Stapled Aortic Anastomoses: A Minimally Invasive, Feasible Alternative to Videoscopic Aortic Suturing?

View through CrossRef
Widespread applications of totally laparoscopic aortic reconstructions have been limited by the long cross-clamp time required to suture the aortic anastomosis despite improvement in instrumentation. The authors' hypothesis was that a “one-step anastomosis concept” using an intraluminal stapler would allow shorter cross-clamp time but similar patency and imperviousness as videoscopic suturing techniques. An intraluminal stapler (Endopath-ILS, Ethicon) with a modified anvil was used to perform videoscopic-assisted thoracic aorta-to-iliac artery bypass with a 21 mm by 8 mm polytetrafluoroethylene (PTFE) graft in 22 sheep through a minimally invasive approach using a 5 cm thoracotomy. The graft-to-iliac artery anastomoses were hand sutured through a flank incision. Twelve sheep were used to establish the technique and 10 subsequent animals constituted the study group. Aortic cross-clamp time, imperviousness, and need for additional sutures were recorded and compared to previously reported data using videoscopic suturing in pigs. Patency was assessed by comparing lower limb arterial pressures. Macroscopic and microscopic examinations of the anastomoses were performed at different time-points within the first 3 months. Videoscopic-assisted stapled anastomoses were also performed on atherosclerotic aortas of 3 human cadavers. Stapled anastomoses between the thoracic aorta and PTFE graft were completed in 8 of 10 animals. Two animals were euthanized after stapler failure and anastomotic bleeding. Sutures to strengthen the anastomosis had to be used in 4 cases. Mean aortic cross-clamp time in 8 successful cases was 4.3 ±2.9 minutes (range 2–11 minutes) and was significantly shorter than clamp time of videoscopic suturing technique (48.7 ±9.4 minutes, p<0.0001). Imperviousness was good or excellent in 4 animals and fair in 4 animals. All anastomoses were patent at the end of the procedure. Examination of the anastomosis of the 2 failed interventions showed medial aortic tear surrounding the anastomosis in 1 case and misfired staples in the other. No graft occlusion was noted during follow-up ranging from 0 to 12 weeks. At the time of harvest, no bleeding was noted after epinephrine and volume infusion to increase mean arterial pressure to 200 mm Hg for 15 minutes. Macroscopic examination of the anastomoses revealed adequate healing with circumferential stapling of the prosthesis to the aortic wall and no stenosis or thrombus except in 1 false aneurysm (1/7, 14%). Surface electron microscopy showed cells coverage of the anastomosis surface. When applied on human cadaver thoracic and abdominal aorta with atherosclerotic changes, clamping times of less than 5 minutes were achieved. However, imperviousness tested with saline was poor. An automatic stapling device allows performance of a graft-to-aorta anastomosis through a minimally invasive approach with shorter clamping time than a videoscopic suturing technique. However, the current technique of aortic stapling is unreliable and further improvements are needed.
Title: Stapled Aortic Anastomoses: A Minimally Invasive, Feasible Alternative to Videoscopic Aortic Suturing?
Description:
Widespread applications of totally laparoscopic aortic reconstructions have been limited by the long cross-clamp time required to suture the aortic anastomosis despite improvement in instrumentation.
The authors' hypothesis was that a “one-step anastomosis concept” using an intraluminal stapler would allow shorter cross-clamp time but similar patency and imperviousness as videoscopic suturing techniques.
An intraluminal stapler (Endopath-ILS, Ethicon) with a modified anvil was used to perform videoscopic-assisted thoracic aorta-to-iliac artery bypass with a 21 mm by 8 mm polytetrafluoroethylene (PTFE) graft in 22 sheep through a minimally invasive approach using a 5 cm thoracotomy.
The graft-to-iliac artery anastomoses were hand sutured through a flank incision.
Twelve sheep were used to establish the technique and 10 subsequent animals constituted the study group.
Aortic cross-clamp time, imperviousness, and need for additional sutures were recorded and compared to previously reported data using videoscopic suturing in pigs.
Patency was assessed by comparing lower limb arterial pressures.
Macroscopic and microscopic examinations of the anastomoses were performed at different time-points within the first 3 months.
Videoscopic-assisted stapled anastomoses were also performed on atherosclerotic aortas of 3 human cadavers.
Stapled anastomoses between the thoracic aorta and PTFE graft were completed in 8 of 10 animals.
Two animals were euthanized after stapler failure and anastomotic bleeding.
Sutures to strengthen the anastomosis had to be used in 4 cases.
Mean aortic cross-clamp time in 8 successful cases was 4.
3 ±2.
9 minutes (range 2–11 minutes) and was significantly shorter than clamp time of videoscopic suturing technique (48.
7 ±9.
4 minutes, p<0.
0001).
Imperviousness was good or excellent in 4 animals and fair in 4 animals.
All anastomoses were patent at the end of the procedure.
Examination of the anastomosis of the 2 failed interventions showed medial aortic tear surrounding the anastomosis in 1 case and misfired staples in the other.
No graft occlusion was noted during follow-up ranging from 0 to 12 weeks.
At the time of harvest, no bleeding was noted after epinephrine and volume infusion to increase mean arterial pressure to 200 mm Hg for 15 minutes.
Macroscopic examination of the anastomoses revealed adequate healing with circumferential stapling of the prosthesis to the aortic wall and no stenosis or thrombus except in 1 false aneurysm (1/7, 14%).
Surface electron microscopy showed cells coverage of the anastomosis surface.
When applied on human cadaver thoracic and abdominal aorta with atherosclerotic changes, clamping times of less than 5 minutes were achieved.
However, imperviousness tested with saline was poor.
An automatic stapling device allows performance of a graft-to-aorta anastomosis through a minimally invasive approach with shorter clamping time than a videoscopic suturing technique.
However, the current technique of aortic stapling is unreliable and further improvements are needed.

