A-345 Effect of Hemolysis on Routine Blood Gas Analytes
View through CrossRef
Abstract
Background
Hemolysis is a major pre-analytical concern for most laboratory analytes. Detection of hemolysis and mitigation efforts are especially important for analytes, such as potassium, with high intracellular concentrations. Routine serum and plasma chemistry tests are performed on analyzers with the capacity to detect and measure the degree of hemolysis. However, for whole blood chemistries and blood gas measurements, the instruments utilized lack this capacity. The aim of this study was to evaluate the effect of hemolysis on whole blood and blood gas analytes and to compare visual assessments of hemolysis to measured hemolysis indices (H-index).
Methods
Remnant whole blood samples were split into two portions and each portion was aspirated and dispensed through a syringe one or more times. For the mock portion, only the syringe was used, while the hemolyzed portion had a needle affixed to the end of the syringe. The needle provided shear stress on the red blood cells to induce hemolysis, while the mock procedure was used to assess the impact of aspirating/dispensing on the sample in the absence of hemolysis. Each portion was then analyzed on a Radiometer ABL800 series instrument, spun down, and the H-index of the plasma portion was measured on a Roche cobas 8000 instrument. Medical technologists recorded their visual assessment of the specimens, with two technologists agreeing to the categorization of the specimen as either slightly, moderately, or severely hemolyzed. Degree of hemolysis was categorized by the delta (hemolyzed - mock) of the measured H-index: slight hemolysis was defined as an H-index delta of <100, moderate as 100-500 and severe as >500.
Results
The effect of hemolysis, with H-indices ranging from 2 to 3861, on 13 routine blood gas analytes was studied for 85 whole blood specimens. Hemolysis had little effect on metabolites, as percent bias was within ±3% at all levels of hemolysis for glucose, creatinine, and lactate. Similarly, most cooximetry components were minimally affected, with total hemoglobin, oxyhemoglobin, carboxyhemoglobin, and oxygen saturation within ±5% bias at all levels of hemolysis. Methemoglobin had a larger overall negative bias, with slight, moderate, and severe hemolysis levels yielding percent biases of -6.6, -12.3, and -13.3%, respectively. As expected, potassium displayed a significant positive bias with increasing hemolysis, with a generally linear trend. At moderate levels of hemolysis, the average potassium bias was 24.0% and at severe levels, over 100%. Sodium and ionized calcium also displayed overall linear trends but with a negative bias. At slight, moderate, and severe levels of hemolysis, sodium had a -0.56, -1.10, and -3.96% bias, and ionized calcium had a -2.99, -5.65, and -15.5% bias respectively.
Conclusions
Hemolysis can falsely increase or decrease a range of blood gas analytes and lead to misinterpretation of results and adversely affect clinical decision-making. Therefore, equipping current blood gas analyzers with hemolysis detection capabilities is crucial to enable laboratories to mitigate this effect and ensure accurate results for patient care.
Title: A-345 Effect of Hemolysis on Routine Blood Gas Analytes
Description:
Abstract
Background
Hemolysis is a major pre-analytical concern for most laboratory analytes.
Detection of hemolysis and mitigation efforts are especially important for analytes, such as potassium, with high intracellular concentrations.
Routine serum and plasma chemistry tests are performed on analyzers with the capacity to detect and measure the degree of hemolysis.
However, for whole blood chemistries and blood gas measurements, the instruments utilized lack this capacity.
The aim of this study was to evaluate the effect of hemolysis on whole blood and blood gas analytes and to compare visual assessments of hemolysis to measured hemolysis indices (H-index).
Methods
Remnant whole blood samples were split into two portions and each portion was aspirated and dispensed through a syringe one or more times.
For the mock portion, only the syringe was used, while the hemolyzed portion had a needle affixed to the end of the syringe.
The needle provided shear stress on the red blood cells to induce hemolysis, while the mock procedure was used to assess the impact of aspirating/dispensing on the sample in the absence of hemolysis.
Each portion was then analyzed on a Radiometer ABL800 series instrument, spun down, and the H-index of the plasma portion was measured on a Roche cobas 8000 instrument.
Medical technologists recorded their visual assessment of the specimens, with two technologists agreeing to the categorization of the specimen as either slightly, moderately, or severely hemolyzed.
Degree of hemolysis was categorized by the delta (hemolyzed - mock) of the measured H-index: slight hemolysis was defined as an H-index delta of <100, moderate as 100-500 and severe as >500.
