Ambulatory 24‐Hour Double‐Probe pH Monitoring: The Importance of Manometry
View through CrossRef
AbstractObjective Historically, manometry has been used for sphincter localization before ambulatory 24‐hour double‐probe pH monitoring to ensure accurate placement of the probes. Recently, direct‐vision placement (DVP), using transnasal fiberoptic laryngoscopy (TFL), has been offered as an alternative technique. Presumably, DVP might be used to precisely place the proximal (pharyngeal) pH probe; however, using DVP, there appears to be no way to accurately position the distal (esophageal) probe. The purpose of this study was to evaluate the accuracy of DVP for pH probe placement using manometric measurement as the gold standard.Methods Thirty patients undergoing pH monitoring participated in this prospective study. Each subject underwent manometric examination of the esophagus to determine the precise location of the upper and lower esophageal sphincters (UES and LES). In addition, external anatomic landmarks were used to estimate interprobe distances. A physician blinded to the manometry results then placed a pH catheter using DVP so that the proximal probe was located just above the UES. The results were recorded and compared with those obtained by manometry.Results Accurate DVP of the proximal pH probe was achieved in 70% (23 of 30) of the subjects. The use of external anatomic landmarks to estimate interprobe distance resulted in accurate positioning of the distal probe in only 40% (12 of 30) of the subjects. Using fixed interprobe distances of 15 cm and 20 cm, distal probe position accuracy was 3% (1 of 30) and 40% (12 of 30), respectively. Therefore, using DVP, the distal esophageal probe was in an incorrect position in 60% to 97% of subjects.Conclusion For double‐probe pH monitoring, the proximal probe can be accurately positioned by DVP; however, there is no precise way to determine the interprobe distance required to correctly position the distal pH probe. Failure to accurately position the distal probe results in grossly inaccurate esophageal acid‐exposure times. Thus, manometry is needed to ensure valid double‐probe pH monitoring data.
Title: Ambulatory 24‐Hour Double‐Probe pH Monitoring: The Importance of Manometry
Description:
AbstractObjective Historically, manometry has been used for sphincter localization before ambulatory 24‐hour double‐probe pH monitoring to ensure accurate placement of the probes.
Recently, direct‐vision placement (DVP), using transnasal fiberoptic laryngoscopy (TFL), has been offered as an alternative technique.
Presumably, DVP might be used to precisely place the proximal (pharyngeal) pH probe; however, using DVP, there appears to be no way to accurately position the distal (esophageal) probe.
The purpose of this study was to evaluate the accuracy of DVP for pH probe placement using manometric measurement as the gold standard.
Methods Thirty patients undergoing pH monitoring participated in this prospective study.
Each subject underwent manometric examination of the esophagus to determine the precise location of the upper and lower esophageal sphincters (UES and LES).
In addition, external anatomic landmarks were used to estimate interprobe distances.
A physician blinded to the manometry results then placed a pH catheter using DVP so that the proximal probe was located just above the UES.
The results were recorded and compared with those obtained by manometry.
Results Accurate DVP of the proximal pH probe was achieved in 70% (23 of 30) of the subjects.
The use of external anatomic landmarks to estimate interprobe distance resulted in accurate positioning of the distal probe in only 40% (12 of 30) of the subjects.
Using fixed interprobe distances of 15 cm and 20 cm, distal probe position accuracy was 3% (1 of 30) and 40% (12 of 30), respectively.
Therefore, using DVP, the distal esophageal probe was in an incorrect position in 60% to 97% of subjects.
Conclusion For double‐probe pH monitoring, the proximal probe can be accurately positioned by DVP; however, there is no precise way to determine the interprobe distance required to correctly position the distal pH probe.
Failure to accurately position the distal probe results in grossly inaccurate esophageal acid‐exposure times.
Thus, manometry is needed to ensure valid double‐probe pH monitoring data.
Related Results
Ambulatory blood pressure of adults in Ohasama, Japan.
Ambulatory blood pressure of adults in Ohasama, Japan.
We performed a cross-sectional study in a small town in northern Japan to evaluate the distribution, reference values, and daily variation in ambulatory blood pressure. A total of ...
Effects of reduced blood pressure dipping on the progression of chronic kidney disease in children
Effects of reduced blood pressure dipping on the progression of chronic kidney disease in children
Abstract
Background: Hypertension is a common complication of chronic kidney disease (CKD) in children. It is related to the progression of CKD. However, current guidelines...
Pleural Manometry—Basics for Clinical Practice
Pleural Manometry—Basics for Clinical Practice
Abstract
Purpose of Review
The aim of this paper is to present basic data on pleural manometry and to outline the advances in its use as both a rese...
Analisis Produktivitas Alat Berat Pada Pekerjaan Peningkatan Jalan Batas Kota Pegaf – Batas Kabupaten Manokwari Selatan
Analisis Produktivitas Alat Berat Pada Pekerjaan Peningkatan Jalan Batas Kota Pegaf – Batas Kabupaten Manokwari Selatan
Roads in Manokwari City still need a lot to be improved, especially on the Pegaf city boundary road - South Manokwari district, because this road connects the border between the Ci...
Day long Activities of Indian Blue Peafowl in Bangladesh National Zoo
Day long Activities of Indian Blue Peafowl in Bangladesh National Zoo
The research work was carried out from July 2016 to June 2017 to explore the day long activities of Indian Blue Peafowl habitat at Bangladesh National Zoo. The objectives of the st...
Environmental Surveillance Protocols for Highly Pathogenic Avian Influenza (HPAI) v2
Environmental Surveillance Protocols for Highly Pathogenic Avian Influenza (HPAI) v2
EnvironmentalSurveillance Protocols for Highly Pathogenic Avian Influenza (HPAI) This comprehensive protocol suite enables systematic environmental surveillance for avian influenza...
Cost-Effectiveness of Ambulatory Blood Pressure
Cost-Effectiveness of Ambulatory Blood Pressure
Accurate diagnosis of hypertension and prognosis for future cardiovascular events can be enhanced through the use of 24-hour ambulatory blood pressure monitoring. It has been sugge...
Subclinical arterial stiffness in prehypertensives
Subclinical arterial stiffness in prehypertensives
Objective
To assess the difference of arterial stiffness in prehypertensive subjects in comparison to normotensive subjects.
...