI by designing ICSI experiments with matured oocytes and sperm; it was found that the degeneration rate was absolutely zero percent for all the matured oocytes.
Conclusions: A novel device to micro manipulate oocytes during intra-cytoplasmic sperm injection is presented in this paper and is based on the gripping principle as opposed to the conventional suction-based pipettes for holding the oocytes gently and firmly in place during ICSI.
The degeneration rate was found to be zero using the gripper-based novel device.

Related Results

O-105 Chromatin accessibility of oocytes contributes to PCOS transgenerational inheritance
O-105 Chromatin accessibility of oocytes contributes to PCOS transgenerational inheritance
Abstract Study question What is the underlying mechanism contributing to the transgenerational defects of oocytes and embryos of...
Optimization of the TLR7/8 Activation-Based Sorting System for Goat Sperm
Optimization of the TLR7/8 Activation-Based Sorting System for Goat Sperm
Background:Current research indicates that the immunological separation method based on differentially expressed proteins in X- and Y-chromosome-bearing sperm represents a novel ap...

Back to Top