Heart Rate Variability from Wearables: A Comparative Analysis Among Standard ECG, a Smart Shirt and a Wristband
View through CrossRef
Heart Rate Variability (HRV) derived from standard one-lead electrocardiography (ECG) was compared with HRV computed by a commercial ECG shirt and with the inter-beat-intervals (IBI) measured by a research-grade photoplethysmographic (PPG) wristband. Signals from 8 subjects were recorded in three experimental phases: during sit; in upright position (“stand”); during controlled respiration. HRV and IBI from both the wearables were computed online (i.e. during the experiment) and stored for subsequent analyses, while the standard ECG was processed offline with state-of-the-art methods to obtain a clean reference HRV. Shirt and wristband signals accuracies were assessed, with respect to the reference HRV, through a between-phase and a beat-to-beat analyses. The former considered several time and frequency domain parameters; the latter was carried out through the Bland-Altman method. Time and frequency domain parameters computed from shirt HRV resulted very similar to the ones extracted from the reference HRV and generally more accurate than the parameters computed from wristband IBI. The Bland-Altman analysis showed that wristband IBI is significantly different from ECG-derived HRV, especially during “stand”. Therefore, our results support the idea that some care should be paid in the interpretation of online PPG-derived IBI, while HRV measures online-derived from ECG-shirts seem to be more reliable in the analyzed conditions. The high number of missing beats also suggest that the design of wristband devices should be taken into account to reduce the rate of incorrect measurements, by maximizing sensor adhesion to the skin.
Title: Heart Rate Variability from Wearables: A Comparative Analysis Among Standard ECG, a Smart Shirt and a Wristband
Description:
Heart Rate Variability (HRV) derived from standard one-lead electrocardiography (ECG) was compared with HRV computed by a commercial ECG shirt and with the inter-beat-intervals (IBI) measured by a research-grade photoplethysmographic (PPG) wristband.
Signals from 8 subjects were recorded in three experimental phases: during sit; in upright position (“stand”); during controlled respiration.
HRV and IBI from both the wearables were computed online (i.
e.
during the experiment) and stored for subsequent analyses, while the standard ECG was processed offline with state-of-the-art methods to obtain a clean reference HRV.
Shirt and wristband signals accuracies were assessed, with respect to the reference HRV, through a between-phase and a beat-to-beat analyses.
The former considered several time and frequency domain parameters; the latter was carried out through the Bland-Altman method.
Time and frequency domain parameters computed from shirt HRV resulted very similar to the ones extracted from the reference HRV and generally more accurate than the parameters computed from wristband IBI.
The Bland-Altman analysis showed that wristband IBI is significantly different from ECG-derived HRV, especially during “stand”.
Therefore, our results support the idea that some care should be paid in the interpretation of online PPG-derived IBI, while HRV measures online-derived from ECG-shirts seem to be more reliable in the analyzed conditions.
The high number of missing beats also suggest that the design of wristband devices should be taken into account to reduce the rate of incorrect measurements, by maximizing sensor adhesion to the skin.
Related Results
Primerjalna književnost na prelomu tisočletja
Primerjalna književnost na prelomu tisočletja
In a comprehensive and at times critical manner, this volume seeks to shed light on the development of events in Western (i.e., European and North American) comparative literature ...
Improvement in transfusion safety using a specially designed transfusion wristband
Improvement in transfusion safety using a specially designed transfusion wristband
Fatal haemolytic transfusion reaction due to ABO incompatibility occurs mainly as a result of clerical error. A blood sample drawn from the wrong patient and labelled as another pa...
Complex Deep Learning Models for Denoising of Human Heart ECG signals
Complex Deep Learning Models for Denoising of Human Heart ECG signals
Effective and powerful methods for denoising real electrocardiogram (ECG) signals are important for wearable sensors and devices. Deep Learning (DL) models have been used extensive...
Development of AI-based method to detect the subtle ECG deviations from the population ECG norm
Development of AI-based method to detect the subtle ECG deviations from the population ECG norm
Abstract
Funding Acknowledgements
Type of funding sources: Public grant(s) – National budget only. Main funding source(s): Natio...
Multichannel ECG Recording from Waist using Textile Sensors
Multichannel ECG Recording from Waist using Textile Sensors
Abstract
Background: The development of wearable health monitoring systems is garnering tremendous interest in research, technology and commercial applications. Their abili...
Heart failure monitoring using the wearable cardioverter defibrillator in patients with newly diagnosed heart failure
Heart failure monitoring using the wearable cardioverter defibrillator in patients with newly diagnosed heart failure
Abstract
Funding Acknowledgements
Type of funding sources: None.
Background
...
DCAE-SR: Design of a Denoising Convolutional Autoencoder for reconstructing Electrocardiograms signals at Super Resolution
DCAE-SR: Design of a Denoising Convolutional Autoencoder for reconstructing Electrocardiograms signals at Super Resolution
AbstractElectrocardiogram (ECG) signals play a pivotal role in cardiovascular diagnostics, providing essential information on the electrical activity of the heart. However, the inh...
Abnormal cardiomyocytes electrophysiology in patients with DM
Abnormal cardiomyocytes electrophysiology in patients with DM
Introduction:. ECG is used for routine evaluation heart function which depends from electrolytes content, especially Calcium and Potassium. In patients with DM ECG not always adequ...