Enhanced fluid drainage using a novel multi-pod drainage catheter: An in vitro evaluation
View through CrossRef
BACKGROUND: A multi-pod catheter (MPC) is a large drainage catheter that can house multiple smaller retractable (MPC-R) and deployable catheters (MPC-D) within the body. OBJECTIVE: The drainage capabilities and resistance to clogging of a novel MPC have been assessed. METHODS: The drainage capabilities are evaluated by placing the MPC in a bag of either a non-clogging (H2O) or clogging medium. The results are then compared to matched-size single-lumen catheters with either a close (CTC) or open tip (OTC). The means of five test runs were used to measure drainage rate, maximum drained volume (MaxDV), and time to drain the first 200 mL (TTD200). RESULTS: In the non-clogging medium, MPC-D had a slightly higher MaxDV than MPC-R, and higher flow rate than CTC and MPC-R. Moreover, MPC-D needed less TTD200 than MPC-R. In the clogging medium, MPC-D had a higher MaxDV than CTC and OTC, higher flow rate, and faster TTD200 than CTC. However, comparison with MPC-R showed no significant difference. CONCLUSION: The novel catheter may offer superior drainage compared to the single-lumen catheter in a clogging medium, implying various clinical applications, particularly when clogging is a potential risk. Further testing may be required to simulate various clinical scenarios.
Title: Enhanced fluid drainage using a novel multi-pod drainage catheter: An in vitro evaluation
Description:
BACKGROUND: A multi-pod catheter (MPC) is a large drainage catheter that can house multiple smaller retractable (MPC-R) and deployable catheters (MPC-D) within the body.
OBJECTIVE: The drainage capabilities and resistance to clogging of a novel MPC have been assessed.
METHODS: The drainage capabilities are evaluated by placing the MPC in a bag of either a non-clogging (H2O) or clogging medium.
The results are then compared to matched-size single-lumen catheters with either a close (CTC) or open tip (OTC).
The means of five test runs were used to measure drainage rate, maximum drained volume (MaxDV), and time to drain the first 200 mL (TTD200).
RESULTS: In the non-clogging medium, MPC-D had a slightly higher MaxDV than MPC-R, and higher flow rate than CTC and MPC-R.
Moreover, MPC-D needed less TTD200 than MPC-R.
In the clogging medium, MPC-D had a higher MaxDV than CTC and OTC, higher flow rate, and faster TTD200 than CTC.
However, comparison with MPC-R showed no significant difference.
CONCLUSION: The novel catheter may offer superior drainage compared to the single-lumen catheter in a clogging medium, implying various clinical applications, particularly when clogging is a potential risk.
Further testing may be required to simulate various clinical scenarios.
Related Results
Blunt Chest Trauma and Chylothorax: A Systematic Review
Blunt Chest Trauma and Chylothorax: A Systematic Review
Abstract
Introduction: Although traumatic chylothorax is predominantly associated with penetrating injuries, instances following blunt trauma, as a rare and challenging condition, ...
CHARACTER ASSOCIATION AND PATH ANALYSIS IN VEGETABLE COWPEA
CHARACTER ASSOCIATION AND PATH ANALYSIS IN VEGETABLE COWPEA
The genotypic and phenotypic correlations of green pod yield with different components were estimated from 20 genotypes of vegetable cowpea pooled over two seasons. The genotypic a...
Epidural Catheter Reconnection
Epidural Catheter Reconnection
Background
An in vitro model of epidural catheter contamination was used to determine if disconnected catheters can be safely reconnected.
...
Reliability Analysis Of Subsea Control Pod
Reliability Analysis Of Subsea Control Pod
ABSTRACT
A subsea production control pod was withdrawn from service, disassembled and analyzed to evaluate pod design. This paper examines the results and compare...
Investigations on Per se Performance, Genetic Variability and Correlations in Vegetable Cowpea [Vigna unguiculata (L.) Walp.] Germplasm for Yield and Its Attributing Traits
Investigations on Per se Performance, Genetic Variability and Correlations in Vegetable Cowpea [Vigna unguiculata (L.) Walp.] Germplasm for Yield and Its Attributing Traits
Background: Cowpea is a multiple-purpose, drought-tolerant, climate-smart and leguminous pulse crop. Recent studies established this crop as a potential vegetable crop with nitroge...
GWAS Of Pod Morphological and Color Characters in Common Bean
GWAS Of Pod Morphological and Color Characters in Common Bean
Abstract
BackgroundCommon bean (Phaseolus vulgaris L.) is an important legume species which can be consumed as immature pods and dry seeds after re-hydration and cooking. M...
#3349 METICULOUS CATHETER CARE CAN REDUCE CATHETER-RELATED BLOODSTREAM INFECTIONS SIGNIFICANTLY IN HAEMODIALYSIS PATIENTS – A 5-YEAR SINGLE CENTRE STUDY
#3349 METICULOUS CATHETER CARE CAN REDUCE CATHETER-RELATED BLOODSTREAM INFECTIONS SIGNIFICANTLY IN HAEMODIALYSIS PATIENTS – A 5-YEAR SINGLE CENTRE STUDY
Abstract
Background and Aims
The use of central venous catheters as hemodialysis (HD) vascular access is a leading contributor t...
Peripherally Inserted Central Catheters in Children: A Prospective Single-Center Analysis of Associated Complications
Peripherally Inserted Central Catheters in Children: A Prospective Single-Center Analysis of Associated Complications
Highlights
Overall, 72% (n = 63) of PICCs remained in situ until intended therapy was completed.
In 24...