Javascript must be enabled to continue!
A Comprehensive Bench Assessment of Automatic Tube Compensation in ICU Ventilators for Better Clinical Management
View through CrossRef
BACKGROUND:
Automatic tube compensation (ATC) unloads endotracheal tube (ETT) resistance. We conducted a bench assessment of ATC functionality in ICU ventilators to improve clinical management.
METHODS:
This study had 2 phases. First, we performed an international survey on the use of ATC in clinical practice, hypothesizing a rate of ATC use of 25%. Second, we tested 7 modern ICU ventilators in a lung model mimicking a normal subject (Normal), a subject with ARDS, and a subject with COPD. Inspiratory effort consisted of esophageal pressure over 30 consecutive breaths obtained in a real patient under weaning. A brand new 8-mm inner diameter ETT was attached to the lung model, and ATC was set at 100% compensation for the ETT. The 30 breaths were first run with ATC off and no ETT (ie, reference period), and then with ATC on and ETT (ie, active period). The primary end point was the difference in tidal volume (V
T
) between reference and active periods. We hypothesized that the V
T
difference should be equal to 0 in an ideally functioning ATC. V
T
difference was compared across ventilators and respiratory mechanics conditions using a linear mixed-effects model.
RESULTS:
The clinical use of ATC was 64% according to 644 individuals who responded to the international survey. The V
T
difference varied significantly across ventilators in all respiratory mechanics configurations. The divergence between V
T
difference and 0 was small but significant: the extreme median (interquartile range) values were −0.013 L (–0.019 to −0.002) in the COPD model and 0.056 L (0.051–0.06) in the Normal model. V
T
difference for all ventilators was 0.015 L (95% CI 0.013–0.018) in the ARDS model, which was significantly different from 0.021 L (95% CI 0.018–0.024) in the Normal model (
P
< .001) and 0.010 L (0.007–0.012) in the COPD model (
P
= .003).
CONCLUSIONS:
ATC is used more frequently in clinical practice than expected. In addition, V
T
delivery by ATC differed slightly though significantly between ventilators.
Title: A Comprehensive Bench Assessment of Automatic Tube Compensation in ICU Ventilators for Better Clinical Management
Description:
BACKGROUND:
Automatic tube compensation (ATC) unloads endotracheal tube (ETT) resistance.
We conducted a bench assessment of ATC functionality in ICU ventilators to improve clinical management.
METHODS:
This study had 2 phases.
First, we performed an international survey on the use of ATC in clinical practice, hypothesizing a rate of ATC use of 25%.
Second, we tested 7 modern ICU ventilators in a lung model mimicking a normal subject (Normal), a subject with ARDS, and a subject with COPD.
Inspiratory effort consisted of esophageal pressure over 30 consecutive breaths obtained in a real patient under weaning.
A brand new 8-mm inner diameter ETT was attached to the lung model, and ATC was set at 100% compensation for the ETT.
The 30 breaths were first run with ATC off and no ETT (ie, reference period), and then with ATC on and ETT (ie, active period).
The primary end point was the difference in tidal volume (V
T
) between reference and active periods.
We hypothesized that the V
T
difference should be equal to 0 in an ideally functioning ATC.
V
T
difference was compared across ventilators and respiratory mechanics conditions using a linear mixed-effects model.
RESULTS:
The clinical use of ATC was 64% according to 644 individuals who responded to the international survey.
The V
T
difference varied significantly across ventilators in all respiratory mechanics configurations.
The divergence between V
T
difference and 0 was small but significant: the extreme median (interquartile range) values were −0.
013 L (–0.
019 to −0.
002) in the COPD model and 0.
056 L (0.
051–0.
06) in the Normal model.
V
T
difference for all ventilators was 0.
015 L (95% CI 0.
013–0.
018) in the ARDS model, which was significantly different from 0.
021 L (95% CI 0.
018–0.
024) in the Normal model (
P
< .
001) and 0.
010 L (0.
007–0.
012) in the COPD model (
P
= .
003).
CONCLUSIONS:
ATC is used more frequently in clinical practice than expected.
In addition, V
T
delivery by ATC differed slightly though significantly between ventilators.
Related Results
Blunt Chest Trauma and Chylothorax: A Systematic Review
Blunt Chest Trauma and Chylothorax: A Systematic Review
Abstract
Introduction: Although traumatic chylothorax is predominantly associated with penetrating injuries, instances following blunt trauma, as a rare and challenging condition, ...
Oxygen management in New Zealand and Australian intensive care units: A knowledge translation study
Oxygen management in New Zealand and Australian intensive care units: A knowledge translation study
<p><b>Background: Knowledge translation literature shows a delay between publication and uptake of research findings into clinical practice. There is uncertainty about ...
Oxygen management in New Zealand and Australian intensive care units: A knowledge translation study
Oxygen management in New Zealand and Australian intensive care units: A knowledge translation study
<p><b>Background: Knowledge translation literature shows a delay between publication and uptake of research findings into clinical practice. There is uncertainty about...
TELE-ICU BERMANFAAT DALAM PENCAPAIAN PELAYANAN BERKUALITAS
TELE-ICU BERMANFAAT DALAM PENCAPAIAN PELAYANAN BERKUALITAS
ABSTRAKKejadian mortalitas di ruang ICU masih tinggi. Pasien kritis membutuhkan perawatan kompleks sehingga membutuhkan perawat terlatih dan kompeten tetapi penyebaran tenaga masih...
Burnout syndrome among Thai intensivists and nurses in pre-COVID19 era
Burnout syndrome among Thai intensivists and nurses in pre-COVID19 era
Background: Burnout syndrome (BOS), a work-related constellation of symptoms and signs, causes individuals emotional stress and is associated with increasing job-related disillusio...
Savable but lost lives when ICU is overloaded: a model from 733 patients in epicenter Wuhan, China
Savable but lost lives when ICU is overloaded: a model from 733 patients in epicenter Wuhan, China
Abstract
Background. Coronavirus Disease (COVID-19) causes a sudden turn over to bad at some check-point and thus needs intervention of intensive care unit (ICU). This resu...
Environmental Surveillance Protocols for Highly Pathogenic Avian Influenza (HPAI) v2
Environmental Surveillance Protocols for Highly Pathogenic Avian Influenza (HPAI) v2
EnvironmentalSurveillance Protocols for Highly Pathogenic Avian Influenza (HPAI) This comprehensive protocol suite enables systematic environmental surveillance for avian influenza...
Intensive care unit-acquired Stenotrophomonas maltophilia: incidence, risk factors, and outcome
Intensive care unit-acquired Stenotrophomonas maltophilia: incidence, risk factors, and outcome
Abstract
Introduction
The aim of this study was to determine incidence, risk factors, and impact on outcome of intensive care unit (...