Analysis of Artifacts and Errors on Intraoral Phosphor Plate Radiographs: A Retrospective Study
View through CrossRef
Introduction. The advantages of phosphor plates (PPs), including their flexibility, thinness, and wireless connectivity, have contributed to their widespread use in dentistry alongside the advancement of digital radiography techniques. Continuous use of PPs and certain errors may lead to artifacts on the images. This retrospective study aimed to determine the frequency and types of intraoral PP artifacts.
Methods. This study was conducted on 814 intraoral PP radiographs, including periapical and bitewing images. The errors and artifacts were classified into 4 main categories: technical errors, plate-related artifacts, scanner-induced artifacts, and ambient light artifacts.
Results. A total of 656 periapical and 158 bitewing radiographs were examined. The most observed artifacts on the radiographs were plate-related artifacts, followed by technical errors (n=542, n=461, respectively). Within the category of plate-related artifacts, the most common artifacts were cracks or scratches (n=418, 77.1%), while within the category of technical errors, these were cone-cut errors (n=188, 40.7%). Parallel or zigzag radiopaque lines were the most common scanner-induced artifacts (n=313, 98.7%) and fading was the most common ambient light artefact (n=93, 49.2%). On more than half of the PPs, more than one artifact group was observed.
Conclusions. Cone-cut, cracks/scratches, parallel zigzag radiopaque lines, and fading were common PP artifacts and errors in this study. The analysis revealed that over half of the PPs exhibited appearance of more than one artifact group. The high frequency of artifacts poses a risk of incorrect and incomplete diagnoses in radiographs, emphasizing the crucial importance of understanding the causes and developing effective prevention strategies for artifacts to enhance the reliability of diagnostic imaging and ensure the accuracy of patient assessments.
Title: Analysis of Artifacts and Errors on Intraoral Phosphor Plate Radiographs: A Retrospective Study
Description:
Introduction.
The advantages of phosphor plates (PPs), including their flexibility, thinness, and wireless connectivity, have contributed to their widespread use in dentistry alongside the advancement of digital radiography techniques.
Continuous use of PPs and certain errors may lead to artifacts on the images.
This retrospective study aimed to determine the frequency and types of intraoral PP artifacts.
Methods.
This study was conducted on 814 intraoral PP radiographs, including periapical and bitewing images.
The errors and artifacts were classified into 4 main categories: technical errors, plate-related artifacts, scanner-induced artifacts, and ambient light artifacts.
Results.
A total of 656 periapical and 158 bitewing radiographs were examined.
The most observed artifacts on the radiographs were plate-related artifacts, followed by technical errors (n=542, n=461, respectively).
Within the category of plate-related artifacts, the most common artifacts were cracks or scratches (n=418, 77.
1%), while within the category of technical errors, these were cone-cut errors (n=188, 40.
7%).
Parallel or zigzag radiopaque lines were the most common scanner-induced artifacts (n=313, 98.
7%) and fading was the most common ambient light artefact (n=93, 49.
2%).
On more than half of the PPs, more than one artifact group was observed.
Conclusions.
Cone-cut, cracks/scratches, parallel zigzag radiopaque lines, and fading were common PP artifacts and errors in this study.
The analysis revealed that over half of the PPs exhibited appearance of more than one artifact group.
The high frequency of artifacts poses a risk of incorrect and incomplete diagnoses in radiographs, emphasizing the crucial importance of understanding the causes and developing effective prevention strategies for artifacts to enhance the reliability of diagnostic imaging and ensure the accuracy of patient assessments.
Related Results
NICU Medication Errors: Describing the Cause and Nature of Medication Errors in a NICU in Qatar
NICU Medication Errors: Describing the Cause and Nature of Medication Errors in a NICU in Qatar
IntroductionA medication error can be defined as “any error occurring in the medication use process” and focuses on problems with the delivery of medication to a patient [1]. Medic...
Screening of color phosphor powders on CRT faceplates
Screening of color phosphor powders on CRT faceplates
Abstract— We studied the screening of color phosphor powders on CRT faceplates by photolithography. For photolithography, the phosphor particles should disperse perfectly in the dr...
Retrospective Study: Evaluating the Positioning Errors in Digital Panoramic Radiographs
Retrospective Study: Evaluating the Positioning Errors in Digital Panoramic Radiographs
To produce a panoramic cardiograph with a constant image quality common errors must be kept to a minimum.This study is conducted with an objective to analyse the most prevalent pos...
The influences of LaVO4:Eu3+,Cr3+ red phosphors on white light-emitting diode applications
The influences of LaVO4:Eu3+,Cr3+ red phosphors on white light-emitting diode applications
<span>In this present paper, the </span><span lang="EN-ID">LaVO<sub>4</sub>:Eu<sup>3+</sup>,Cr<sup>3+</sup> (</span><...
An Efficient Broadband Cyan Phosphor for Full‐Spectrum pc‐WLEDs
An Efficient Broadband Cyan Phosphor for Full‐Spectrum pc‐WLEDs
AbstractCurrent phosphor‐converted white light‐emitting diodes (WLEDs) exhibit an unexpected gap in the cyan spectral range (480–520 nm), which hinders the realization of high colo...
Assessment of Common Errors and Subjective Quality of Digital Panoramic Radiographs in Dental Institution, Riyadh
Assessment of Common Errors and Subjective Quality of Digital Panoramic Radiographs in Dental Institution, Riyadh
Aim:
The study aims to determine the frequency of positioning errors, record the number of diagnostically inappropriate images, and assess the quality of panoramic radi...
Efficient bright green phosphor for white light-emitting diode using in backlight application
Efficient bright green phosphor for white light-emitting diode using in backlight application
The (Ba,Ca)ScO2F:Bi3+,K+ phosphor with narrow emission band and green emission color peaking at 510 nm is demonstrated in the study. This phosphor is prepared using a solid-state ...
Awareness of intraoral photography among dental students
Awareness of intraoral photography among dental students
Precise shade matching is one of the most testing parts of dental restorations and stylish dentistry. Because of the incredible assortment of regular tooth shading accomplishing a ...