Smart watch for the detection of exercise-induced transient ST segment changes
View through CrossRef
Abstract
Background
Smartwatches are largely used and allowed to perform electrocardiogram one lead, usually a modified DI, used to diagnose atrial fibrillation. It is unknown whether or not smartwatches could be used to identify electrocardiographic transient ST changes during exercise-induced myocardial ischemia. The present study assessed the feasibility and reliability of using an Apple watch to record V5 in subjects scheduled for standard ECG stress tests.
Methods
One hundred eighty-nine subjects were included in this study. The mean age was 61+9 years, and 21.6 % were women (41 subjects). A single thoracic lead V5 was obtained with an Apple watch, positioned at the fifth intercostal space anterior axillary line, by creating a circuit between the back of the watch and the right index resting on the crown for 30 seconds. Four subjects were excluded because of the poor quality of the Apple watch signal. Forty-nine patients (25,92%) had a positive stress test with standard 12 lead ECG, 136 (71,96 %) negative stress test, and 4 (2,12%) the ECGs were non-diagnostic. The mean difference or bias between the Apple Watch and standard exercise-induced ST changes was quantified as 0.054, suggesting a generally minimal discrepancy in their readings. This consistency is critically affirmed by Cohen's Kappa (k) coefficient of 0.665. 9-1.0)
Results
A total of 189 subjects were included in the study; the mean age was 61 years (SD 9), and 21.6 % were women (41 subjects). Insightful outcomes were gleaned from the Bland-Altman analysis, which probed the concordance between the smartwatch ECG and the conventional standard ECG in measuring ST-segment deviations (Bias 0,054; SD 0.31, from -0.56 to 0.67, 95% limits of agreement), as vividly depicted in Figure. The dashed blue lines in Figure demarcate the 95% limits of agreement, which span from -0.56 to 0.67, offering a tangible visualization of the range within which most individual differences resided. A notable harmony was unveiled in the intricate exploration of the concordance between the two diagnostic methodologies, substantiated by an agreement of 87.6%. This demonstrates that, in most instances, the methodologies provided analogous results, exhibiting a commendable consistency between them in a substantial proportion of cases.
Assuming the results of standard ECGs as the reference values, Apple Watch ST deviation showed a sensitivity of 69% and specificity of 94 %. The positive predictive value was 0.80952381, and the negative predictive value was 0,8951049.
Conclusions
In conclusion, this study demonstrated that the Apple Watch could detect transient ST changes with a sensitivity of 69% and a specificity of 94 %. Further prospective studies should be performed to assess the potential role of detecting transient myocardial ischemia with a smartwatch in a general population at risk.
Title: Smart watch for the detection of exercise-induced transient ST segment changes
Description:
Abstract
Background
Smartwatches are largely used and allowed to perform electrocardiogram one lead, usually a modified DI, used to diagnose atrial fibrillation.
It is unknown whether or not smartwatches could be used to identify electrocardiographic transient ST changes during exercise-induced myocardial ischemia.
The present study assessed the feasibility and reliability of using an Apple watch to record V5 in subjects scheduled for standard ECG stress tests.
Methods
One hundred eighty-nine subjects were included in this study.
The mean age was 61+9 years, and 21.
6 % were women (41 subjects).
A single thoracic lead V5 was obtained with an Apple watch, positioned at the fifth intercostal space anterior axillary line, by creating a circuit between the back of the watch and the right index resting on the crown for 30 seconds.
Four subjects were excluded because of the poor quality of the Apple watch signal.
Forty-nine patients (25,92%) had a positive stress test with standard 12 lead ECG, 136 (71,96 %) negative stress test, and 4 (2,12%) the ECGs were non-diagnostic.
The mean difference or bias between the Apple Watch and standard exercise-induced ST changes was quantified as 0.
054, suggesting a generally minimal discrepancy in their readings.
This consistency is critically affirmed by Cohen's Kappa (k) coefficient of 0.
665.
9-1.
0)
Results
A total of 189 subjects were included in the study; the mean age was 61 years (SD 9), and 21.
6 % were women (41 subjects).
Insightful outcomes were gleaned from the Bland-Altman analysis, which probed the concordance between the smartwatch ECG and the conventional standard ECG in measuring ST-segment deviations (Bias 0,054; SD 0.
