Evaluation of tube current modulation programms for the optimization of scan protocols in computed tomography
View through CrossRef
Computed tomography (CT) is commonly associated with relatively high patient doses. In order to keep the patient doses from the CT examinations on the acceptable level it is necessary to apply the principle of optimization. An essential part of optimization is the achievement of the compromise between the patient dose reduction and the maintenance of the image quality that provides accurate diagnostic information. The aim of the study was to determine the relations between the patient doses, CT image quality and the parameters of the tube current modulation program (Auto mA, CareDose and DoseRight) for the examination of the chest. The study was performed on the three most common modern tomographs in Russia: Ingenuity Core 128, Philips; Optima 64, General Electric; Definition AS, Siemens. The anthropomorphic phantom Lungman (Kyoto Kagaku CO., LTD) was used in the study. In order to assess the tube current modulation for the range of tube voltage (80–120 kV), the reference mA∙s (ref. mAs), Noise Index (NI) and Dose Right Index (DRI) were changed for the CareDose (Siemens), Auto mA (GE) and RightDose (Philips), respectively. Estimation of the effective dose was performed using the method from Methodical guidance 2.6.1.2944-11 (MU 2.6.1.2944-11). In order to evaluate the image quality, the noise of a CT image in the mediastinum was selected as the most homogenous for chest region. It was estimated, that for the GE units with Auto mA, the noise of CT image had a liner relationship with NI; the patient dose decreased with the increase of NI. For the Siemens units with CareDose, the noise of CT image decreased with the ref.mAs for the range of tube voltage (80–120 kV); the patient dose was directly proportional to the ref.mAs and increased with the tube voltage. For the Philips units with DoseRight, the noise of CT image decreased with the DRI for the range of tube voltage (80–120 kV); the patient dose was directly proportional to the DRI and had no dependence on the tube voltage. The tube current modulations are proprietary for each manufacturer; it is necessary to consider them for the protocol development. The obtained dependences could be useful for optimization of CT protocols.
SPRI of Radiation Hygiene Prof. PV Ramzaev
Title: Evaluation of tube current modulation programms for the optimization of scan protocols in computed tomography
Description:
Computed tomography (CT) is commonly associated with relatively high patient doses.
In order to keep the patient doses from the CT examinations on the acceptable level it is necessary to apply the principle of optimization.
An essential part of optimization is the achievement of the compromise between the patient dose reduction and the maintenance of the image quality that provides accurate diagnostic information.
The aim of the study was to determine the relations between the patient doses, CT image quality and the parameters of the tube current modulation program (Auto mA, CareDose and DoseRight) for the examination of the chest.
The study was performed on the three most common modern tomographs in Russia: Ingenuity Core 128, Philips; Optima 64, General Electric; Definition AS, Siemens.
The anthropomorphic phantom Lungman (Kyoto Kagaku CO.
, LTD) was used in the study.
In order to assess the tube current modulation for the range of tube voltage (80–120 kV), the reference mA∙s (ref.
mAs), Noise Index (NI) and Dose Right Index (DRI) were changed for the CareDose (Siemens), Auto mA (GE) and RightDose (Philips), respectively.
Estimation of the effective dose was performed using the method from Methodical guidance 2.
6.
1.
2944-11 (MU 2.
6.
1.
2944-11).
In order to evaluate the image quality, the noise of a CT image in the mediastinum was selected as the most homogenous for chest region.
It was estimated, that for the GE units with Auto mA, the noise of CT image had a liner relationship with NI; the patient dose decreased with the increase of NI.
For the Siemens units with CareDose, the noise of CT image decreased with the ref.
mAs for the range of tube voltage (80–120 kV); the patient dose was directly proportional to the ref.
mAs and increased with the tube voltage.
For the Philips units with DoseRight, the noise of CT image decreased with the DRI for the range of tube voltage (80–120 kV); the patient dose was directly proportional to the DRI and had no dependence on the tube voltage.
The tube current modulations are proprietary for each manufacturer; it is necessary to consider them for the protocol development.
The obtained dependences could be useful for optimization of CT protocols.
.
Related Results
Blunt Chest Trauma and Chylothorax: A Systematic Review
Blunt Chest Trauma and Chylothorax: A Systematic Review
Abstract
Introduction: Although traumatic chylothorax is predominantly associated with penetrating injuries, instances following blunt trauma, as a rare and challenging condition, ...
Environmental Surveillance Protocols for Highly Pathogenic Avian Influenza (HPAI) v2
Environmental Surveillance Protocols for Highly Pathogenic Avian Influenza (HPAI) v2
EnvironmentalSurveillance Protocols for Highly Pathogenic Avian Influenza (HPAI) This comprehensive protocol suite enables systematic environmental surveillance for avian influenza...
Lectin C gene analysis v1
Lectin C gene analysis v1
Mammalian Tissue Total RNA Purification Protocol by GeneJET RNA Purification Kit (Thermo Scientific, USA) Before starting: • Supplement the required amount of Lysis Buffer with β-...
Optimizing IETF multimedia signaling protocols and architectures in 3GPP networks : an evolutionary approach
Optimizing IETF multimedia signaling protocols and architectures in 3GPP networks : an evolutionary approach
Signaling in Next Generation IP-based networks heavily relies in the family of multimedia signaling protocols defined by IETF. Two of these signaling protocols are RTSP and SIP, wh...
Tube Shunt Related Complications in Pediatrics
Tube Shunt Related Complications in Pediatrics
Although medical therapy is usually an excellent therapeutic option in the adult population, in children it is often ineffective or associated with an undesirable risk:benefit rati...
Complex Collision Tumors: A Systematic Review
Complex Collision Tumors: A Systematic Review
Abstract
Introduction: A collision tumor consists of two distinct neoplastic components located within the same organ, separated by stromal tissue, without histological intermixing...
Characterization of surgically transposed ovaries in integrated PET/CT scan in patients with cervical cancer
Characterization of surgically transposed ovaries in integrated PET/CT scan in patients with cervical cancer
AbstractBackground. The purpose of this study was to determine the ovarian findings on integrated positron emission tomography/computed tomography scans during follow‐up in cervica...
J-Tube Design for Flexible Umbilicals
J-Tube Design for Flexible Umbilicals
ABSTRACT
The design of J-tubes with one or more bends under functional loads is dictated by a single force TL, the cable tension at the bottom of the J-tube. This...