Javascript must be enabled to continue!
MR (Diffusion-Weighted Imaging (DWI) of the Apparent Diffusion Coefficient (ADC), Clinically Feasible Profile
View through CrossRef
The goal of a QIBA Profile is to help achieve a useful level of performance for a given biomarker. The Claim (Section 2) describes the biomarker performance and is derived from the body of scientific literature meeting specific requirements, in particular test-retest studies. The Activities (Section 3) contribute to generating the biomarker. Requirements are placed on the Actors that participate in those activities as necessary to achieve the Claim. Assessment Procedures (Section 4) for evaluating specific requirements are defined as needed to ensure acceptable performance. Diffusion-Weighted Imaging (DWI) and the Apparent Diffusion Coefficient (ADC) are being used clinically as qualitative (DWI) and quantitative (ADC) indicators of disease presence, progression or response to treatment. Use of ADC as a robust quantitative biomarker with finite confidence intervals places additional requirements on Sites, Acquisition Devices and Protocols, Field Engineers, Scanner Operators (MR Technologists, Radiologists, Physicists and other Scientists), Image Analysts, Reconstruction Software and Image Analysis Tools. Additionally, due to the intrinsic dependence of measured ADC values on biophysical tissue properties, both the Profile Claims and the associated scan protocols (Section 3.6.2) are organ-specific. All of these are considered Actors involved in Activities of Acquisition Device Pre-delivery and Installation, Subject Handling, Image Data Acquisition, Reconstruction, Registration, ADC map generation, Quality Assurance (QA), Distribution, Analysis, and Interpretation. The requirements addressed in this Profile are focused on achieving ADC values with minimal systematic bias and measurement variability. DISCLAIMER: Technical performance of the MRI system can be assessed using a phantom having known diffusion properties, such as the QIBA DWI phantom. The clinical performance target is to achieve a 95% confidence interval for measurement of ADC with a variable precision depending on the organ being imaged and assuming adequate technical performance requirements are met. While in vivo DWI/ADC measurements have been performed throughout the human body, this Profile focused on four organ systems, namely brain, liver, prostate, and breast as having high clinical utilization of ADC with a sufficient level of statistical evidence to support the Profile Claims derived from the current peer-reviewed literature. In due time, new DWI technologies with proven greater performance levels, as well as more organ systems will be incorporated in future Profiles. This document is intended to help a variety of users: clinicians using this biomarker to aid patient management; imaging staff generating this biomarker; MRI system architects developing related products; purchasers of such products; and investigators designing clinical trials utilizing quantitative diffusion-based imaging endpoints. Note that this document only states requirements specific to DWI to achieve the claim, not requirements that pertain to clinical standard of care. Conforming to this Profile is secondary to proper patient care.
Radiological Society of North America (RSNA) / Quantitative Imaging Biomarkers Alliance (QIBA)
Title: MR (Diffusion-Weighted Imaging (DWI) of the Apparent Diffusion Coefficient (ADC), Clinically Feasible Profile
Description:
The goal of a QIBA Profile is to help achieve a useful level of performance for a given biomarker.
The Claim (Section 2) describes the biomarker performance and is derived from the body of scientific literature meeting specific requirements, in particular test-retest studies.
The Activities (Section 3) contribute to generating the biomarker.
Requirements are placed on the Actors that participate in those activities as necessary to achieve the Claim.
Assessment Procedures (Section 4) for evaluating specific requirements are defined as needed to ensure acceptable performance.
Diffusion-Weighted Imaging (DWI) and the Apparent Diffusion Coefficient (ADC) are being used clinically as qualitative (DWI) and quantitative (ADC) indicators of disease presence, progression or response to treatment.
Use of ADC as a robust quantitative biomarker with finite confidence intervals places additional requirements on Sites, Acquisition Devices and Protocols, Field Engineers, Scanner Operators (MR Technologists, Radiologists, Physicists and other Scientists), Image Analysts, Reconstruction Software and Image Analysis Tools.
Additionally, due to the intrinsic dependence of measured ADC values on biophysical tissue properties, both the Profile Claims and the associated scan protocols (Section 3.
6.
2) are organ-specific.
All of these are considered Actors involved in Activities of Acquisition Device Pre-delivery and Installation, Subject Handling, Image Data Acquisition, Reconstruction, Registration, ADC map generation, Quality Assurance (QA), Distribution, Analysis, and Interpretation.
The requirements addressed in this Profile are focused on achieving ADC values with minimal systematic bias and measurement variability.
DISCLAIMER: Technical performance of the MRI system can be assessed using a phantom having known diffusion properties, such as the QIBA DWI phantom.
