Comparative Efficacy of Humidifiers for Noninvasive Infant Respiratory Support
View through CrossRef
Abstract
Delivery of cold, dry air to infants while supporting their breathing can lead to hypothermia in addition to irritating and damaging their sensitive nares and negatively impacting outcomes with these therapies. In high resource settings, electric heated humidifiers are used to mitigate this problem. In many resource-constrained settings, passive nonelectric bubbling humidifiers are instead used. We here compare the efficacy of conventional electric heated humidification, custom-built low cost heated humidification, passive nonelectric bubbling humidification and a control of no humidification. In a hospital patient room (temperature 22 °C, humidity 50%), the temperature and humidity delivered to a simulated patient lung via a BC161-10 Fisher Paykel bubble continuous positive airway pressure (CPAP) system were measured with conventional electric heated humidification, low cost custom-built heated humidification, passive bubbling humidification and no humidification. (Delivered CPAP: 5 cm H2O; flowrate varied from 4 to 8 liters per minute (LPM) in 2 LPM increments.) As the flowrate was varied from 4 to 8 LPM, delivered relative humidity (standard deviation) with each humidifier was as follows: control 10% (3.6%), passive bubbler 44% (3.7%), custom-built humidifier 67% (1.7%), electric heated humidifier 91% (0.86%). Delivered temperature with the electric heated humidifier was 38 °C (0.21 °C) versus 33 °C for all other setups. Conventional electric heating humidification is more effective than passive bubbling humidification, and the custom-built low cost humidifier provides an intermediate degree of humidification. Through further improvement of this concept with a heated inspiratory circuit and sensor based control of the heating element, an effective yet low cost solution heating humidification could be developed.
ASME International
Title: Comparative Efficacy of Humidifiers for Noninvasive Infant Respiratory Support
Description:
Abstract
Delivery of cold, dry air to infants while supporting their breathing can lead to hypothermia in addition to irritating and damaging their sensitive nares and negatively impacting outcomes with these therapies.
In high resource settings, electric heated humidifiers are used to mitigate this problem.
In many resource-constrained settings, passive nonelectric bubbling humidifiers are instead used.
We here compare the efficacy of conventional electric heated humidification, custom-built low cost heated humidification, passive nonelectric bubbling humidification and a control of no humidification.
In a hospital patient room (temperature 22 °C, humidity 50%), the temperature and humidity delivered to a simulated patient lung via a BC161-10 Fisher Paykel bubble continuous positive airway pressure (CPAP) system were measured with conventional electric heated humidification, low cost custom-built heated humidification, passive bubbling humidification and no humidification.
(Delivered CPAP: 5 cm H2O; flowrate varied from 4 to 8 liters per minute (LPM) in 2 LPM increments.
) As the flowrate was varied from 4 to 8 LPM, delivered relative humidity (standard deviation) with each humidifier was as follows: control 10% (3.
6%), passive bubbler 44% (3.
7%), custom-built humidifier 67% (1.
7%), electric heated humidifier 91% (0.
86%).
Delivered temperature with the electric heated humidifier was 38 °C (0.
21 °C) versus 33 °C for all other setups.
Conventional electric heating humidification is more effective than passive bubbling humidification, and the custom-built low cost humidifier provides an intermediate degree of humidification.
Through further improvement of this concept with a heated inspiratory circuit and sensor based control of the heating element, an effective yet low cost solution heating humidification could be developed.
Related Results
LB2306. Population Pharmacokinetic (PPK), Pharmacokinetic/Pharmacodynamic attainment (PTA), and Clinical Pharmacokinetic/Pharmacodynamic (PK/PD) Analyses for Sulbactam-Durlobactam (SUL-DUR) to Support Dose Selection for the Treatment of Acinetobacter baum
LB2306. Population Pharmacokinetic (PPK), Pharmacokinetic/Pharmacodynamic attainment (PTA), and Clinical Pharmacokinetic/Pharmacodynamic (PK/PD) Analyses for Sulbactam-Durlobactam (SUL-DUR) to Support Dose Selection for the Treatment of Acinetobacter baum
Abstract
Background
SUL-DUR is a β-lactam/β-lactamase inhibitor combination in development for the treatment of ABC infections, ...
619. Pharmacokinetic-Pharmacodynamic (PK-PD) Target Attainment Analyses to Support Epetraborole Dose Selection for the Treatment of Patients with Mycobacterium avium Complex (MAC) Lung Disease
619. Pharmacokinetic-Pharmacodynamic (PK-PD) Target Attainment Analyses to Support Epetraborole Dose Selection for the Treatment of Patients with Mycobacterium avium Complex (MAC) Lung Disease
Abstract
Background
Epetraborole (EBO) is an orally available, bacterial leucyl transfer RNA synthetase inhibitor that concentra...
593. Population Pharmacokinetic Model Development for Epetraborole and Mycobacterium avium Complex (MAC) Lung Disease Patients Using Data from Phase 1 and 2 Studies
593. Population Pharmacokinetic Model Development for Epetraborole and Mycobacterium avium Complex (MAC) Lung Disease Patients Using Data from Phase 1 and 2 Studies
Abstract
Background
Epetraborole (EBO), an orally available bacterial leucyl transfer RNA synthetase inhibitor with potent activ...
592. Impact of Elevated MIC Values on Echinocandin Pharmacokinetic-Pharmacodynamic (PK-PD) Candida glabrata Target Attainment (TA)
592. Impact of Elevated MIC Values on Echinocandin Pharmacokinetic-Pharmacodynamic (PK-PD) Candida glabrata Target Attainment (TA)
Abstract
Background
Given the increasing prevalence of non-albicans Candida species, including C. glabrata and C. auris, which h...
Primerjalna književnost na prelomu tisočletja
Primerjalna književnost na prelomu tisočletja
In a comprehensive and at times critical manner, this volume seeks to shed light on the development of events in Western (i.e., European and North American) comparative literature ...
Impact of Humidification Systems on Ventilator-associated Pneumonia
Impact of Humidification Systems on Ventilator-associated Pneumonia
Abstract
Rationale and objectives
The respective influence on the incidence of ventilator-associated pneumonia of current...
Pilarowski–Bjornsson Syndrome with Congenital Heart Defect: A Case Report and Literature Review
Pilarowski–Bjornsson Syndrome with Congenital Heart Defect: A Case Report and Literature Review
Abstract
Introduction
Pilarowski–Bjornsson syndrome (PILBOS) is a rare autosomal dominant neurodevelopmental disorder caused by heterozygous variants in chromodomain helicase DNA-b...
Clinical Efficacy of Infantile Massage in the Treatment of Infant Functional Constipation: A Meta-Analysis
Clinical Efficacy of Infantile Massage in the Treatment of Infant Functional Constipation: A Meta-Analysis
Background:Functional constipation in children is a common disease that causes a psychological burden on infants and young children across the world. It will greatly affect infant ...