Can Video Head Impulse Test Determine the Early Effect of Cochlear Implantation on the Semicircular Canals Function?
View through CrossRef
Background and Aim: Although cochlear implantation (CI) is a safe surgical procedure for severe to profound sensorineural hearing loss (SNHL) but, due to the embryological and anatomical connection between the vestibular and cochlear structures, vestibular dysfunction may occur after CI. Video head impulse test (vHIT) is a reliable test for assessing the function of semicircular canals (SCCs). This study aimed to determine the early effect of CI on SCCs function, by comparing pre- and post-operative vHIT results.
Methods: In this cross-sectional study, participants were 22 adults with SNHL scheduled for unilateral CI in the right ear and 22 age-matched healthy subjects as a control group. The vHIT was conducted before and two weeks after CI.
Results: The mean vHIT gains in the SNHL group were significantly lower than in controls, with a large effect size. Furthermore, the mean vHIT gains in the right lateral SCCs (p<0.001) and right anterior SCCs (p=0.003) were significantly reduced after CI, compared to the gain values before CI, with a large pooled effect size. However, these differences were not statistically significant for the right posterior SCCs. The comparison of vHIT gains in the non- implanted ear showed no statistically significant difference between pre- and post-operative phases.
Conclusion: The vHIT is a useful clinical method to detect the early effects of CI on the function of SCCs. These effects are more obvious in the lateral and superior SCCs in the implanted ear.
Keywords: Cochlear implantation; semicircular canals; video head impulse test
Title: Can Video Head Impulse Test Determine the Early Effect of Cochlear Implantation on the Semicircular Canals Function?
Description:
Background and Aim: Although cochlear implantation (CI) is a safe surgical procedure for severe to profound sensorineural hearing loss (SNHL) but, due to the embryological and anatomical connection between the vestibular and cochlear structures, vestibular dysfunction may occur after CI.
Video head impulse test (vHIT) is a reliable test for assessing the function of semicircular canals (SCCs).
This study aimed to determine the early effect of CI on SCCs function, by comparing pre- and post-operative vHIT results.
Methods: In this cross-sectional study, participants were 22 adults with SNHL scheduled for unilateral CI in the right ear and 22 age-matched healthy subjects as a control group.
The vHIT was conducted before and two weeks after CI.
Results: The mean vHIT gains in the SNHL group were significantly lower than in controls, with a large effect size.
Furthermore, the mean vHIT gains in the right lateral SCCs (p<0.
001) and right anterior SCCs (p=0.
003) were significantly reduced after CI, compared to the gain values before CI, with a large pooled effect size.
However, these differences were not statistically significant for the right posterior SCCs.
The comparison of vHIT gains in the non- implanted ear showed no statistically significant difference between pre- and post-operative phases.
Conclusion: The vHIT is a useful clinical method to detect the early effects of CI on the function of SCCs.
These effects are more obvious in the lateral and superior SCCs in the implanted ear.
Keywords: Cochlear implantation; semicircular canals; video head impulse test.
Related Results
Chronic Cochlear Implantation with and without Electric Stimulation in a Mouse Model Induces Robust Cochlear Influx of CX3CR1+/GFP Macrophages
Chronic Cochlear Implantation with and without Electric Stimulation in a Mouse Model Induces Robust Cochlear Influx of CX3CR1+/GFP Macrophages
AbstractBackgroundCochlear implantation is an effective auditory rehabilitation strategy for those with profound hearing loss, including those with residual low frequency hearing t...
Vestibular Function After Cochlear Implantation in Partial Deafness Treatment
Vestibular Function After Cochlear Implantation in Partial Deafness Treatment
Introduction: Cochlear implantation is a fully accepted method of treating individuals with profound hearing loss. Since the indications for cochlear implantation have broadened an...
Cochlear Patency after Translabyrinthine and Retrosigmoid Vestibular Schwannoma Surgery
Cochlear Patency after Translabyrinthine and Retrosigmoid Vestibular Schwannoma Surgery
OBJECTIVES: To assess the incidence and onset of cochlear obliteration after translabyrinthine and retrosigmoid vestibular schwannoma surgery.
MATERIALS and METHODS: We retrospect...
LIFESPAN OF NITI ROTARY FILES WITH 360°ROTATION AND RECIPROCATING MOTION
LIFESPAN OF NITI ROTARY FILES WITH 360°ROTATION AND RECIPROCATING MOTION
Objective:
To assess the life of two different rotary nickel-titanium (NiTi) instruments in curved canals with complete protocol, by using 360°rotation motion and reciprocati...
Assessment of Benefits of Cochlear Implantation in Children With Single-Sided Deafness
Assessment of Benefits of Cochlear Implantation in Children With Single-Sided Deafness
Objectives:
To assess the hearing benefits of cochlear implantation and its daily usage in children with single-sided deafness (SSD). We compared the results of...
Radiographic Measurement of Cochlear in Sudanese Population using High Resolution Computed Tomography (HRCT)
Radiographic Measurement of Cochlear in Sudanese Population using High Resolution Computed Tomography (HRCT)
This study aimed to determine the measurements of the cochlea among healthy subjects and hearing deafness subjects using a High Resolution Computed Tomography (HRCT). A total of 23...
Convenience and short comings among paediatric cochlear implant candidates
Convenience and short comings among paediatric cochlear implant candidates
Background: Cochlear implants have revolutionized the treatment of severe-to-profound sensorineural hearing loss in children. However, the convenience and shortcomings experienced ...
Assessing Array-Type Differences in Cochlear Implant Users Using the Panoramic ECAP Method
Assessing Array-Type Differences in Cochlear Implant Users Using the Panoramic ECAP Method
OBJECTIVES: Cochlear-implant companies manufacture devices with different electrode array types. Some arrays have a straight geometry designed for minimal neuronal trauma, while ot...