The impact of AI correction on Centiloid in a same subject visual read comparison of 18 F‐Florbetapir and 18 F‐NAV4694 Aβ PET

View through CrossRef
Abstract Background Visual read of Aβ PET remains standard clinical practice. 18 F‐NAV4694 (NAV) has high affinity for Aβ and low non‐specific binding so may be more accurate than older Aβ tracers. We compared visual read of NAV to 18 F‐Florbetapir (FBP) PET using standard and AI corrected Centiloid (CL) thresholds as gold standard. Method 150 participants (71.3±6.0 years) in the AIBL study underwent both scans within 2 years. Scans were read by 6 nuclear medicine physicians blinded to CL. Reader performance was compared against various CL thresholds. CL was measured with CapAIBL using a) the standard method, b) a composite cerebellum plus hemispheric white matter reference region for FBP and c) with an AI correction method for the CL SUVR (DeepSUVR). The DeepSUVR correction was learnt from a separate dataset of longitudinal scans where unexpected temporal changes were penalized and a tracer specific SUVR correction factor was developed for application to single scans. Result CL frequency distribution graphs differed for NAV and FBP. Applying the DeepSUVR correction improved the correlation (R 2 ) between NAV and FBP CL from 0.82 to 0.94, predominantly by boosting many mid‐range FBP CL values and reducing FBP variance around zero CL (Figure 1). The FBP composite reference region had a similar though less marked effect. The thresholds that produced peak visual read accuracy with standard and DeepSUVR CL methods were 15CL and 11CL for NAV, and 29CL and 21CL for FBP (Figure 2). Mean sensitivity at peak accuracy thresholds was higher in NAV with 97% for standard and 96% for DeepSUVR methods compared to 93% for both methods in FBP. Mean specificity at the same thresholds was similar with 95% for NAV and FBP with standard CL while DeepSUVR produced a specificity of 97% for both tracers. Conclusion Visual read of amyloid PET can accurately detect levels of amyloid below the traditional thresholds of 20 or 25CL for a positive scan. Visual read thresholds vary by tracer and reader experience. AI correction of CL with the DeepSUVR method significantly improves the performance of Florbetapir but NAV4694 remains more sensitive to low amyloid levels.
Title: The impact of AI correction on Centiloid in a same subject visual read comparison of 18 F‐Florbetapir and 18 F‐NAV4694 Aβ PET
Description:
Abstract Background Visual read of Aβ PET remains standard clinical practice.
18 F‐NAV4694 (NAV) has high affinity for Aβ and low non‐specific binding so may be more accurate than older Aβ tracers.
We compared visual read of NAV to 18 F‐Florbetapir (FBP) PET using standard and AI corrected Centiloid (CL) thresholds as gold standard.
Method 150 participants (71.
3±6.
0 years) in the AIBL study underwent both scans within 2 years.
Scans were read by 6 nuclear medicine physicians blinded to CL.
Reader performance was compared against various CL thresholds.
CL was measured with CapAIBL using a) the standard method, b) a composite cerebellum plus hemispheric white matter reference region for FBP and c) with an AI correction method for the CL SUVR (DeepSUVR).
The DeepSUVR correction was learnt from a separate dataset of longitudinal scans where unexpected temporal changes were penalized and a tracer specific SUVR correction factor was developed for application to single scans.
Result CL frequency distribution graphs differed for NAV and FBP.
Applying the DeepSUVR correction improved the correlation (R 2 ) between NAV and FBP CL from 0.
82 to 0.
94, predominantly by boosting many mid‐range FBP CL values and reducing FBP variance around zero CL (Figure 1).
The FBP composite reference region had a similar though less marked effect.
The thresholds that produced peak visual read accuracy with standard and DeepSUVR CL methods were 15CL and 11CL for NAV, and 29CL and 21CL for FBP (Figure 2).
Mean sensitivity at peak accuracy thresholds was higher in NAV with 97% for standard and 96% for DeepSUVR methods compared to 93% for both methods in FBP.
Mean specificity at the same thresholds was similar with 95% for NAV and FBP with standard CL while DeepSUVR produced a specificity of 97% for both tracers.
Conclusion Visual read of amyloid PET can accurately detect levels of amyloid below the traditional thresholds of 20 or 25CL for a positive scan.
Visual read thresholds vary by tracer and reader experience.
AI correction of CL with the DeepSUVR method significantly improves the performance of Florbetapir but NAV4694 remains more sensitive to low amyloid levels.

Related Results

SEMANA DE ENFERMAGEM E SEUS ASPECTOS SOCIAIS NA VALORIZAÇÃO PROFISSIONAL: UM RELATO DE EXPERIÊNCIA DO GRUPO PET-ENFERMAGEM
SEMANA DE ENFERMAGEM E SEUS ASPECTOS SOCIAIS NA VALORIZAÇÃO PROFISSIONAL: UM RELATO DE EXPERIÊNCIA DO GRUPO PET-ENFERMAGEM
A enfermagem é o pilar da assistência pois está na linha de frente do cuidado holístico, todavia esta é estigmatizada e desvalorizada, assim como não possui reconhecimento consider...
283 Quantifying muscle amyloid in inclusion body myositis using pet
283 Quantifying muscle amyloid in inclusion body myositis using pet
ObjectivesInclusion body myositis (IBM) shares some histopathological features with polymyositis (PM). Current investigations have low sensitivity for identification of amyloid dep...
Comparative study of total-body PET and PET/MR in the diagnosis of liver metastases
Comparative study of total-body PET and PET/MR in the diagnosis of liver metastases
ObjectiveTo compare the diagnostic differences between total-body PET/CT (positron emission tomography/computed tomography) and PET/MR (positron emission tomography/magnetic resona...
Baseline Staging Evaluation in Lymphoma: The Role of FDG PET, CT, and Bone Marrow Biopsy
Baseline Staging Evaluation in Lymphoma: The Role of FDG PET, CT, and Bone Marrow Biopsy
Abstract Abstract 2640 BACKGROUND: The revised response criteria for malignant lymphoma (Cheson et al JCO 25:579 ...
Tekstualni subjekt u poeziji Marije Stepanove od 2001. do 2017. godine
Tekstualni subjekt u poeziji Marije Stepanove od 2001. do 2017. godine
Maria Stepanova (b. 1972) is a contemporary Russian poet who has emerged in recent decades as one of the most original and complex voices on the poetically highly heterogeneous and...

Back to Top