Remote auditing of radiotherapy facilities using optically stimulated luminescence dosimeters

View through CrossRef
Purpose:On 1 July 2012, the Australian Clinical Dosimetry Service (ACDS) released its Optically Stimulated Luminescent Dosimeter (OSLD) Level I audit, replacing the previous TLD based audit. The aim of this work is to present the results from this new service and the complete uncertainty analysis on which the audit tolerances are based.Methods:The audit release was preceded by a rigorous evaluation of the InLight® nanoDot OSLD system from Landauer (Landauer, Inc., Glenwood, IL). Energy dependence, signal fading from multiple irradiations, batch variation, reader variation, and dose response factors were identified and quantified for each individual OSLD. The detectors are mailed to the facility in small PMMA blocks, based on the design of the existing Radiological Physics Centre audit. Modeling and measurement were used to determine a factor that could convert the dose measured in the PMMA block, to dose in water for the facilityˈs reference conditions. This factor is dependent on the beam spectrum. The TPR20,10 was used as the beam quality index to determine the specific block factor for a beam being audited. The audit tolerance was defined using a rigorous uncertainty calculation. The audit outcome is then determined using a scientifically based two tiered action level approach. Audit outcomes within two standard deviations were defined as Pass (Optimal Level), within three standard deviations as Pass (Action Level), and outside of three standard deviations the outcome is Fail (Out of Tolerance).Results:To‐date the ACDS has audited 108 photon beams with TLD and 162 photon beams with OSLD. The TLD audit results had an average deviation from ACDS of 0.0% and a standard deviation of 1.8%. The OSLD audit results had an average deviation of −0.2% and a standard deviation of 1.4%. The relative combined standard uncertainty was calculated to be 1.3% (1σ). Pass (Optimal Level) was reduced to ≤2.6% (2σ), and Fail (Out of Tolerance) was reduced to >3.9% (3σ) for the new OSLD audit. Previously with the TLD audit the Pass (Optimal Level) and Fail (Out of Tolerance) were set at ≤4.0% (2σ) and >6.0% (3σ).Conclusions:The calculated standard uncertainty of 1.3% at one standard deviation is consistent with the measured standard deviation of 1.4% from the audits and confirming the suitability of the uncertainty budget derived audit tolerances. The OSLD audit shows greater accuracy than the previous TLD audit, justifying the reduction in audit tolerances. In the TLD audit, all outcomes were Pass (Optimal Level) suggesting that the tolerances were too conservative. In the OSLD audit 94% of the audits have resulted in Pass (Optimal level) and 6% of the audits have resulted in Pass (Action Level). All Pass (Action level) results have been resolved with a repeat OSLD audit, or an on‐site ion chamber measurement.
Title: Remote auditing of radiotherapy facilities using optically stimulated luminescence dosimeters
Description:
Purpose:On 1 July 2012, the Australian Clinical Dosimetry Service (ACDS) released its Optically Stimulated Luminescent Dosimeter (OSLD) Level I audit, replacing the previous TLD based audit.
The aim of this work is to present the results from this new service and the complete uncertainty analysis on which the audit tolerances are based.
Methods:The audit release was preceded by a rigorous evaluation of the InLight® nanoDot OSLD system from Landauer (Landauer, Inc.
, Glenwood, IL).
Energy dependence, signal fading from multiple irradiations, batch variation, reader variation, and dose response factors were identified and quantified for each individual OSLD.
The detectors are mailed to the facility in small PMMA blocks, based on the design of the existing Radiological Physics Centre audit.
Modeling and measurement were used to determine a factor that could convert the dose measured in the PMMA block, to dose in water for the facilityˈs reference conditions.
This factor is dependent on the beam spectrum.
The TPR20,10 was used as the beam quality index to determine the specific block factor for a beam being audited.
The audit tolerance was defined using a rigorous uncertainty calculation.
The audit outcome is then determined using a scientifically based two tiered action level approach.
Audit outcomes within two standard deviations were defined as Pass (Optimal Level), within three standard deviations as Pass (Action Level), and outside of three standard deviations the outcome is Fail (Out of Tolerance).
Results:To‐date the ACDS has audited 108 photon beams with TLD and 162 photon beams with OSLD.
The TLD audit results had an average deviation from ACDS of 0.
0% and a standard deviation of 1.
8%.
The OSLD audit results had an average deviation of −0.
2% and a standard deviation of 1.
4%.
The relative combined standard uncertainty was calculated to be 1.
3% (1σ).
Pass (Optimal Level) was reduced to ≤2.
6% (2σ), and Fail (Out of Tolerance) was reduced to >3.
9% (3σ) for the new OSLD audit.
Previously with the TLD audit the Pass (Optimal Level) and Fail (Out of Tolerance) were set at ≤4.
0% (2σ) and >6.
0% (3σ).
Conclusions:The calculated standard uncertainty of 1.
3% at one standard deviation is consistent with the measured standard deviation of 1.
4% from the audits and confirming the suitability of the uncertainty budget derived audit tolerances.
The OSLD audit shows greater accuracy than the previous TLD audit, justifying the reduction in audit tolerances.
In the TLD audit, all outcomes were Pass (Optimal Level) suggesting that the tolerances were too conservative.
In the OSLD audit 94% of the audits have resulted in Pass (Optimal level) and 6% of the audits have resulted in Pass (Action Level).
All Pass (Action level) results have been resolved with a repeat OSLD audit, or an on‐site ion chamber measurement.

Related Results

Current Developments in Radiotherapy Quality Control Methods
Current Developments in Radiotherapy Quality Control Methods
Quality control is of critical importance to ensure the accurate and reliable implementation of treatment in radiotherapy. Radiotherapy is a primary method of treating diseases suc...
Investigation of alanine dosimeter for output measurement in radiotherapy: Results from multicenter in Thailand
Investigation of alanine dosimeter for output measurement in radiotherapy: Results from multicenter in Thailand
Background: Alanine dosimeters are generally used in high-dose industrial applications (kGy). Later, research into employing alanine as a dosimeter in radiotherapy (1-20 Gy) has in...
Use of data auditing for encrypted data stored in cloud environment
Use of data auditing for encrypted data stored in cloud environment
The exponential growth of data generation and the increasing reliance on cloud-based storage solutions have raised significant concerns regarding data security and privacy. The wid...
RESEARCH ON VALUE-ADDED INTERNAL AUDITING IN ENTERPRISES BASED ON VALUE CHAIN THEORY
RESEARCH ON VALUE-ADDED INTERNAL AUDITING IN ENTERPRISES BASED ON VALUE CHAIN THEORY
With the continuous development of the market economy, the business environment of enterprises has become increasingly complex. As a critical component of corporate governance, int...
Radiotherapy refusal in breast cancer with breast-conserving surgery
Radiotherapy refusal in breast cancer with breast-conserving surgery
Abstract Background Although radiotherapy after breast-conserving surgery has been the standard treatment for breast cancer, some people still refus...
Potential of feldspars single-grain luminescence to reveal sediment input and transport in fluvial systems
Potential of feldspars single-grain luminescence to reveal sediment input and transport in fluvial systems
Le potentiel de la luminescence mono-grain des feldspaths pour révéler l'apport et le transport de sédiments dans les systèmes fluviatiles Les rivières jouent un rô...

Back to Top