Results
The effect of hemolysis, with H-indices ranging from 2 to 3861, on 13 routine blood gas analytes was studied for 85 whole blood specimens.
Hemolysis had little effect on metabolites, as percent bias was within ±3% at all levels of hemolysis for glucose, creatinine, and lactate.
Similarly, most cooximetry components were minimally affected, with total hemoglobin, oxyhemoglobin, carboxyhemoglobin, and oxygen saturation within ±5% bias at all levels of hemolysis.
Methemoglobin had a larger overall negative bias, with slight, moderate, and severe hemolysis levels yielding percent biases of -6.
6, -12.
3, and -13.
3%, respectively.
As expected, potassium displayed a significant positive bias with increasing hemolysis, with a generally linear trend.
At moderate levels of hemolysis, the average potassium bias was 24.
0% and at severe levels, over 100%.
Sodium and ionized calcium also displayed overall linear trends but with a negative bias.
At slight, moderate, and severe levels of hemolysis, sodium had a -0.
56, -1.
10, and -3.
96% bias, and ionized calcium had a -2.
99, -5.
65, and -15.
5% bias respectively.
Conclusions
Hemolysis can falsely increase or decrease a range of blood gas analytes and lead to misinterpretation of results and adversely affect clinical decision-making.
Therefore, equipping current blood gas analyzers with hemolysis detection capabilities is crucial to enable laboratories to mitigate this effect and ensure accurate results for patient care.
Related Results
[RETRACTED] Guardian Blood Balance –Feel the difference Guardian Blood Balance makes! v1
[RETRACTED] Guardian Blood Balance –Feel the difference Guardian Blood Balance makes! v1
[RETRACTED]Guardian Blood Balance Reviews (Works Or Hoax) Does Guardian Botanicals Blood Balance AU Really Works? Read Updated Report! Diabetes and Hypertension is such a health p...
Ghost cells as a two-phase blood analog fluid —fluorescent mechanical hemolysis detection
Ghost cells as a two-phase blood analog fluid —fluorescent mechanical hemolysis detection
Abstract
Background
This study investigated fluorescent hemolysis detection as an optical method to detect local hemolysis in mechanical circulatory support systems, addres...
[RETRACTED] Guardian Blood Balance Australia- Reviews - Guardian Botanicals Blood Balance [AU] SCAM ALERT! Read Real Critical Reports.. Price in Australia v1
[RETRACTED] Guardian Blood Balance Australia- Reviews - Guardian Botanicals Blood Balance [AU] SCAM ALERT! Read Real Critical Reports.. Price in Australia v1
[RETRACTED]Guardian Blood Balance Australia Reviews - Diabetes and blood sugar are some of the common problems that are attacking so many adult individuals nowadays. Obesity is t...
Association between inflammatory and hemolysis markers in children with sickle cell disease
Association between inflammatory and hemolysis markers in children with sickle cell disease
Abstract
Background: Sickle cell disease (SCD) is a severe hemoglobin disorder characterized by recurrent vaso-occlusive...
Three in One: Systemic Lupus Erythematosus, HELLP Syndrome, and Antiphospholipid Syndrome: A Case Report and Literature Review
Three in One: Systemic Lupus Erythematosus, HELLP Syndrome, and Antiphospholipid Syndrome: A Case Report and Literature Review
Abstract
Introduction
Systemic lupus erythematosus (SLE) is a multisystem autoimmune disease commonly affecting women of reproductive age. Its overlap with HELLP syndrome (Hemolysi...
Design Optimization of Artificial Heart Blood Pumps Based on Hemolysis Testing
Design Optimization of Artificial Heart Blood Pumps Based on Hemolysis Testing
Abstract
A measurement of blood hemolysis level (damage of blood cells) during pumping is one of the most effective methods to determine an ideal design feature o...
Red cell serological characterization of autoimmune hemolytic anemia in systemic lupus erythematosus patients
Red cell serological characterization of autoimmune hemolytic anemia in systemic lupus erythematosus patients
Abstract:
INTRODUCTION:
Autoimmune hemolytic anemia (AIHA) is a clinical condition with increased red cell destruction or decreased red cell surv...
Inhibition of the Complement Alternative Pathway Attenuates Hemolysis and Preserves Renal Function in a Mouse Model of Sickle Cell Disease
Inhibition of the Complement Alternative Pathway Attenuates Hemolysis and Preserves Renal Function in a Mouse Model of Sickle Cell Disease
Introduction: the alternative pathway (AP) of complement activation plays a significant role in the pathophysiology of sickle cell disease (SCD), contributing to hemolysis and subs...