Real-time dynamic changes in intraocular pressure after occlusion break: comparing 2 phacoemulsification systems
View through CrossRef
Purpose:
To compare real-time intraocular pressure (IOP) response to occlusion break event in 2 phacoemulsification systems.
Setting:
Iladevi Cataract and IOL Research Center, India.
Design:
Randomized, experimental study.
Methods:
Rabbits were randomized to:Group I (n = 10 eyes), Centurion Vision system with active fluidics, or, Group II (n = 10 eyes), Centurion with Active Sentry. Within each group, parameter set 1 (IOP 30 mm Hg, aspiration flow rate [AFR] 20 mL/min, and vacuum 600 mm Hg) and parameter set 2 (IOP 50 mm Hg, AFR 25 mL/min, and vacuum 600 mm Hg) were tested. Real-time rate of drop and rise of IOP after occlusion break event (mm Hg per second) and percentage reduction of IOP from maximum during nuclear fragment removal were compared.
Results:
10 rabbits (20 eyes) were included. Rate of drop of IOP after occlusion break was not significantly different between groups. Rate of rise of IOP was statistically significantly higher in Group II with both parameter sets (199.09 ± 69.28 vs 94.33 ± 45.66 in parameter set 1, P = .006; and 256 ± 45.05 vs 165.25 ± 51.80 in parameter set 2, P = .005), suggesting faster recovery to baseline IOP after occlusion break. The mean percentage reduction of IOP from maximum was significantly higher in Group I (P = .003).
Conclusions:
Rise of IOP to baseline after occlusion break event was faster and mean percentage reduction of IOP from maximum during nuclear fragment removal was lower when using the Centurion Vision system with the Active Sentry upgrade compared with the traditional handpiece. The ability to sense IOP at the level of the handpiece with the Active Sentry upgrade allows faster mitigation of surge response.
Ovid Technologies (Wolters Kluwer Health)
Title: Real-time dynamic changes in intraocular pressure after occlusion break: comparing 2 phacoemulsification systems
Description:
Purpose:
To compare real-time intraocular pressure (IOP) response to occlusion break event in 2 phacoemulsification systems.
Setting:
Iladevi Cataract and IOL Research Center, India.
Design:
Randomized, experimental study.
Methods:
Rabbits were randomized to:Group I (n = 10 eyes), Centurion Vision system with active fluidics, or, Group II (n = 10 eyes), Centurion with Active Sentry.
Within each group, parameter set 1 (IOP 30 mm Hg, aspiration flow rate [AFR] 20 mL/min, and vacuum 600 mm Hg) and parameter set 2 (IOP 50 mm Hg, AFR 25 mL/min, and vacuum 600 mm Hg) were tested.
Real-time rate of drop and rise of IOP after occlusion break event (mm Hg per second) and percentage reduction of IOP from maximum during nuclear fragment removal were compared.
Results:
10 rabbits (20 eyes) were included.
Rate of drop of IOP after occlusion break was not significantly different between groups.
Rate of rise of IOP was statistically significantly higher in Group II with both parameter sets (199.
09 ± 69.
28 vs 94.
33 ± 45.
66 in parameter set 1, P = .
006; and 256 ± 45.
05 vs 165.
25 ± 51.
80 in parameter set 2, P = .
005), suggesting faster recovery to baseline IOP after occlusion break.
The mean percentage reduction of IOP from maximum was significantly higher in Group I (P = .
003).
Conclusions:
Rise of IOP to baseline after occlusion break event was faster and mean percentage reduction of IOP from maximum during nuclear fragment removal was lower when using the Centurion Vision system with the Active Sentry upgrade compared with the traditional handpiece.
The ability to sense IOP at the level of the handpiece with the Active Sentry upgrade allows faster mitigation of surge response.
Related Results
Woningcorporaties en Vastgoedontwikkeling
Woningcorporaties en Vastgoedontwikkeling
This summary highlights the findings of the PhD-thesis ‘Woningcorporaties en Vastgoedontwikkeling: Fit for Use’ (‘Housing associations and Real Estate Development: Fit for Use?’). ...
Changes in Intraocular Pressure After Phacoemulsification: A pilot study in Uganda
Changes in Intraocular Pressure After Phacoemulsification: A pilot study in Uganda
Objective: To determine the changes in intraocular pressure after phacoemulsification surgery among patients with cataracts in Southwestern Uganda.Methods: This was a retrospective...
Effect of Fasting on Intraocular Pressure
Effect of Fasting on Intraocular Pressure
Purpose: To determine the effect of fasting on intraocular pressure in fasting individuals.
Study Design: Cross sectional observational study.
Place and Duration of S...
Effect of Supra Choroid Triamcinolone Acetate On Intraocular Pressure
Effect of Supra Choroid Triamcinolone Acetate On Intraocular Pressure
Intraocular pressure is a critical parameter in maintaining the physiological function of the eye. Dysregulation of Intraocular pressure can lead to ocular pathologies. Objective: ...
Intraocular pressure variation during different surgical procedures in rabbit eyes.
Intraocular pressure variation during different surgical procedures in rabbit eyes.
Abstract Purpose To analyse intraocular and infusion line pressures during vitrectomy and cataract extraction (manual extracapsular extraction, phacofragmentation and phacoemulsifi...
Corneal endothelial cell loss after trabeculectomy and phacoemulsification in one or two steps: a prospective study
Corneal endothelial cell loss after trabeculectomy and phacoemulsification in one or two steps: a prospective study
Abstract
Objective
The objective of this study was to analyse the results of the surgical treatment of coexisting cataract and glaucoma and its effe...
Correlation study of changes in visual acuity, pupil diameter, Kappa angle, and Alpha angle after phacoemulsification surgery combined with multifocal IOL implantation
Correlation study of changes in visual acuity, pupil diameter, Kappa angle, and Alpha angle after phacoemulsification surgery combined with multifocal IOL implantation
Abstract
Purpose To study the changes and correlation of visual acuity, pupil size, kappa angle and Alpha angle after phacoemulsification combined with multifocal intraocul...
Predicting Fluid Responsiveness in Critically Ill Patients by Using Combined End-Expiratory and End-Inspiratory Occlusions With Echocardiography
Predicting Fluid Responsiveness in Critically Ill Patients by Using Combined End-Expiratory and End-Inspiratory Occlusions With Echocardiography
Objectives:
First, we aimed at assessing whether fluid responsiveness is predicted by the effects of an end-expiratory occlusion on the velocity-time integral of the le...