Javascript must be enabled to continue!
Clinical Testing of a New Pacemaker Function to Monitor Ventricular Capture
View through CrossRef
Automatic beat‐by‐beat capture functions are designed to minimize the pacing energy delivered, while maintaining the highest safety by delivering an immediate back‐up stimulus in case of loss of capture. The objective of this study was to estimate the lowering of ventricular pacing amplitude allowed by such a function, compared to amplitudes usually set manually in routine practice. An automatic ventricular pacing threshold test is launched every 6 hours to measure the automatic capture threshold (AT). From AT the function calculates: (1) the“capture amplitude”(Vc) = AT + 0.5 V at a minimum output of 1 V and (2) the“safety amplitude” (Vs) = twice AT at a minimum output of 2.5 V. The function preferentially uses Vc and verifies capture after each paced beat. In case of loss of capture, a back‐up spike is delivered and Vs is implemented until the next threshold measurement. We estimated the ventricular amplitude delivered by the pacemaker from data stored in the pacemaker memory. We compared these values with the pacing amplitude typically programmed manually (MPA) by physicians at twice AT and a minimum of 2.5 V. Data from 57 recipients of Talent 3 DR pacemakers were analyzed. Complete data sets were available in 25 patients at 1 day, 28 at 1 month, and 39 between 1 day and 1 month. No loss of capture or ventricular pause was observed on 53 ambulatory electrocardiograms (ECG); and pulse amplitude automatically delivered by the device was significantly lower than the MPA at each of the three time points analyzed. This new beat‐by‐beat capture function allows a significant lowering of the pacing amplitude compared to manual settings, while preserving a 100% safety.
Title: Clinical Testing of a New Pacemaker Function to Monitor Ventricular Capture
Description:
Automatic beat‐by‐beat capture functions are designed to minimize the pacing energy delivered, while maintaining the highest safety by delivering an immediate back‐up stimulus in case of loss of capture.
The objective of this study was to estimate the lowering of ventricular pacing amplitude allowed by such a function, compared to amplitudes usually set manually in routine practice.
An automatic ventricular pacing threshold test is launched every 6 hours to measure the automatic capture threshold (AT).
From AT the function calculates: (1) the“capture amplitude”(Vc) = AT + 0.
5 V at a minimum output of 1 V and (2) the“safety amplitude” (Vs) = twice AT at a minimum output of 2.
5 V.
The function preferentially uses Vc and verifies capture after each paced beat.
In case of loss of capture, a back‐up spike is delivered and Vs is implemented until the next threshold measurement.
We estimated the ventricular amplitude delivered by the pacemaker from data stored in the pacemaker memory.
We compared these values with the pacing amplitude typically programmed manually (MPA) by physicians at twice AT and a minimum of 2.
5 V.
Data from 57 recipients of Talent 3 DR pacemakers were analyzed.
Complete data sets were available in 25 patients at 1 day, 28 at 1 month, and 39 between 1 day and 1 month.
No loss of capture or ventricular pause was observed on 53 ambulatory electrocardiograms (ECG); and pulse amplitude automatically delivered by the device was significantly lower than the MPA at each of the three time points analyzed.
This new beat‐by‐beat capture function allows a significant lowering of the pacing amplitude compared to manual settings, while preserving a 100% safety.
Related Results
Patients’ Perspective on Termination of Pacemaker Therapy—A Cross-Sectional Anonymous Survey among Patients Carrying a Pacemaker in Germany
Patients’ Perspective on Termination of Pacemaker Therapy—A Cross-Sectional Anonymous Survey among Patients Carrying a Pacemaker in Germany
Objective: To determine the opinions of patients regarding the withdrawal of pacemaker therapy. Participants and methods: A cross-sectional anonymous questionnaire was administered...
Device–device interference triggered by an abandoned pacemaker: a case report
Device–device interference triggered by an abandoned pacemaker: a case report
Abstract
Background
Cardiac implantable electronic devices (CIEDs) are prone to electromagnetic interference. Common sources inc...
Vagus nerve stimulation reduces ventricular arrhythmias and increases ventricular electrical stability
Vagus nerve stimulation reduces ventricular arrhythmias and increases ventricular electrical stability
AbstractBackgroundTranscutaneous stimulation of the auricular branch of the vagus nerve (AB‐VNS) is a potentially noninvasive, inexpensive, and safe approach for vagus nerve stimul...
Long-term implications of pacemaker insertion in younger adults: a single centre experience
Long-term implications of pacemaker insertion in younger adults: a single centre experience
Abstract
Background
The long-term implications of pacemaker insertion in younger adults are poorly described in the literature.
...
The importance of the Autostrain RV technique in the treatment of right ventricular myocardial alterations in patients with breast cancer receiving chemotherapy
The importance of the Autostrain RV technique in the treatment of right ventricular myocardial alterations in patients with breast cancer receiving chemotherapy
Abstract
BACKGROUND
To research the value of Autostrain right ventricular(RV) technology in detecting and preventing right ventricular myocardial injury in patients underg...
Subacute Right Ventricle Perforation by Pacemaker Lead Presenting with Left Hemothorax and Shock
Subacute Right Ventricle Perforation by Pacemaker Lead Presenting with Left Hemothorax and Shock
Cardiac perforation by pacemaker is a rare but potentially fatal complication. Acute perforations occurring within twenty-four hours of insertion of pacemaker can lead to hemoperic...
Leadless Pacemaker tine fracture and dislocation: two case reports
Leadless Pacemaker tine fracture and dislocation: two case reports
Abstract
The rate of leadless pacemaker implantation is increasing worldwide. To date, there have been few reports of leadless pacemaker dislocation and extraction. This ar...
Aveir Leadless Pacemaker implantation in pediatric population: a case series
Aveir Leadless Pacemaker implantation in pediatric population: a case series
Abstract
Background
While the Medtronic Micra pacemaker provided a small device for leadless pacemaker implantation, the Aveir d...

