Alternate Site Testing for HbA1c Using the Primus CLC330 GHb Analyzer

View through CrossRef
OBJECTIVE To determine whether the Primus high-pressure liquid chromatography (HPLC) is suited to alternate site testing (AST) for HbA1c in a hospital diabetes outpatient clinic. RESEARCH DESIGN AND METHODS Patients were attending the clinic for routine management of their diabetes. A number of diabetic patients with uremia (n = 11) were also investigated. HbA1c levels were measured in the outpatient setting by the Primus HPLC and in a more limited study the DCA-2000 instrument using the new 6-min assay cartridge. HbA1c measurements were also performed with Pierce affinity minicolumns and a Bio-Rad Variant HPLC. RESULTS The Primus HPLC assay had low imprecision of 2.3, 1.6, and 1.0% for HbA1c levels of 4.7, 7.3, and 11.1%, respectively, and was not prone to interference by carbamylated hemoglobin as found for the ion-exchange Variant HPLC method. Method comparison studies showed that the bias and proportional error between the Pierce affinity minicolumn procedure (standardized with respect to an external quality control program) and the Primus HPLC (Y) was −0.4 and 1.2% respectively (n = 32). Similarly the bias and proportional error between the Primus and DCA-2000 methods was 0.7 and −2.5%. The Primus was shown to give falsely elevated HbA1c concentrations if the time between sequential injections was > 28 min. CONCLUSIONS The Primus HPLC has a decided advantage over specialty AST instruments, like the DCA-2000, in not only meeting AST requirements but also allowing rapid automated batch processing of all laboratory HbA1c samples.
Title: Alternate Site Testing for HbA1c Using the Primus CLC330 GHb Analyzer
Description:
OBJECTIVE To determine whether the Primus high-pressure liquid chromatography (HPLC) is suited to alternate site testing (AST) for HbA1c in a hospital diabetes outpatient clinic.
RESEARCH DESIGN AND METHODS Patients were attending the clinic for routine management of their diabetes.
A number of diabetic patients with uremia (n = 11) were also investigated.
HbA1c levels were measured in the outpatient setting by the Primus HPLC and in a more limited study the DCA-2000 instrument using the new 6-min assay cartridge.
HbA1c measurements were also performed with Pierce affinity minicolumns and a Bio-Rad Variant HPLC.
RESULTS The Primus HPLC assay had low imprecision of 2.
3, 1.
6, and 1.
0% for HbA1c levels of 4.
7, 7.
3, and 11.
1%, respectively, and was not prone to interference by carbamylated hemoglobin as found for the ion-exchange Variant HPLC method.
Method comparison studies showed that the bias and proportional error between the Pierce affinity minicolumn procedure (standardized with respect to an external quality control program) and the Primus HPLC (Y) was −0.
4 and 1.
2% respectively (n = 32).
Similarly the bias and proportional error between the Primus and DCA-2000 methods was 0.
7 and −2.
5%.
The Primus was shown to give falsely elevated HbA1c concentrations if the time between sequential injections was > 28 min.
CONCLUSIONS The Primus HPLC has a decided advantage over specialty AST instruments, like the DCA-2000, in not only meeting AST requirements but also allowing rapid automated batch processing of all laboratory HbA1c samples.

Related Results

Microchemical Identification of Gamma-Hydroxybutyrate (GHB)
Microchemical Identification of Gamma-Hydroxybutyrate (GHB)
Abstract A new microcrystal test for the detection of gamma-hydroxybutyrate (GHB) is described. The silver/copper reagent consists of an aqueous solution of 0.1 g of...
Pharmacological Treatment of GHB Withdrawal Syndrome
Pharmacological Treatment of GHB Withdrawal Syndrome
Abstract Purpose of Review Gamma-hydroxybutyrate (GHB) is an illicit drug used for many reasons: during music festivals or parties, for self-managem...
Hubungan Kadar HbA1c dengan Komplikasi Retinopati Diabetik di RSUD Al-Ihsan
Hubungan Kadar HbA1c dengan Komplikasi Retinopati Diabetik di RSUD Al-Ihsan
Abstract. Diabetic retinopathy is a microvascular complication of diabetes mellitus that may lead to visual impairment and blindness. Hemoglobin A1c (HbA1c) is commonly used as an ...
Inverted U-Shaped Relationship between HbA1c and Diabetic Retinopathy in Diabetic Patients: A Cross-Sectional Study
Inverted U-Shaped Relationship between HbA1c and Diabetic Retinopathy in Diabetic Patients: A Cross-Sectional Study
Abstract Background: Diabetic retinopathy (DR) is a leading cause of blindness among adults with diabetes. Glycated hemoglobin A1C (HbA1C) is a critical biomarker for long-...
Which combination is best for HbA1c, fasting plasma glucose, and 2-h post-challenge glucose for diagnosis of diabetes mellitus
Which combination is best for HbA1c, fasting plasma glucose, and 2-h post-challenge glucose for diagnosis of diabetes mellitus
Abstract Background We evaluated reliability among glycated hemoglobin (HbA1c), fasting plasma glucose (FPG), and a 75 g oral glucose tolerance test (75 g OGTT) where 2-h ...
Inpatient management of gamma-hydroxybutyrate withdrawal
Inpatient management of gamma-hydroxybutyrate withdrawal
Objectives: To describe the baseline characteristics, treatment and retention in patients electively admitted for gamma-hydroxybutyrate (GHB) withdrawal management. Methods: All pa...

Back to Top