Validation of a Novel Noninvasive Technology to Estimate Blood Oxygen Saturation Using Green Light: Observational Study (Preprint)
View through CrossRef
BACKGROUND
Pulse oximeters work within the red-infrared wavelengths. Therefore, these oximeters produce erratic results in dark-skinned subjects and in subjects with cold extremities. Pulse oximetry is routinely performed in patients with fever; however, an elevation in body temperature decreases the affinity of hemoglobin for oxygen, causing a drop in oxygen saturation or oxyhemoglobin concentrations.
OBJECTIVE
We aimed to determine whether our new investigational device, the Shani device or SH1 (US Patent 11191460), detects a drop in oxygen saturation or a decrease in oxyhemoglobin concentrations.
METHODS
An observational study (phase 1) was performed in two separate groups to validate measurements of hemoglobin and oxygen concentrations, including 39 participants recruited among current university students and staff aged 20-40 years. All volunteers completed baseline readings using the SH1 device and the commercially available Food and Drug Administration–approved pulse oximeter Masimo. SH1 uses two light-emitting diodes in which the emitted wavelengths match with absorption peaks of oxyhemoglobin (hemoglobin combined with oxygen) and deoxyhemoglobin (hemoglobin without oxygen or reduced hemoglobin). Total hemoglobin was calculated as the sum of oxyhemoglobin and deoxyhemoglobin. Subsequently, 16 subjects completed the “heat jacket study” and the others completed the “blood donation study.” Masimo was consistently used on the finger for comparison. The melanin level was accounted for using the von Luschan skin color scale (VLS) and a specifically designed algorithm. We here focus on the results of the heat jacket study, in which the subject wore a double-layered heated jacket and pair of trousers including a network of polythene tubules along with an inlet and outlet. Warm water was circulated to increase the body temperature by 0.5-0.8 °C above the baseline body temperature. We expected a slight drop in oxyhemoglobin concentrations in the heating phase at the tissue level.
RESULTS
The mean age of the participants was 24.1 (SD 0.8) years. The skin tone varied from 12 to 36 on the VLS, representing a uniform distribution with one-third of the participants having fair skin, brown skin, and dark skin, respectively. Using a specific algorithm and software, the reflection ratio for oxyhemoglobin was displayed on the screen of the device along with direct hemoglobin values. The SH1 device picked up more minor changes in oxyhemoglobin levels after a change in body temperature compared to the pulse oximeter, with a maximum drop in oxyhemoglobin concentration detected of 6.5% and 2.54%, respectively.
CONCLUSIONS
Our new investigational device SH1 measures oxygen saturation at the tissue level by reflectance spectroscopy using green wavelengths. This device fared well regardless of skin color. This device can thus eliminate racial disparity in these key biomarker assessments. Moreover, since the light is shone on the wrist, SH1 can be readily miniaturized into a wearable device.
Title: Validation of a Novel Noninvasive Technology to Estimate Blood Oxygen Saturation Using Green Light: Observational Study (Preprint)
Description:
BACKGROUND
Pulse oximeters work within the red-infrared wavelengths.
Therefore, these oximeters produce erratic results in dark-skinned subjects and in subjects with cold extremities.
Pulse oximetry is routinely performed in patients with fever; however, an elevation in body temperature decreases the affinity of hemoglobin for oxygen, causing a drop in oxygen saturation or oxyhemoglobin concentrations.
OBJECTIVE
We aimed to determine whether our new investigational device, the Shani device or SH1 (US Patent 11191460), detects a drop in oxygen saturation or a decrease in oxyhemoglobin concentrations.
METHODS
An observational study (phase 1) was performed in two separate groups to validate measurements of hemoglobin and oxygen concentrations, including 39 participants recruited among current university students and staff aged 20-40 years.
All volunteers completed baseline readings using the SH1 device and the commercially available Food and Drug Administration–approved pulse oximeter Masimo.
SH1 uses two light-emitting diodes in which the emitted wavelengths match with absorption peaks of oxyhemoglobin (hemoglobin combined with oxygen) and deoxyhemoglobin (hemoglobin without oxygen or reduced hemoglobin).
Total hemoglobin was calculated as the sum of oxyhemoglobin and deoxyhemoglobin.
Subsequently, 16 subjects completed the “heat jacket study” and the others completed the “blood donation study.