31, from -0.
56 to 0.
67, 95% limits of agreement), as vividly depicted in Figure.
The dashed blue lines in Figure demarcate the 95% limits of agreement, which span from -0.
56 to 0.
67, offering a tangible visualization of the range within which most individual differences resided.
A notable harmony was unveiled in the intricate exploration of the concordance between the two diagnostic methodologies, substantiated by an agreement of 87.
6%.
This demonstrates that, in most instances, the methodologies provided analogous results, exhibiting a commendable consistency between them in a substantial proportion of cases.
Assuming the results of standard ECGs as the reference values, Apple Watch ST deviation showed a sensitivity of 69% and specificity of 94 %.
The positive predictive value was 0.
80952381, and the negative predictive value was 0,8951049.
Conclusions
In conclusion, this study demonstrated that the Apple Watch could detect transient ST changes with a sensitivity of 69% and a specificity of 94 %.
Further prospective studies should be performed to assess the potential role of detecting transient myocardial ischemia with a smartwatch in a general population at risk.
Related Results
Studi Komparatif Efektivitas Mckenzie Exercises dan William Flexion Exercises pada Myogenic Low Back Pain
Studi Komparatif Efektivitas Mckenzie Exercises dan William Flexion Exercises pada Myogenic Low Back Pain
McKenzie and William Flexion exercises are physiotherapeutic interventions for LBPM (Low Back Myogenic Pain) to reduce pain and improve functional activity. This study aims to comp...
Perceived barriers and facilitators to workplace exercise participation
Perceived barriers and facilitators to workplace exercise participation
Purpose
Workplace exercise programmes have been shown to increase employee participation in physical activities and improve health and fitness in the short-term. However, the limit...
Reinventing Smart Water Management System through ICT and IoT Driven Solution for Smart Cities
Reinventing Smart Water Management System through ICT and IoT Driven Solution for Smart Cities
Purpose: Worldwide water scarcity is one of the major problems to deal with. Smart Cities also faces this challenging problem due to its ever-increasing population and limited sour...
Automatic Responsiveness Testing in Epilepsy with Wearable Technology: The ARTiE Watch
Automatic Responsiveness Testing in Epilepsy with Wearable Technology: The ARTiE Watch
AbstractObjectiveAn accurate evaluation of behavioral responsiveness during and after seizures in people with epilepsy is critical for accurate diagnosis and management. Current me...
Generative AI-Driven Smart Contract Optimization for Secure and Scalable Smart City Services
Generative AI-Driven Smart Contract Optimization for Secure and Scalable Smart City Services
Smart cities use advanced infrastructure and technology to improve the quality of life for their citizens. Collaborative services in smart cities are making the smart city ecosyste...
SEGMENTATION OF CRUSTAL VELOCITY STRUCTURE BENEATH THE SHANDONG‐JIANGSU‐ANHUI SEGMENT OF THE TANLU FAULT ZONE AND ADJACENT AREAS AND ITS GEOLOGICAL INTERPRETATIONS
SEGMENTATION OF CRUSTAL VELOCITY STRUCTURE BENEATH THE SHANDONG‐JIANGSU‐ANHUI SEGMENT OF THE TANLU FAULT ZONE AND ADJACENT AREAS AND ITS GEOLOGICAL INTERPRETATIONS
AbstractThe Yishu segment of the Tanlu fault zone is the seismogenic structure of the 1668 Tancheng M8½ earthquake, and is also the research focus of the potential strong earthquak...
Effect of one hour kick boxing exercise on lipid profile
Effect of one hour kick boxing exercise on lipid profile
Background: Physical inactivity and a sedentary lifestyle are contributory factors for increased risk of obesity, hypertension, hyperlipidemia, and related complications, namely ca...
Influence of pre‐exercise muscle glycogen content on exercise‐induced transcriptional regulation of metabolic genes
Influence of pre‐exercise muscle glycogen content on exercise‐induced transcriptional regulation of metabolic genes
Transcription of metabolic genes is transiently induced during recovery from exercise in skeletal muscle of humans. To determine whether pre‐exercise muscle glycogen content influe...