Related Results

AORTIC ROOT ANEURYSM OR ECTASIA TREATED WITH AORTIC ROOT WRAPPED PROCEDURE
AORTIC ROOT ANEURYSM OR ECTASIA TREATED WITH AORTIC ROOT WRAPPED PROCEDURE
Objectives To develop a relatively simple and effective and less risk operation, aortic root wrapped procedure, to treat with aortic root aneurysm or ectasia. ...
Evaluation of the MMI Symani® Robotic Microsurgical System for coronary-bypass anastomoses in a cadaveric porcine model
Evaluation of the MMI Symani® Robotic Microsurgical System for coronary-bypass anastomoses in a cadaveric porcine model
Abstract Objective The MMI Symani® is a recently approved robotic microsurgical system for surgical procedures in adults. The system enables the surgeon to create microanas...
Breast Carcinoma within Fibroadenoma: A Systematic Review
Breast Carcinoma within Fibroadenoma: A Systematic Review
Abstract Introduction Fibroadenoma is the most common benign breast lesion; however, it carries a potential risk of malignant transformation. This systematic review provides an ove...
Diseases of the Aorta
Diseases of the Aorta
This review covers the major presentations affecting the aorta: aortic aneurysms (abdominal aortic aneurysms and thoracic aortic aneurysms), acute aortic syndromes (including aorti...
Dissecting the geometric and hydrophobic constraints of stapled peptides
Dissecting the geometric and hydrophobic constraints of stapled peptides
Stapled peptides are a promising class of molecules with potential as highly specific probes of protein-protein interactions and as therapeutics. Hydrocarbon stapling affects the p...
Technique‐associated outcomes in horses following large colon resection
Technique‐associated outcomes in horses following large colon resection
AbstractObjectiveTo compare survival and complications in horses undergoing large colon resection with either sutured end‐to‐end or stapled functional end‐to‐end anastomoses.Study ...
Twin anemia polycythemia sequence after fetoscopy: predictors of development
Twin anemia polycythemia sequence after fetoscopy: predictors of development
Objective. To identify predictors of post-fetoscopic twin anemia polycythemia sequence (pTAPS) after fetoscopic laser photocoagulation (FLP) of placental anastomoses for twin–twin ...

Back to Top