Loop Securities of Arthroscopic Sliding‐Knot Techniques When the Suture Loop Is Not Evenly Tensioned

View through CrossRef
Purpose The purpose of this study was to evaluate the loop security of arthroscopic sliding knots when tension is only applied to the post strand and not the loop strand. Methods Six different locking sliding knots (Weston, Nicky, Roeder, SMC, San Diego, and Dines) were included. Loop securities were evaluated in 2 ways: with a conventional method (equal tension applied to the suture loop) and with a worst‐case scenario (WCS) method (only the post strand of the suture loop was tensioned). Differences between test methods were evaluated for significance. To help assess the applicability of each test method, loop‐security testing in a cadaveric shoulder was performed with 1 type of knot (SMC). Results Loop securities with the conventional method versus the WCS method were as follows: 10.74 ± 4.20 N versus 6.90 ± 3.90 N for Weston, 21.25 ± 14.74 N versus 8.73 ± 3.35 N for Nicky, 26.14 ± 15.57 N versus 7.95 ± 4.23 N for Roeder, 42.67 ± 22.96 N versus 8.67 ± 4.33 N for SMC, 52.99 ± 21.36 N versus 18.25 ± 10.58 N for San Diego, and 89.27 ± 27.96 N versus 12.48 ± 3.40 N for Dines ( P < .05 for each knot). All knots failed at significantly lower loads when the suture loop was not evenly tensioned. Cadaveric testing (SMC) resulted in a loop security of 5.53 ± 6.06 N, which was similar to the WCS setting. Conclusions The locking mechanism of the sliding knots is maintained when the suture loop is evenly tensioned at both post and non‐post strands. When tension is not applied to the non‐post strand side, the knots slide more easily and fail at lower loads than previously reported. Clinical Relevance When surgeons tie locking sliding knots in single‐row rotator cuff repair, they should be aware that the knots could fail at much lower loads than previously reported.
Title: Loop Securities of Arthroscopic Sliding‐Knot Techniques When the Suture Loop Is Not Evenly Tensioned
Description:
Purpose The purpose of this study was to evaluate the loop security of arthroscopic sliding knots when tension is only applied to the post strand and not the loop strand.
Methods Six different locking sliding knots (Weston, Nicky, Roeder, SMC, San Diego, and Dines) were included.
Loop securities were evaluated in 2 ways: with a conventional method (equal tension applied to the suture loop) and with a worst‐case scenario (WCS) method (only the post strand of the suture loop was tensioned).
Differences between test methods were evaluated for significance.
To help assess the applicability of each test method, loop‐security testing in a cadaveric shoulder was performed with 1 type of knot (SMC).
Results Loop securities with the conventional method versus the WCS method were as follows: 10.
74 ± 4.
20 N versus 6.
90 ± 3.
90 N for Weston, 21.
25 ± 14.
74 N versus 8.
73 ± 3.
35 N for Nicky, 26.
14 ± 15.
57 N versus 7.
95 ± 4.
23 N for Roeder, 42.
67 ± 22.
96 N versus 8.
67 ± 4.
33 N for SMC, 52.
99 ± 21.
36 N versus 18.
25 ± 10.
58 N for San Diego, and 89.
27 ± 27.
96 N versus 12.
48 ± 3.
40 N for Dines ( P < .
05 for each knot).
All knots failed at significantly lower loads when the suture loop was not evenly tensioned.
Cadaveric testing (SMC) resulted in a loop security of 5.
53 ± 6.
06 N, which was similar to the WCS setting.
Conclusions The locking mechanism of the sliding knots is maintained when the suture loop is evenly tensioned at both post and non‐post strands.
When tension is not applied to the non‐post strand side, the knots slide more easily and fail at lower loads than previously reported.
Clinical Relevance When surgeons tie locking sliding knots in single‐row rotator cuff repair, they should be aware that the knots could fail at much lower loads than previously reported.

Related Results

Concomitant Medial Meniscal Root Repair with Extrusion Repair (Centralization Technique)
Concomitant Medial Meniscal Root Repair with Extrusion Repair (Centralization Technique)
Background: Meniscal extrusion is a phenomenon in which a degenerative posterior horn tear, radial tear, or root tear results in displacement of the body of the meniscu...
Arthroscopic Suture Anchor Capsulorrhaphy Versus Suture Capsulorrhaphy in a Cadaveric Model (SS‐01)
Arthroscopic Suture Anchor Capsulorrhaphy Versus Suture Capsulorrhaphy in a Cadaveric Model (SS‐01)
Introduction Recurrent anterior glenohumeral instability is challenging to treat. Arthroscopic capsulorrhaphy utilizing either a solitary suture or suture ancho...
The HU Knot: A New Sliding Knot for Arthroscopic Surgery
The HU Knot: A New Sliding Knot for Arthroscopic Surgery
Abstract Sliding knots are commonly used in arthroscopic knot tying. The Surgical knot as a hand‐tied knot is very secure. We found that the Surgical knot could b...
The “Cruciate Suture” for Arthroscopic Meniscal Repair: A New Technique
The “Cruciate Suture” for Arthroscopic Meniscal Repair: A New Technique
Abstract Meniscal repair has become the treatment of choice whenever a reparable tear is diagnosed. Fixation strength of the repair technique is always of paramou...
The Future of Securities Law in the Supreme Court
The Future of Securities Law in the Supreme Court
Since the enactment of the first federal securities statute in 1933, securities law has illustrated key shifts in the Supreme Court’s jurisprudence. During the New Deal, the Court’...
Suture-Induced Tubo-Ovarian Abscess: A Case Report with Literature Review
Suture-Induced Tubo-Ovarian Abscess: A Case Report with Literature Review
Abstract Introduction Suture is an underreported cause for tubo-ovarian abscess (TOA) that can cause significant morbidity. This report describes a case of TOA arising from a silk ...
Integral Suture‐Handling Techniques for Arthroscopic Sliding Knots
Integral Suture‐Handling Techniques for Arthroscopic Sliding Knots
Abstract In arthroscopic tissue repair, the final step is achieving adequate tissue approximation with a secure knot. Th...

Back to Top