Search engine for discovering works of Art, research articles, and books related to Art and Culture
ShareThis
Javascript must be enabled to continue!

Heart Rate Variability from Wearables: A Comparative Analysis Among Standard ECG, a Smart Shirt and a Wristband

View through CrossRef
Heart Rate Variability (HRV) derived from standard one-lead electrocardiography (ECG) was compared with HRV computed by a commercial ECG shirt and with the inter-beat-intervals (IBI) measured by a research-grade photoplethysmographic (PPG) wristband. Signals from 8 subjects were recorded in three experimental phases: during sit; in upright position (“stand”); during controlled respiration. HRV and IBI from both the wearables were computed online (i.e. during the experiment) and stored for subsequent analyses, while the standard ECG was processed offline with state-of-the-art methods to obtain a clean reference HRV. Shirt and wristband signals accuracies were assessed, with respect to the reference HRV, through a between-phase and a beat-to-beat analyses. The former considered several time and frequency domain parameters; the latter was carried out through the Bland-Altman method. Time and frequency domain parameters computed from shirt HRV resulted very similar to the ones extracted from the reference HRV and generally more accurate than the parameters computed from wristband IBI. The Bland-Altman analysis showed that wristband IBI is significantly different from ECG-derived HRV, especially during “stand”. Therefore, our results support the idea that some care should be paid in the interpretation of online PPG-derived IBI, while HRV measures online-derived from ECG-shirts seem to be more reliable in the analyzed conditions. The high number of missing beats also suggest that the design of wristband devices should be taken into account to reduce the rate of incorrect measurements, by maximizing sensor adhesion to the skin.
Title: Heart Rate Variability from Wearables: A Comparative Analysis Among Standard ECG, a Smart Shirt and a Wristband
Description:
Heart Rate Variability (HRV) derived from standard one-lead electrocardiography (ECG) was compared with HRV computed by a commercial ECG shirt and with the inter-beat-intervals (IBI) measured by a research-grade photoplethysmographic (PPG) wristband.
Signals from 8 subjects were recorded in three experimental phases: during sit; in upright position (“stand”); during controlled respiration.
HRV and IBI from both the wearables were computed online (i.
e.
during the experiment) and stored for subsequent analyses, while the standard ECG was processed offline with state-of-the-art methods to obtain a clean reference HRV.
Shirt and wristband signals accuracies were assessed, with respect to the reference HRV, through a between-phase and a beat-to-beat analyses.
The former considered several time and frequency domain parameters; the latter was carried out through the Bland-Altman method.
Time and frequency domain parameters computed from shirt HRV resulted very similar to the ones extracted from the reference HRV and generally more accurate than the parameters computed from wristband IBI.
The Bland-Altman analysis showed that wristband IBI is significantly different from ECG-derived HRV, especially during “stand”.
Therefore, our results support the idea that some care should be paid in the interpretation of online PPG-derived IBI, while HRV measures online-derived from ECG-shirts seem to be more reliable in the analyzed conditions.
The high number of missing beats also suggest that the design of wristband devices should be taken into account to reduce the rate of incorrect measurements, by maximizing sensor adhesion to the skin.

Related Results

Primerjalna književnost na prelomu tisočletja
Primerjalna književnost na prelomu tisočletja
In a comprehensive and at times critical manner, this volume seeks to shed light on the development of events in Western (i.e., European and North American) comparative literature ...
Inter-beat and intra-beat ECG interval analysis based on state space and hidden markov models
Inter-beat and intra-beat ECG interval analysis based on state space and hidden markov models
Analyse des intervalles ECG inter- et intra-battement sur des modèles d'espace d'état et de Markov cachés Les maladies cardiovasculaires sont l'une des principales ...
Complex Deep Learning Models for Denoising of Human Heart ECG signals
Complex Deep Learning Models for Denoising of Human Heart ECG signals
Effective and powerful methods for denoising real electrocardiogram (ECG) signals are important for wearable sensors and devices. Deep Learning (DL) models have been used extensive...
How Can We Identify the Best Implantation Site for an ECG Event Recorder?
How Can We Identify the Best Implantation Site for an ECG Event Recorder?
ZELLERHOFF, C., et al.: How Can We Identify the Best Implantation Site for an ECG Event Recorder? The aim of this study was to show how to find the preferable implantation site for...
Development of AI-based method to detect the subtle ECG deviations from the population ECG norm
Development of AI-based method to detect the subtle ECG deviations from the population ECG norm
Abstract Funding Acknowledgements Type of funding sources: Public grant(s) – National budget only. Main funding source(s): Natio...
Abstract P821: A New Era for Cerebral Angiograms: TransRadial Approach in NeuroInterventional Radiology
Abstract P821: A New Era for Cerebral Angiograms: TransRadial Approach in NeuroInterventional Radiology
Background: Historically, femoral artery access has been the standard for neurointerventional radiology (NIR) procedures. However, recent advances in cerebral angiograp...

Back to Top