Contrast enhancement in abdominal computed tomography: influence of photon energy of different scanners
View through CrossRef
Objective:
Different CT scanners have different X-ray spectra and photon energies indicating that contrast enhancement vary among scanners. However, this issue has not been fully validated; therefore, we performed phantom and clinical studies to assess this difference.
Methods:
Two scanners were used: scanner-A and scanner-B. In the phantom study, we compared the contrast enhancement between the scanners at tube voltage peaks of 80, 100 and 120 kVp. Then, we calculated the effective energies of the two CT scanners. In the clinical study, 40 patients underwent abdominal scanning with scanner-A and another 40 patients with scanner-B, with each group using the same scanning protocol. The contrast enhancement of abdominal organs was assessed quantitatively (based on the absolute difference between the attenuation of unenhanced scans and contrast-enhanced scans) and qualitatively. A two-tailed independent Student's t-test and or the Mann–Whitney U test were used to compare the discrepancies.
Results:
In the phantom study, contrast enhancement for scanner-B was 36.9, 32.6 and 30.8% higher than that for scanner-A at 80, 100 and 120 kVp, respectively. The effective energies were higher for scanner-A than for scanner-B. In the quantitative analysis for the clinical study, scanner-B yielded significantly better contrast enhancement of the hepatic parenchyma, pancreas, kidney, portal vein and inferior vena cava compared with that of scanner-A. The mean visual scores for contrast enhancement were also significantly higher on images obtained by scanner-B than those by scanner-A.
Conclusion:
There were significant differences in contrast enhancement of the abdominal organs between the compared CT scanners from two different vendors even at the same scanning and contrast parameters.
Advances in knowledge:
Awareness of the impact of different X-ray energies on the resultant attenuation of contrast material is important when interpreting clinical CT images.
Oxford University Press (OUP)
Title: Contrast enhancement in abdominal computed tomography: influence of photon energy of different scanners
Description:
Objective:
Different CT scanners have different X-ray spectra and photon energies indicating that contrast enhancement vary among scanners.
However, this issue has not been fully validated; therefore, we performed phantom and clinical studies to assess this difference.
Methods:
Two scanners were used: scanner-A and scanner-B.
In the phantom study, we compared the contrast enhancement between the scanners at tube voltage peaks of 80, 100 and 120 kVp.
Then, we calculated the effective energies of the two CT scanners.
In the clinical study, 40 patients underwent abdominal scanning with scanner-A and another 40 patients with scanner-B, with each group using the same scanning protocol.
The contrast enhancement of abdominal organs was assessed quantitatively (based on the absolute difference between the attenuation of unenhanced scans and contrast-enhanced scans) and qualitatively.
A two-tailed independent Student's t-test and or the Mann–Whitney U test were used to compare the discrepancies.
Results:
In the phantom study, contrast enhancement for scanner-B was 36.
9, 32.
6 and 30.
8% higher than that for scanner-A at 80, 100 and 120 kVp, respectively.
The effective energies were higher for scanner-A than for scanner-B.
In the quantitative analysis for the clinical study, scanner-B yielded significantly better contrast enhancement of the hepatic parenchyma, pancreas, kidney, portal vein and inferior vena cava compared with that of scanner-A.
The mean visual scores for contrast enhancement were also significantly higher on images obtained by scanner-B than those by scanner-A.
Conclusion:
There were significant differences in contrast enhancement of the abdominal organs between the compared CT scanners from two different vendors even at the same scanning and contrast parameters.
Advances in knowledge:
Awareness of the impact of different X-ray energies on the resultant attenuation of contrast material is important when interpreting clinical CT images.
Related Results
Hydatid Disease of The Brain Parenchyma: A Systematic Review
Hydatid Disease of The Brain Parenchyma: A Systematic Review
Abstarct
Introduction
Isolated brain hydatid disease (BHD) is an extremely rare form of echinococcosis. A prompt and timely diagnosis is a crucial step in disease management. This ...
Atypical Sigmoid Volvulus in an Adolescent: A Case Report and Literature Review
Atypical Sigmoid Volvulus in an Adolescent: A Case Report and Literature Review
Abstract
Introduction
Sigmoid volvulus is an underrecognized diagnosis with potentially fatal outcomes in adolescents; the current study aims to present an adolescent with mild sym...
[RETRACTED] Rhino XL Male Enhancement v1
[RETRACTED] Rhino XL Male Enhancement v1
[RETRACTED]Rhino XL Reviews, NY USA: Studies show that testosterone levels in males decrease constantly with growing age. There are also many other problems that males face due ...
Emerging Evidence of IgG4-Related Disease in Pericarditis: A Systematic Review
Emerging Evidence of IgG4-Related Disease in Pericarditis: A Systematic Review
Abstract
Introduction
Immunoglobulin G4-related disease (IgG4-RD) is a recently identified immune-mediated condition that is debilitating and often overlooked. While IgG4-RD has be...
The photon blockade effect of a complete Buck-Sukumar model
The photon blockade effect of a complete Buck-Sukumar model
The Buck-Sukumar (BS) model, with a nonlinear coupling between the atom and the light field, is well defined only when its coupling strength is lower than a critical coupling. Its ...
ACCURACY AND PRECISION OF SIX DIGITAL SCANNERS IN FULL ARCH MAXILLARY SCANS: AN IN VITRO COMPARATIVE STUDY
ACCURACY AND PRECISION OF SIX DIGITAL SCANNERS IN FULL ARCH MAXILLARY SCANS: AN IN VITRO COMPARATIVE STUDY
The objective of this study is to evaluate the accuracy and precision of full arch maxillary digital scans produced by six digital scanners. Materials and Methods: A standard model...
PLACE DE L’IMAGERIE DANS LE DIAGNOSTIC DES DOULEURS ABDOMINALES AIGUËS DE L’ADULTE AUX URGENCES CHIRURGICALES DE L’HOPITAL ARISTIDE LE DANTEC
PLACE DE L’IMAGERIE DANS LE DIAGNOSTIC DES DOULEURS ABDOMINALES AIGUËS DE L’ADULTE AUX URGENCES CHIRURGICALES DE L’HOPITAL ARISTIDE LE DANTEC
Aim : To report the imaging strategies in the diagnosis of acute non-traumatic acute abdominal pain in adult patient. Patients and methods: We performed aforward-looking study over...
Feasibility of energy-resolved angiography
Feasibility of energy-resolved angiography
<p dir="ltr">Cardiovascular disease (CVD) is the leading cause of death globally, with coronary heart disease being one of the most common types of CVD. Two-dimensional x-ray...