Activated Pyrolytic Carbon Tip Pacing Leads: An Alternative to Steroid‐Eluting Pacing Leads?
View through CrossRef
Pyrolytic carbon technology is known for its excellent mechanical properties and electrical conductivity; it is particularly biocompatible and does not require high production costs. The Sorin S100/4 lead is a ventricular passive fixation lead with a hemispherical electrode surface area of 4 mm2; the stimulating tip is made of a graphite core coated by a thin activated pyrolytic carbon layer. We evaluated the acute and medium‐term performance of the unipolar version of this lead in 65 patients. At implantation, pacing threshold (at 0.5 ms) was 0.26 ± 0.08 V; pacing impedance (at 5 V and 0.5 ms) was 537 ± 94 Ω and Ft wave amplitude was 15.0 ± 5.5 mV. No lead related complications (dislodgment, perforation, exit block, etc.) occurred in any patient; one patient presented with a wire fracture after 26 months, due to subclavian crush syndrome. Follow‐up procedures were performed at 1 week, and 2‐, 10‐, 18‐, and 30‐months postimplant. Since the leads were connected to pacemakers from different manufacturers, either voltage or duration thresholds were measured, In approximately two thirds of the patients, with an output of 2.5 V, a mean duration threshold of 0.16 ± 0.13 ms at 1 week, 0.12 ± 0.08 ms at 2 months, 0.11 ± 0.06 ms at 10 months, 0.09 ± 0.06 ms at 18 months, and 0.07 ± 0.03 ms at 30 months, was measured. In the remaining one‐third of the patients, a comparable voltage threshold trend was measured. The mean pacing impedance showed a transient drop at 1 week, and then increased to a plateau of about 600 fl reached after 10 months. No sensing defect occurred in any patient. Our data show good acute and intermediate‐term results of the S100/4 lead; the early rise in threshold was remarkably blunted. Activated pyrolytic carbon tip leads might therefore be considered as a possible, inexpensive alternative to steroideluting leads.
Title: Activated Pyrolytic Carbon Tip Pacing Leads: An Alternative to Steroid‐Eluting Pacing Leads?
Description:
Pyrolytic carbon technology is known for its excellent mechanical properties and electrical conductivity; it is particularly biocompatible and does not require high production costs.
The Sorin S100/4 lead is a ventricular passive fixation lead with a hemispherical electrode surface area of 4 mm2; the stimulating tip is made of a graphite core coated by a thin activated pyrolytic carbon layer.
We evaluated the acute and medium‐term performance of the unipolar version of this lead in 65 patients.
At implantation, pacing threshold (at 0.
5 ms) was 0.
26 ± 0.
08 V; pacing impedance (at 5 V and 0.
5 ms) was 537 ± 94 Ω and Ft wave amplitude was 15.
0 ± 5.
5 mV.
No lead related complications (dislodgment, perforation, exit block, etc.
) occurred in any patient; one patient presented with a wire fracture after 26 months, due to subclavian crush syndrome.
Follow‐up procedures were performed at 1 week, and 2‐, 10‐, 18‐, and 30‐months postimplant.
Since the leads were connected to pacemakers from different manufacturers, either voltage or duration thresholds were measured, In approximately two thirds of the patients, with an output of 2.
5 V, a mean duration threshold of 0.
16 ± 0.
13 ms at 1 week, 0.
12 ± 0.
08 ms at 2 months, 0.
11 ± 0.
06 ms at 10 months, 0.
09 ± 0.
06 ms at 18 months, and 0.
07 ± 0.
03 ms at 30 months, was measured.
In the remaining one‐third of the patients, a comparable voltage threshold trend was measured.
The mean pacing impedance showed a transient drop at 1 week, and then increased to a plateau of about 600 fl reached after 10 months.
No sensing defect occurred in any patient.
Our data show good acute and intermediate‐term results of the S100/4 lead; the early rise in threshold was remarkably blunted.
Activated pyrolytic carbon tip leads might therefore be considered as a possible, inexpensive alternative to steroideluting leads.
Related Results
Crush Stenting With Paclitaxel-Eluting or Sirolimus-Eluting Stents for the Treatment of Coronary Bifurcation Lesions
Crush Stenting With Paclitaxel-Eluting or Sirolimus-Eluting Stents for the Treatment of Coronary Bifurcation Lesions
Two hundred forty-six patients with 252 bifurcation lesions were enrolled into a prospective, nonrandomized study to use paclitaxel-eluting or sirolimus-eluting stent for crush ste...
Comparative life cycle assessment of biomass-based and coal-based activated carbon production
Comparative life cycle assessment of biomass-based and coal-based activated carbon production
Activated carbon is an effective adsorbent due to its high porosity, large surface area and high surface reactivity. Activated carbon is commonly produced from coal which is a non-...
Long‐Term Follow‐up Impact of Dual‐Chamber Pacing on Patients with Hypertrophic Obstructive Cardiomyopathy
Long‐Term Follow‐up Impact of Dual‐Chamber Pacing on Patients with Hypertrophic Obstructive Cardiomyopathy
BackgroundPacing has been proposed as a treatment for patients with hypertrophic obstructive cardiomyopathy (HOCM), but there are few studies with long‐term follow‐up. We evaluated...
Recipient and Graft Outcomes in Simultaneous Kidney and Pancreas Transplantation With Steroid Avoidance in the United States
Recipient and Graft Outcomes in Simultaneous Kidney and Pancreas Transplantation With Steroid Avoidance in the United States
Background.
Steroid avoidance in kidney transplantation has been proven noninferior. Long-term outcome data on steroid avoidance in simultaneous pancreas-kidney (SPK) r...
Effects of PIP targeting LOX-1 eluting stents on in-stent restenosis and re-endothelialisation in rat abdominal aorta stenting models
Effects of PIP targeting LOX-1 eluting stents on in-stent restenosis and re-endothelialisation in rat abdominal aorta stenting models
Background and Objectives
Pyrrole-Imidazole polyamide (PIP) is a novel gene silencer that can be readily designed and synthesised to target any gene. Contrary to ...
Right ventricular pacing is not harmful as perceived previously
Right ventricular pacing is not harmful as perceived previously
Abstract
Background: Pacing from the apex of the RV is not optimal. This study aimed to assess and compare RV and LV pacing of LV function and dyssynchrony.
Thirty-six pati...
Childhood idiopathic nephrotic syndrome in Turkey
Childhood idiopathic nephrotic syndrome in Turkey
Abstract
Background
: It has been reported that there are racial and regional differences in peak incidence age, histopathological features and response to s...
ANALYSIS OF QUALITY OF ACTIVATED CARBON (Myristica fragrans) SHELL AS ADSORPTIVE AGENT
ANALYSIS OF QUALITY OF ACTIVATED CARBON (Myristica fragrans) SHELL AS ADSORPTIVE AGENT
The activated carbon made in this study is activated carbon in the form of powder and granules (a mixture of the two) by a chemical activation process using calcium oxide (CaO). Th...