The clinical performance target is to achieve a 95% confidence interval for measurement of ADC with a variable precision depending on the organ being imaged and assuming adequate technical performance requirements are met.
While in vivo DWI/ADC measurements have been performed throughout the human body, this Profile focused on four organ systems, namely brain, liver, prostate, and breast as having high clinical utilization of ADC with a sufficient level of statistical evidence to support the Profile Claims derived from the current peer-reviewed literature.
In due time, new DWI technologies with proven greater performance levels, as well as more organ systems will be incorporated in future Profiles.
This document is intended to help a variety of users: clinicians using this biomarker to aid patient management; imaging staff generating this biomarker; MRI system architects developing related products; purchasers of such products; and investigators designing clinical trials utilizing quantitative diffusion-based imaging endpoints.
Note that this document only states requirements specific to DWI to achieve the claim, not requirements that pertain to clinical standard of care.
Conforming to this Profile is secondary to proper patient care.
Related Results
Application of Diffusion-Weighted Imaging in the Detection of Active Sacroiliitis and the Comparison of Apparent Diffusion Coefficient and Relative Apparent Diffusion Coefficient Values
Application of Diffusion-Weighted Imaging in the Detection of Active Sacroiliitis and the Comparison of Apparent Diffusion Coefficient and Relative Apparent Diffusion Coefficient Values
Objectives: This study aims to evaluate the role of diffusion-weighted imaging in detection of active sacroiliitis and compare the apparent diffusion coefficient (ADC) and normaliz...
Abstract PS05-08: Restriction Spectrum Imaging MRI for automated evaluation of response to neoadjuvant therapy in breast cancer
Abstract PS05-08: Restriction Spectrum Imaging MRI for automated evaluation of response to neoadjuvant therapy in breast cancer
Abstract
Introduction:
Dynamic contrast-enhanced MRI (DCE) is currently used to evaluate neoadjuvant therapy response of breast cancer (1). However, D...
Characteristics and Role of Apparent Diffusion Coefficient (ADC) Value in Musculoskeletal Soft Tissue Tumor
Characteristics and Role of Apparent Diffusion Coefficient (ADC) Value in Musculoskeletal Soft Tissue Tumor
Introduction: Musculoskeletal soft tissue tumors (STTs) consist of a group of malignant and benign tumors that present challenges in diagnosis and therapy. Adding diffusion-weighte...
MRI with diffusion‐weighted imaging (DWI) and apparent diffusion coefficient (ADC) assessment in the evaluation of normal and abnormal fetal kidneys: preliminary experience
MRI with diffusion‐weighted imaging (DWI) and apparent diffusion coefficient (ADC) assessment in the evaluation of normal and abnormal fetal kidneys: preliminary experience
AbstractObjectiveTo investigate apparent diffusion coefficient (ADC) mapping with the measurement of renal tissue ADC value of normal and pathological fetal kidneys at various gest...
Prognostic value of apparent diffusion coefficient in neuroendocrine carcinomas of the uterine cervix
Prognostic value of apparent diffusion coefficient in neuroendocrine carcinomas of the uterine cervix
Objectives
This research was designed to examine the associations between the apparent diffusion coefficient (ADC) values and clinicopathological parameters, an...
Diagnostic Utility of Diffusion-Weighted Magnetic Resonance Imaging in Differentiating Benign and Malignant Focal Liver Lesions
Diagnostic Utility of Diffusion-Weighted Magnetic Resonance Imaging in Differentiating Benign and Malignant Focal Liver Lesions
Background: Accurate characterization of focal liver lesions (FLLs) is essential for appropriate staging in oncologic patients and to avoid unnecessary invasive procedures in benig...
Reproducibility of measurement of apparent diffusion coefficients of malignant hepatic tumors: Effect of DWI techniques and calculation methods
Reproducibility of measurement of apparent diffusion coefficients of malignant hepatic tumors: Effect of DWI techniques and calculation methods
AbstractPurpose:To evaluate the effect of diffusion‐weighted imaging (DWI) methods, apparent diffusion coefficient (ADC) calculation methods, and selection of b‐values on the ADCs ...
Role of diffusion weighted MRI imaging in characterization of hepatic focal lesions in cirrhotic patients in comparison to dynamic cross sectional imaging
Role of diffusion weighted MRI imaging in characterization of hepatic focal lesions in cirrhotic patients in comparison to dynamic cross sectional imaging
Abstract
Aim of work
The aim of the study is to show the feasibility of diffusion weighted (DWI) MRI with calculation of the app...

