An in vivo comparative evaluation to determine the accuracy of working length between radiographic and electronic apex locators
View through CrossRef
Background:
An
in vivo
comparative evaluation to determine the accuracy of working length between radiographic and electronic apex locators.
Aim:
The study was aimed at evaluating the accuracy of electronic apex locator, to determine the working length of root canal, and to compare it with the radiographic method of working length determination.
Materials and Methods:
A total of 20 teeth selected for the study had to go for extraction because of periodontal or orthodontic reasons. Access cavity was prepared and the clinical estimated working length (CEWL) was determined with 10-25 no. K-file. A radiograph was then taken for determining the radiographic estimated working length (REWL). For electronic measurement of root canal, a 10 no. K-file was advanced toward the apex until it reached a 0.5 mm short of apex as shown by the apex locator. After fixing the file with a light cured composite, the tooth was extracted, the tooth surface was then longitudinally grounded using straight fissure diamond bur until the root canal and the tip of the file were visible. The distance of file from the minor constriction was measured with help of stereomicroscope.
Statistical analysis:
The chi-square test was used for statistical analysis for this study.
Results:
The chi-square test where χ
2
= 21.034 with
P
= 0.000 indicated that a significant difference exists among the groups. The electronic method showed highest number of cases with the working length at the minor constrictor.
Conclusion:
The electronic method for determining the working length of root canal was found to be more accurate than the radiographic method.
Ovid Technologies (Wolters Kluwer Health)
Title: An in vivo comparative evaluation to determine the accuracy of working length between radiographic and electronic apex locators
Description:
Background:
An
in vivo
comparative evaluation to determine the accuracy of working length between radiographic and electronic apex locators.
Aim:
The study was aimed at evaluating the accuracy of electronic apex locator, to determine the working length of root canal, and to compare it with the radiographic method of working length determination.
Materials and Methods:
A total of 20 teeth selected for the study had to go for extraction because of periodontal or orthodontic reasons.
Access cavity was prepared and the clinical estimated working length (CEWL) was determined with 10-25 no.
K-file.
A radiograph was then taken for determining the radiographic estimated working length (REWL).
For electronic measurement of root canal, a 10 no.
K-file was advanced toward the apex until it reached a 0.
5 mm short of apex as shown by the apex locator.
After fixing the file with a light cured composite, the tooth was extracted, the tooth surface was then longitudinally grounded using straight fissure diamond bur until the root canal and the tip of the file were visible.
The distance of file from the minor constriction was measured with help of stereomicroscope.
Statistical analysis:
The chi-square test was used for statistical analysis for this study.
Results:
The chi-square test where χ
2
= 21.
034 with
P
= 0.
000 indicated that a significant difference exists among the groups.
The electronic method showed highest number of cases with the working length at the minor constrictor.
Conclusion:
The electronic method for determining the working length of root canal was found to be more accurate than the radiographic method.
Related Results
Comparative clinical assessment of working length endo-motor apex locator versus radiographic method in endo-dontic therapy.
Comparative clinical assessment of working length endo-motor apex locator versus radiographic method in endo-dontic therapy.
Introduction: The technological and therapeutic advancements suggests that, choosing an appropriate endodontic approach is the key parameter to mitigate over radiation exposure and...
In vitro evaluation of the accuracy of the fourth generation electronic apex locator: Apex ID
In vitro evaluation of the accuracy of the fourth generation electronic apex locator: Apex ID
Background: To achieve success in endodontic treatment, the working length must be precisely defined and maintained during preparation. The introduction of the electronic apex loca...
Accuracy evaluation of three electronic apex locators in teeth with immature apices
Accuracy evaluation of three electronic apex locators in teeth with immature apices
The aim of this randomized controlled study was to evaluate the accuracy of Root ZX, iPex and YS-RZ-A apex locators in tooth with incomplete apices. Material and methods: Thirty hu...
Primerjalna književnost na prelomu tisočletja
Primerjalna književnost na prelomu tisočletja
In a comprehensive and at times critical manner, this volume seeks to shed light on the development of events in Western (i.e., European and North American) comparative literature ...
OSIRIS-APEX: NASA's Apophis Explorer Mission
OSIRIS-APEX: NASA's Apophis Explorer Mission
Introduction: NASA’s Origins, Spectral Interpretation, Resource Identification, and Security– Regolith Explorer (OSIRIS–REx) mission characterized and collected a sample from aster...
Accuracy of Different Apex Locaters. (A Comparative in Vitro Study)
Accuracy of Different Apex Locaters. (A Comparative in Vitro Study)
Aims: The objective of this in vitro study was to compare the accuracy of the Root-ZX II, Root ZX Mini,i Pex II, Kerr apex ID , Root PI and YD2 electronic apex locators (EALs) in l...
Ex vivo Accuracy of Root Zx Ii, Root Zx Mini and RomiApex A-15 Apex Locators in Extracted Vital Pulp Teeth
Ex vivo Accuracy of Root Zx Ii, Root Zx Mini and RomiApex A-15 Apex Locators in Extracted Vital Pulp Teeth
ABSTRACT
The objective of this study was to compare, ex vivo, the accuracy of three electronic apex locators (EALs), Root ZX II, Root ZX Mini and RomiApex A-15, in detecting the ap...
Effectiveness of an electronic apex locator used after preflaring of cervical and middle third
Effectiveness of an electronic apex locator used after preflaring of cervical and middle third
The electronic method has been studied and improved aiming to add precision, speed and reliability of the measurement technique to determine the exact location of the working lengt...