Standard views do not suffice in assessing distal scaphoid articular cannulated screw penetration
View through CrossRef
Abstract
Background
Articular screw penetration is one of the most common hardware-related problems after scaphoid fracture fixation, occurring in up to two-thirds of patients, in particular into the scaphotrapezotrapezoidal (STT) joint. The aim of this study was to investigate whether this clinically important issue could be detected using standard anteroposterior (AP) and lateral, as well as additional nonstandard fluoroscopic views using direct open visualization with magnifying loupes as reference standard.
Materials and methods
Ten fresh cadaver wrists were used for this imaging study. A 2.2 mm cannulated compression screws with a length of 24 mm was placed in the scaphoid and incrementally left to protrude at the STT joint up to 2 mm. Eight fluoroscopic views of the wrist were then obtained by rotating the forearm using goniometric measurements, keeping the image beam parallel to the floor: (1) anteroposterior with the wrist in neutral rotation, (2) anteroposterior with the wrist in ulnar deviation, (3) supinated oblique 60° from neutral (60° supinated oblique), (4) supinated oblique 45° from neutral (45° supinated oblique), (5) a true lateral, (6) a true lateral with the wrist in radial deviation, (7) pronated oblique 45° from neutral (45° pronated oblique), and (8) a pronated oblique 60° from neutral (60° pronated oblique).
Results
Standard anteroposterior and lateral fluoroscopy views (radiographically calibrated) of a percutaneous cannulated screw fixation of a scaphoid fracture were insufficient to detect distal articular penetration, missing half the amount of screw penetrations in the current study. The 45° pronated oblique view was found as the most sensitive in detecting STT penetration (p < 0.0001).
Conclusions
Standard anteroposterior and lateral fluoroscopy views of a percutaneous cannulated screw fixation of a scaphoid waist fracture are insufficient to detect STT screw penetration. According to the current study, standard views would have missed half the amount of screw penetrations, which seems to reflect the high incidence of this problem in current practice. The most sensitive view was the 45° pronated oblique view, which detected STT screw penetration in all cases.
Level of Evidence Not applicable.
Springer Science and Business Media LLC
Title: Standard views do not suffice in assessing distal scaphoid articular cannulated screw penetration
Description:
Abstract
Background
Articular screw penetration is one of the most common hardware-related problems after scaphoid fracture fixation, occurring in up to two-thirds of patients, in particular into the scaphotrapezotrapezoidal (STT) joint.
The aim of this study was to investigate whether this clinically important issue could be detected using standard anteroposterior (AP) and lateral, as well as additional nonstandard fluoroscopic views using direct open visualization with magnifying loupes as reference standard.
Materials and methods
Ten fresh cadaver wrists were used for this imaging study.
A 2.
2 mm cannulated compression screws with a length of 24 mm was placed in the scaphoid and incrementally left to protrude at the STT joint up to 2 mm.
Eight fluoroscopic views of the wrist were then obtained by rotating the forearm using goniometric measurements, keeping the image beam parallel to the floor: (1) anteroposterior with the wrist in neutral rotation, (2) anteroposterior with the wrist in ulnar deviation, (3) supinated oblique 60° from neutral (60° supinated oblique), (4) supinated oblique 45° from neutral (45° supinated oblique), (5) a true lateral, (6) a true lateral with the wrist in radial deviation, (7) pronated oblique 45° from neutral (45° pronated oblique), and (8) a pronated oblique 60° from neutral (60° pronated oblique).
Results
Standard anteroposterior and lateral fluoroscopy views (radiographically calibrated) of a percutaneous cannulated screw fixation of a scaphoid fracture were insufficient to detect distal articular penetration, missing half the amount of screw penetrations in the current study.
The 45° pronated oblique view was found as the most sensitive in detecting STT penetration (p < 0.
0001).
Conclusions
Standard anteroposterior and lateral fluoroscopy views of a percutaneous cannulated screw fixation of a scaphoid waist fracture are insufficient to detect STT screw penetration.
According to the current study, standard views would have missed half the amount of screw penetrations, which seems to reflect the high incidence of this problem in current practice.
The most sensitive view was the 45° pronated oblique view, which detected STT screw penetration in all cases.
Level of Evidence Not applicable.
Related Results
Percutaneous Fixation for Scaphoid Nonunion With Bone Grafting Through the Distal Insertion Hole of a Fully Threaded Headless Screw
Percutaneous Fixation for Scaphoid Nonunion With Bone Grafting Through the Distal Insertion Hole of a Fully Threaded Headless Screw
Objective: Percutaneous scaphoid screw fixation is a popular treatment for acute scaphoid fractures with no or minimal displacement. For treating scaphoid nonunions, however, open ...
Scaphoid nonunion and SNAC treatment
Scaphoid nonunion and SNAC treatment
Introduction Scaphoid nonunion can result in progressive scaphoid nonunion advanced collapse (SNAC) and have an impact on the quality of life in younger patients. The social signi...
Diagnosis and Treatment of Scaphoid Waist Fractures: A Literature Review
Diagnosis and Treatment of Scaphoid Waist Fractures: A Literature Review
Scaphoid fracture can cause serious complications and its diagnosis and treatment approaches are still contentious. Tenderness of anatomical snuffbox (ASB), longitudinal compressio...
Osseous microarchitecture of the scaphoid: Cadaveric study of regional variations and clinical implications
Osseous microarchitecture of the scaphoid: Cadaveric study of regional variations and clinical implications
AbstractBone strength and structure are closely associated with fracture and screw fixation, however osseous micro architecture on scaphoid has not been clearly addressed. We condu...
Percutaneous Screw Fixation for Acute Scaphoid Fractures Through K-wire-assisted Reduction and Maintenance
Percutaneous Screw Fixation for Acute Scaphoid Fractures Through K-wire-assisted Reduction and Maintenance
Abstract
Background: Percutaneous screw fixation was introduced for acute scaphoid fractures through K-wire-assisted reduction and maintenance, and the effectiveness of the...
Volar Barton Fracture Treatment with T Plate Using Distal End Screws and Without Distal End Screws; A Comparative Study
Volar Barton Fracture Treatment with T Plate Using Distal End Screws and Without Distal End Screws; A Comparative Study
Background: Volar Barton's fractures are the fracture of distal radius with intra-articular extension associated withdislocation and subluxation of radiocarpal joint requiring surg...
PANORAMIC REVIEW OF ACUTE SCAPHOID FRACTURES: DESCRIPTION, EPIDEMIOLOGY, ETIOLOGY, PRESENTATION, DIAGNOSIS, TREATMENT AND PROGNOSIS
PANORAMIC REVIEW OF ACUTE SCAPHOID FRACTURES: DESCRIPTION, EPIDEMIOLOGY, ETIOLOGY, PRESENTATION, DIAGNOSIS, TREATMENT AND PROGNOSIS
Introduction: Scaphoid fractures are the most common fractures of the carpal bone and commonly occur in young males. Inadequate treatment of these fractures commonly leads to pseud...
Study on Positioning Accuracy of Aerostatic Lead Screw
Study on Positioning Accuracy of Aerostatic Lead Screw
The aerostatic lead screw is a non-contact type lead screw in which the contact surface of a male screw and a female screw is supported by a pressurized air film. It is characteriz...