Quantitative assessment of variation in CBCT image technical parameters related to CBCT detector lateral-offset position
View through CrossRef
Objectives: To assess the effect of CBCT detector position (aligned/lateral-offset) on image technical parameters (mean voxel value - MVV, standard deviation of voxel value (SDVV) distribution), comparing peripheral regions of interest (ROIs) to the central ROI in CBCT volumes. Methods: 40 CBCT volumes of a wax phantom were acquired in six units with aligned and/or lateral-offset detectors: Cranex 3Dx (CRA), Ortophos SL (ORT), Picasso Trio (PIC), Promax 3D Mid (PRO), Scanora 3D (SCA), and X1. Four image-acquisition protocols used an aligned detector, and four a lateral-offset detector. In each volume, 13 ROIs (12 peripheral and 1 central) were evaluated. MVV and SDVV of the peripheral ROI were compared to those of the central ROI in the volume. MVVD (the difference in percentage, between the MVV of a peripheral and the central ROI) was calculated. Results: For aligned-detectors, MVV increased (ORT and PRO) or decreased (CRA and X1) in the ROIs farther from the centre. For lateral-offset detectors, ROIs farther from the centre showed increased MVV. SDVV for most aligned detectors was lower, the nearer the ROI was to the centre. For lateral-offset detectors, it was lower for the peripheral ROIs, except with PIC. Range for MVVD was −32.8% to 22.8% for units with aligned detectors, and −20.7% to 69.5% for lateral-offset detectors. Conclusion: Lateral-offset detectors to acquire CBCT images significantly change SDVV distribution within the field-of-view, and lead to MVVD with increased range, compared to aligned detectors. This must be taken in consideration in the clinic, if voxel-value dependent measurements are to be performed.
Oxford University Press (OUP)
Title: Quantitative assessment of variation in CBCT image technical parameters related to CBCT detector lateral-offset position
Description:
Objectives: To assess the effect of CBCT detector position (aligned/lateral-offset) on image technical parameters (mean voxel value - MVV, standard deviation of voxel value (SDVV) distribution), comparing peripheral regions of interest (ROIs) to the central ROI in CBCT volumes.
Methods: 40 CBCT volumes of a wax phantom were acquired in six units with aligned and/or lateral-offset detectors: Cranex 3Dx (CRA), Ortophos SL (ORT), Picasso Trio (PIC), Promax 3D Mid (PRO), Scanora 3D (SCA), and X1.
Four image-acquisition protocols used an aligned detector, and four a lateral-offset detector.
In each volume, 13 ROIs (12 peripheral and 1 central) were evaluated.
MVV and SDVV of the peripheral ROI were compared to those of the central ROI in the volume.
MVVD (the difference in percentage, between the MVV of a peripheral and the central ROI) was calculated.
Results: For aligned-detectors, MVV increased (ORT and PRO) or decreased (CRA and X1) in the ROIs farther from the centre.
For lateral-offset detectors, ROIs farther from the centre showed increased MVV.
SDVV for most aligned detectors was lower, the nearer the ROI was to the centre.
For lateral-offset detectors, it was lower for the peripheral ROIs, except with PIC.
Range for MVVD was −32.
8% to 22.
8% for units with aligned detectors, and −20.
7% to 69.
5% for lateral-offset detectors.
Conclusion: Lateral-offset detectors to acquire CBCT images significantly change SDVV distribution within the field-of-view, and lead to MVVD with increased range, compared to aligned detectors.
This must be taken in consideration in the clinic, if voxel-value dependent measurements are to be performed.
Related Results
Comparison between hybrid CBCT and CBCT in detection of vertical root fracture
Comparison between hybrid CBCT and CBCT in detection of vertical root fracture
Cone beam computerized tomography (CBCT) has been wildly used in many fields of dentistry, especially to investigate the vertical root fracture (VRF). CBCT systems were consisted o...
Characterization of a novel HgCdTe focal plane array for ground and space astronomy through innovative infrared setups
Characterization of a novel HgCdTe focal plane array for ground and space astronomy through innovative infrared setups
(English) Nowadays, mercury-cadmium-telluride (MCT) short-wave infrared (SWIR) detectors are widely used in cutting-edge space
missions and ground-based telescopes. They take adva...
Evaluation of a novel CBCT conversion method implemented in a treatment planning system
Evaluation of a novel CBCT conversion method implemented in a treatment planning system
Abstract
Background
To evaluate a novel CBCT conversion algorithm for dose calculation implemented in a research version ...
Comparing Unet training with three different datasets to correct CBCT images for prostate radiotherapy dose calculations
Comparing Unet training with three different datasets to correct CBCT images for prostate radiotherapy dose calculations
Abstract
Image intensity correction is crucial to enable cone beam computed tomography (CBCT) based radiotherapy dose calculations. This study evaluated three dif...
Barriers Toward Better CBCT Knowledge, Attitude, and Practice
Barriers Toward Better CBCT Knowledge, Attitude, and Practice
Objective: This study attempts to evaluate the knowledge, attitude, and clinical practice of Iraqi dentists toward CBCT use, referral patterns, observed challenges, and the impacts...
Conceptual design report of the MPD Cosmic Ray Detector (MCORD)
Conceptual design report of the MPD Cosmic Ray Detector (MCORD)
Abstract
This report presents a concept of constructing a detector
dedicated for detection of muons observed during measurements
carried out at the MPD (Multi-Pu...
Radiographic imaging in relation to the mandibular third molar: tooth characteristics, modality choice, optimization, and absorbed dose
Radiographic imaging in relation to the mandibular third molar: tooth characteristics, modality choice, optimization, and absorbed dose
Radiographic imaging of the mandibular third molar (M3) is a valuable diagnostic tool. It provides information on tooth position, root morphology, and relations to surrounding anat...
Impact of movement and motion-artefact correction on image quality and interpretability in CBCT units with aligned and lateral-offset detectors
Impact of movement and motion-artefact correction on image quality and interpretability in CBCT units with aligned and lateral-offset detectors
Objectives: To evaluate the impact of movement and motion-artefact correction systems on CBCT image quality and interpretability of simulated diagnostic tasks for aligned and later...