Technical note: Energy dependence of the Gafchromic EBT4 film: Dose‐response curves for 70 kV, 6 MV, 6 MV FFF, 10 MV FFF, and 15 MV x‐ray beams

View through CrossRef
AbstractBackgroundEBT4 was newly released for radiotherapy quality assurance to improve the signal‐to‐noise ratio in radiochromic film dosimetry. It is important to know its dose‐response characteristics before its use in the clinic.PurposeThis study aims to investigate and compare the dose‐response curves of the Gafchromic EBT4 film for megavoltage and kilovoltage x‐ray beams with different dose levels, scanning spatial resolutions, and sizes of region of interest (ROI).MethodsEBT4 film (Lot#07052201) calibration strips (3.5 × 20 cm2) were exposed to a 10×10 cm2 open field at doses of 0, 63, 125, 500, 750, 1000 cGy using 6 MV photon beam. EBT4 film strips from the same lot were then exposed to each x‐ray beam (6 MV, 6 MV FFF, 10 MV FFF, 15 MV, and 70 kV) at six dose values (50, 100, 300, 600, 800, 1000 cGy). A full sheet (25 × 20 cm2) of EBT4 film was irradiated at each energy with 300 cGy for profile comparison with the treatment planning calculation. At two different spatial resolutions of 72 and 300 dpi, each film piece was scanned three consecutive times in the center of an Epson 10000XL flatbed scanner in 48‐bit color. The scanned images were analyzed using FilmQA Pro. For each scanned image, an ROI of 2 × 2 cm2 at the field center was selected to obtain the average pixel value with its standard deviation in the ROI. An additional ROI of 1 cm diameter circle was also used to evaluate the impact of ROI shape and size, especially for FFF beams. The dose value, average dose‐response value, and associated uncertainty were determined for each energy and relative responses were analyzed. The Student's t‐test was performed to evaluate the statistical significance of the dose‐response values with different color channels, ROI shapes, and spatial resolutions.ResultsThe dose‐response curves for the five x‐ray energies were compared in three color channels. Weak energy dependence was found among the megavoltage beams. No significant differences (average ∼1.1%) were observed for all doses in this study among 6 MV, 6 MV FFF, 10 MV FFF, and 15 MV beams, regardless of spatial resolution and color channel. However, a statistically significant difference in dose‐response was observed up to 12% between 70 kV and 6 MV beams.ConclusionsThe dose‐response curves for Gafchromic EBT4 films were nearly independent of the energy of the photon beams among 6 MV, 6 MV FFF, 10 MV FFF, and 15 MV. For very low‐energy photons (e.g., 70 kV), a separate calibration from the same low‐energy x‐ray is necessary.
Title: Technical note: Energy dependence of the Gafchromic EBT4 film: Dose‐response curves for 70 kV, 6 MV, 6 MV FFF, 10 MV FFF, and 15 MV x‐ray beams
Description:
AbstractBackgroundEBT4 was newly released for radiotherapy quality assurance to improve the signal‐to‐noise ratio in radiochromic film dosimetry.
It is important to know its dose‐response characteristics before its use in the clinic.
PurposeThis study aims to investigate and compare the dose‐response curves of the Gafchromic EBT4 film for megavoltage and kilovoltage x‐ray beams with different dose levels, scanning spatial resolutions, and sizes of region of interest (ROI).
MethodsEBT4 film (Lot#07052201) calibration strips (3.
5 × 20 cm2) were exposed to a 10×10 cm2 open field at doses of 0, 63, 125, 500, 750, 1000 cGy using 6 MV photon beam.
EBT4 film strips from the same lot were then exposed to each x‐ray beam (6 MV, 6 MV FFF, 10 MV FFF, 15 MV, and 70 kV) at six dose values (50, 100, 300, 600, 800, 1000 cGy).
A full sheet (25 × 20 cm2) of EBT4 film was irradiated at each energy with 300 cGy for profile comparison with the treatment planning calculation.
At two different spatial resolutions of 72 and 300 dpi, each film piece was scanned three consecutive times in the center of an Epson 10000XL flatbed scanner in 48‐bit color.
The scanned images were analyzed using FilmQA Pro.
For each scanned image, an ROI of 2 × 2 cm2 at the field center was selected to obtain the average pixel value with its standard deviation in the ROI.
An additional ROI of 1 cm diameter circle was also used to evaluate the impact of ROI shape and size, especially for FFF beams.
The dose value, average dose‐response value, and associated uncertainty were determined for each energy and relative responses were analyzed.
The Student's t‐test was performed to evaluate the statistical significance of the dose‐response values with different color channels, ROI shapes, and spatial resolutions.
ResultsThe dose‐response curves for the five x‐ray energies were compared in three color channels.
Weak energy dependence was found among the megavoltage beams.
No significant differences (average ∼1.
1%) were observed for all doses in this study among 6 MV, 6 MV FFF, 10 MV FFF, and 15 MV beams, regardless of spatial resolution and color channel.
However, a statistically significant difference in dose‐response was observed up to 12% between 70 kV and 6 MV beams.
ConclusionsThe dose‐response curves for Gafchromic EBT4 films were nearly independent of the energy of the photon beams among 6 MV, 6 MV FFF, 10 MV FFF, and 15 MV.
For very low‐energy photons (e.
g.
, 70 kV), a separate calibration from the same low‐energy x‐ray is necessary.

Related Results

Reclaiming the Wasteland: Samson and Delilah and the Historical Perception and Construction of Indigenous Knowledges in Australian Cinema
Reclaiming the Wasteland: Samson and Delilah and the Historical Perception and Construction of Indigenous Knowledges in Australian Cinema
It was always based on a teenage love story between the two kids. One is a sniffer and one is not. It was designed for Central Australia because we do write these kids off there. N...
Alternative Entrances: Phillip Noyce and Sydney’s Counterculture
Alternative Entrances: Phillip Noyce and Sydney’s Counterculture
Phillip Noyce is one of Australia’s most prominent film makers—a successful feature film director with both iconic Australian narratives and many a Hollywood blockbuster under his ...
Odd version Mathieu-Gaussian beam based on Green function
Odd version Mathieu-Gaussian beam based on Green function
Like the theoretical pattern of non-diffracting Bessel beams, ideal non-diffracting Mathieu beams also carry infinite energy, but cannot be generated as a physically realizable ent...
Ary Scheffer, een Nederlandse Fransman
Ary Scheffer, een Nederlandse Fransman
AbstractAry Scheffer (1795-1858) is so generally included in the French School (Note 2)- unsurprisingly, since his career was confined almost entirely to Paris - that the fact that...
Construction of Bi-Pearcey beams and their mathematical mechanism
Construction of Bi-Pearcey beams and their mathematical mechanism
We present a theoretical expression in the form of the Pearcey function by deducing the Fresnel diffraction distribution of an elliptic line. Then, we numerically simulate and expe...
Dynamic Characteristics Analysis of Three-Layer Steel–Concrete Composite Beams
Dynamic Characteristics Analysis of Three-Layer Steel–Concrete Composite Beams
The dynamic behavior of three-layer composite beams, consisting of concrete slabs and steel beams, is influenced by the structural configuration of each layer as well as the shear ...
Experimental study and computational analysis of structural performance of reinforced geopolymer concrete beams
Experimental study and computational analysis of structural performance of reinforced geopolymer concrete beams
PurposeIn this study, the aim is to explore the effects of geopolymer concrete (GPC) strength and reinforcement ratio on the flexural performance of reinforced GPC beams. Furthermo...

Back to Top