TU‐H‐206‐02: Novel Linearly‐Filled Derenzo PET Phantom Design

View through CrossRef
Purpose:To design a linearly‐filled Derenzo positron emission tomography (PET) phantom, eliminating the extraneous radioisotope volumes in a conventional reservoir‐type design. This activity reduction combined with the elimination of bubbles in smaller phantom channels would significantly reduce personnel dose, radioisotope cost, and would improve image quality by reducing out‐of‐slice activity scatter.Methods:A computer‐aided design (CAD) was created of a modular Derenzo phantom consisting of three phantom layers with gaskets between the layers. The central piece contains the active pattern volume and channels connecting adjacent rods in a serpentine pattern. The two end‐pieces contained an inlet and an outlet for filling purposes. Phantom prototypes were 3D printed on a Viper Si2 stereolithography machine. The two gaskets were fabricated from silicon sheets using a PLS 6.75 laser cutter. Phantoms were held together by pass‐through glass‐filled nylon bolts and nuts. Phantoms were filled with 52Mn, 64Cu, 74Br, and 124I for testing, and were imaged on a Siemens Inveon MicroPET scanner.Results:Four phantom prototypes were constructed using male Leur Lock fittings for inlet/outlet ports. 3D printed layers were sanded to ensure proper coupling to the silicon gaskets. The filling volume for each prototype was approximately 2.4 mL. The filling process was found to be rapid, leak‐tight, and with minimal back‐pressure. PET images were reconstructed by OSEM3D, and axial slices along the phantom pattern length were averaged to provide final images. Image distortion was isotope dependent with 52Mn and 64Cu having the least distortion and 124I having the most distortion.Conclusion:These results indicate that the linearlyfilled Derenzo design improves on conventional reservoir‐type designs by eliminating potential bubbles in small channels and by reducing activity level, radioisotope volume, radioisotope cost, personnel dose, filling time, and out‐of‐slice activity scatterThe method described in this abstract has been filed as a patent application to the US Patent and Trade Office by the Wisconsin Alumni Research Foundation (WARF).
Title: TU‐H‐206‐02: Novel Linearly‐Filled Derenzo PET Phantom Design
Description:
Purpose:To design a linearly‐filled Derenzo positron emission tomography (PET) phantom, eliminating the extraneous radioisotope volumes in a conventional reservoir‐type design.
This activity reduction combined with the elimination of bubbles in smaller phantom channels would significantly reduce personnel dose, radioisotope cost, and would improve image quality by reducing out‐of‐slice activity scatter.
Methods:A computer‐aided design (CAD) was created of a modular Derenzo phantom consisting of three phantom layers with gaskets between the layers.
The central piece contains the active pattern volume and channels connecting adjacent rods in a serpentine pattern.
The two end‐pieces contained an inlet and an outlet for filling purposes.
Phantom prototypes were 3D printed on a Viper Si2 stereolithography machine.
The two gaskets were fabricated from silicon sheets using a PLS 6.
75 laser cutter.
Phantoms were held together by pass‐through glass‐filled nylon bolts and nuts.
Phantoms were filled with 52Mn, 64Cu, 74Br, and 124I for testing, and were imaged on a Siemens Inveon MicroPET scanner.
Results:Four phantom prototypes were constructed using male Leur Lock fittings for inlet/outlet ports.
3D printed layers were sanded to ensure proper coupling to the silicon gaskets.
The filling volume for each prototype was approximately 2.
4 mL.
The filling process was found to be rapid, leak‐tight, and with minimal back‐pressure.
PET images were reconstructed by OSEM3D, and axial slices along the phantom pattern length were averaged to provide final images.
Image distortion was isotope dependent with 52Mn and 64Cu having the least distortion and 124I having the most distortion.
Conclusion:These results indicate that the linearlyfilled Derenzo design improves on conventional reservoir‐type designs by eliminating potential bubbles in small channels and by reducing activity level, radioisotope volume, radioisotope cost, personnel dose, filling time, and out‐of‐slice activity scatterThe method described in this abstract has been filed as a patent application to the US Patent and Trade Office by the Wisconsin Alumni Research Foundation (WARF).

Related Results

SEMANA DE ENFERMAGEM E SEUS ASPECTOS SOCIAIS NA VALORIZAÇÃO PROFISSIONAL: UM RELATO DE EXPERIÊNCIA DO GRUPO PET-ENFERMAGEM
SEMANA DE ENFERMAGEM E SEUS ASPECTOS SOCIAIS NA VALORIZAÇÃO PROFISSIONAL: UM RELATO DE EXPERIÊNCIA DO GRUPO PET-ENFERMAGEM
A enfermagem é o pilar da assistência pois está na linha de frente do cuidado holístico, todavia esta é estigmatizada e desvalorizada, assim como não possui reconhecimento consider...
Adjektiivin ja adverbin erosta
Adjektiivin ja adverbin erosta
On the limit between adjectives and adverbs in Finnish (englanti)Kielenaineksetaikaisin (kieli: suomi, sivulla: 208)enimmin (kieli: suomi, sivulla: 201)eniten (kieli: suomi, sivull...
Comparative study of total-body PET and PET/MR in the diagnosis of liver metastases
Comparative study of total-body PET and PET/MR in the diagnosis of liver metastases
ObjectiveTo compare the diagnostic differences between total-body PET/CT (positron emission tomography/computed tomography) and PET/MR (positron emission tomography/magnetic resona...
Body image and perception among adults with and without phantom limb pain
Body image and perception among adults with and without phantom limb pain
AbstractBackgroundFollowing lower‐limb amputation, phantom limb pain (i.e., pain perceived as coming from the amputated portion of the limb) is common. Phantom limb pain may be ass...
Baseline Staging Evaluation in Lymphoma: The Role of FDG PET, CT, and Bone Marrow Biopsy
Baseline Staging Evaluation in Lymphoma: The Role of FDG PET, CT, and Bone Marrow Biopsy
Abstract Abstract 2640 BACKGROUND: The revised response criteria for malignant lymphoma (Cheson et al JCO 25:579 ...
Expression and prognostic value of hsa-miR-206 in non-triple-negative breast cancer
Expression and prognostic value of hsa-miR-206 in non-triple-negative breast cancer
Abstract Objective This study aims to analyze the expression and prognostic value of hsa-miR-206 in non-triple-negative breast cancer. Methods The expression of has-miR-206...
Positron Emission Tomography
Positron Emission Tomography
The aim of this study to evaluate the positron emission tomography (PET) scan technology and its role as a diagnostic tool in health care. New PET technology provide the integrated...

Back to Top