Abstract 9064: The Increase of Tissue Echogenicity as the Pre-State of Steam Pops During Radiofrequency Catheter Ablation

View through CrossRef
Introduction: Steam pops (SPs) during radiofrequency catheter ablation (RFA) using force sensing catheters are infrequent but represent a potential severe complication such as embolic stroke, cardiac perforation. SPs remains unpredictable, and unclear in detailed phenomenon. Methods: In an ex vivo, we set the bovine myocardium to 36 °C in a constant temperature bath and performed an experiment in a water tank that generated a non-pulsatile flow. RFA using irrigation catheter was applied at catheter-tissue contact force (CF) 5, 10, 20g, 30g, 40g, and 50g for 30sec. The increase of tissue echogenicity (increased echo brightness) observed by intracardiac echocardiographic images using ViewFlex™ Xtra and impedance during RFA were evaluated with and without SPs. Results: SPs phenomenon appeared in 48 out of 137 sites (35%), and the increased echo brightness during RFA was observed in 89 lesions (65%). All lesions with SPs showed increased echo brightness before SPs (6.4±6.1sec before SPs). Sites with SPs showed stronger CF (30.8±14.8g vs. 22.9±15.9g, p=0.0047), higher initial impedance (78±12Ω vs. 73±9Ω, p=0.0063), greater impedance drops in 5 seconds after the beginning of RFA (13±7Ω, vs. 6±5Ω, p<0.0001), and earlier appearance of increased echo brightness (9.7±6.2 vs 17.4±6.3 sec, p<0.0001). Among 89 lesions with increased echo brightness, sites with SPs showed greater impedance drops in 5 seconds after the beginning of RFA (13±7Ω, vs. 7±6Ω, P<0.0001). In ROC curve analysis of the impedance drops in 5 seconds after the beginning of RFA (AUC, 0.81; sensitivity, 85%; specificity, 71%), optimal cut-off value of SPs was 7.6Ω, and SPs were not observed in lesions with the impedance drops of 3Ω or less. (Figure) Conclusions: The increase of tissue echogenicity observed by ICE images show the pre-state of all steam pops during RFA. Steam pops could be avoided by checking the impedance drop at the initial stage of RFA under ICE monitoring.
Title: Abstract 9064: The Increase of Tissue Echogenicity as the Pre-State of Steam Pops During Radiofrequency Catheter Ablation
Description:
Introduction: Steam pops (SPs) during radiofrequency catheter ablation (RFA) using force sensing catheters are infrequent but represent a potential severe complication such as embolic stroke, cardiac perforation.
SPs remains unpredictable, and unclear in detailed phenomenon.
Methods: In an ex vivo, we set the bovine myocardium to 36 °C in a constant temperature bath and performed an experiment in a water tank that generated a non-pulsatile flow.
RFA using irrigation catheter was applied at catheter-tissue contact force (CF) 5, 10, 20g, 30g, 40g, and 50g for 30sec.
The increase of tissue echogenicity (increased echo brightness) observed by intracardiac echocardiographic images using ViewFlex™ Xtra and impedance during RFA were evaluated with and without SPs.
Results: SPs phenomenon appeared in 48 out of 137 sites (35%), and the increased echo brightness during RFA was observed in 89 lesions (65%).
All lesions with SPs showed increased echo brightness before SPs (6.
4±6.
1sec before SPs).
Sites with SPs showed stronger CF (30.
8±14.
8g vs.
22.
9±15.
9g, p=0.
0047), higher initial impedance (78±12Ω vs.
73±9Ω, p=0.
0063), greater impedance drops in 5 seconds after the beginning of RFA (13±7Ω, vs.
6±5Ω, p<0.
0001), and earlier appearance of increased echo brightness (9.
7±6.
2 vs 17.
4±6.
3 sec, p<0.
0001).
Among 89 lesions with increased echo brightness, sites with SPs showed greater impedance drops in 5 seconds after the beginning of RFA (13±7Ω, vs.
7±6Ω, P<0.
0001).
In ROC curve analysis of the impedance drops in 5 seconds after the beginning of RFA (AUC, 0.
81; sensitivity, 85%; specificity, 71%), optimal cut-off value of SPs was 7.
6Ω, and SPs were not observed in lesions with the impedance drops of 3Ω or less.
(Figure) Conclusions: The increase of tissue echogenicity observed by ICE images show the pre-state of all steam pops during RFA.
Steam pops could be avoided by checking the impedance drop at the initial stage of RFA under ICE monitoring.

Related Results

Remobilització dels contaminants orgànics persistents en els ecosistemes polars costaners
Remobilització dels contaminants orgànics persistents en els ecosistemes polars costaners
Persistent organic pollutants (POPs) constitute an heterogeneous group of chemical substances of great environmental relevance. Its persistence, bioaccumulation, long-range transpo...
Echogenicity unit: A mathematical formula based on grayscale imagery
Echogenicity unit: A mathematical formula based on grayscale imagery
Abstract Ultrasonography is a non-invasive imaging technique widely used in medical practice for diagnostic purposes. One crucial aspect of ultrasound imaging is the interp...
Steam ablation needles based on porous radiation clusters
Steam ablation needles based on porous radiation clusters
Steam thermal ablation (STA) is an emerging minimally invasive treatment technique that utilizes the energy of high-temperature steam to treat diseased tissues. Compared to traditi...
The advantage of mini electrode-equipped catheter for the radiofrequency ablation of paroxysmal supraventricular tachycardia
The advantage of mini electrode-equipped catheter for the radiofrequency ablation of paroxysmal supraventricular tachycardia
Introduction: Novel ablation catheters equipped with mini-electrodes (ME) offer high resolution mapping for target tissue. This study aimed to evaluate the mappin...
TINBOP Cyclic Steam Injection Enhances Oil Recovery in Mature Steamfloods
TINBOP Cyclic Steam Injection Enhances Oil Recovery in Mature Steamfloods
Abstract In a previous paper (SPE/PS-CIM/CHOA 97850), we presented our findings on a novel cyclic steam injection method that we developed. The Top-Injection Bott...
Microwave Ablation with or Without Chemotherapy in Management of Non-Small Cell Lung Cancer: A Systematic Review
Microwave Ablation with or Without Chemotherapy in Management of Non-Small Cell Lung Cancer: A Systematic Review
Abstract Introduction  Microwave ablation (MWA) has emerged as a minimally invasive treatment for patients with inoperable non-small cell lung cancer (NSCLC). However, whether it i...

Back to Top