” Masimo was consistently used on the finger for comparison.
The melanin level was accounted for using the von Luschan skin color scale (VLS) and a specifically designed algorithm.
We here focus on the results of the heat jacket study, in which the subject wore a double-layered heated jacket and pair of trousers including a network of polythene tubules along with an inlet and outlet.
Warm water was circulated to increase the body temperature by 0.
5-0.
8 °C above the baseline body temperature.
We expected a slight drop in oxyhemoglobin concentrations in the heating phase at the tissue level.
RESULTS
The mean age of the participants was 24.
1 (SD 0.
8) years.
The skin tone varied from 12 to 36 on the VLS, representing a uniform distribution with one-third of the participants having fair skin, brown skin, and dark skin, respectively.
Using a specific algorithm and software, the reflection ratio for oxyhemoglobin was displayed on the screen of the device along with direct hemoglobin values.
The SH1 device picked up more minor changes in oxyhemoglobin levels after a change in body temperature compared to the pulse oximeter, with a maximum drop in oxyhemoglobin concentration detected of 6.
5% and 2.
54%, respectively.
CONCLUSIONS
Our new investigational device SH1 measures oxygen saturation at the tissue level by reflectance spectroscopy using green wavelengths.
This device fared well regardless of skin color.
This device can thus eliminate racial disparity in these key biomarker assessments.
Moreover, since the light is shone on the wrist, SH1 can be readily miniaturized into a wearable device.
Related Results
[RETRACTED] Guardian Blood Balance –Feel the difference Guardian Blood Balance makes! v1
[RETRACTED] Guardian Blood Balance –Feel the difference Guardian Blood Balance makes! v1
[RETRACTED]Guardian Blood Balance Reviews (Works Or Hoax) Does Guardian Botanicals Blood Balance AU Really Works? Read Updated Report! Diabetes and Hypertension is such a health p...
Retinal Oximetry
Retinal Oximetry
Abstract.Purpose:Malfunction of retinal blood flow or oxygenation is believed to be involved in various diseases. Among them are retinal vessel occlusions, diabetic retinopathy and...
[RETRACTED] Guardian Blood Balance Australia- Reviews - Guardian Botanicals Blood Balance [AU] SCAM ALERT! Read Real Critical Reports.. Price in Australia v1
[RETRACTED] Guardian Blood Balance Australia- Reviews - Guardian Botanicals Blood Balance [AU] SCAM ALERT! Read Real Critical Reports.. Price in Australia v1
[RETRACTED]Guardian Blood Balance Australia Reviews - Diabetes and blood sugar are some of the common problems that are attacking so many adult individuals nowadays. Obesity is t...
High Concentration Oxygen and Hypercapnia in Respiratory Disease
High Concentration Oxygen and Hypercapnia in Respiratory Disease
<p>Oxygen-induced elevations in arterial carbon dioxide tension have been demonstrated in patients with chronic obstructive pulmonary disease (COPD), asthma, pneumonia, obesi...
Retinal oxygen saturation in health and disease
Retinal oxygen saturation in health and disease
AbstractPurpose Measurements of retinal vessel oxygen saturation can provide insight into the nature of several eye diseases.Methods Dual wavelength oximeters are used by our group...
0864 Severe Central Sleep Apnea
0864 Severe Central Sleep Apnea
Abstract
Introduction
Central sleep apnea (CSA) is a rare form of sleep disordered breathing with repeated apneic episodes with ...
Korelasi Kadar Karboksihemoglobin terhadap Tekanan Darah Penduduk di Sekitar Terminal Bus Tirtonadi Surakarta
Korelasi Kadar Karboksihemoglobin terhadap Tekanan Darah Penduduk di Sekitar Terminal Bus Tirtonadi Surakarta
<table width="645" border="1" cellspacing="0" cellpadding="0"><tbody><tr><td valign="top" width="408"><p> </p><p>Carbon monoxide is a gas ...
Validation in Doctoral Education: Exploring PhD Students’ Perceptions of Belonging to Scaffold Doctoral Identity Work
Validation in Doctoral Education: Exploring PhD Students’ Perceptions of Belonging to Scaffold Doctoral Identity Work
Aim/Purpose: The aim of this article is to make a case of the role of validation in doctoral education. The purpose is to detail findings from three studies which explore PhD stude...