GATE Simulation Study of the Siemens Biograph mCT 20 Excel PET/CT System

View through CrossRef
Abstract We used GATE simulation to study the effect of the coincidence time window (CTW) along with the block gap and the intercrystal gap on the count rate performance and the spatial resolution of the Biograph ™ mCT 20 Excel. We ran simulations on our local cluster to reduce computation time. The task was split into several jobs that were then triggered simultaneously on the cluster nodes. The Biograph TM mCT 20 Excel was validated using the NEMA NU 2-2012 protocol. Our results showed good agreement with experimental data. The simulated sensitivity, peak true count rate, peak noise equivalent count rate (NECR), and scatter fraction showed agreement within 3.62%, 5.77%, 0.6%, and 2.69%, respectively. In addition, the spatial resolution agreed within <0.51 mm. The results showed that a decrease in the coincidence time window and the block gap and an increase in the intercrystal gap increase the count rate performance and improve the spatial resolution. The results also showed that decreasing the coincidence time window increased the NECR by 27.37%. Changing the intercrystal gap from 0 to 0.2 mm and the block gap and from 4 to 0.4 mm increased the NECR by 5.53% and improved the spatial resolution at 1 cm by 2.91% and that at 10 cm by 3.85%. The coincidence time window, crystal gap, and block gap are important parameters with respect to improving the spatial resolution.
Title: GATE Simulation Study of the Siemens Biograph mCT 20 Excel PET/CT System
Description:
Abstract We used GATE simulation to study the effect of the coincidence time window (CTW) along with the block gap and the intercrystal gap on the count rate performance and the spatial resolution of the Biograph ™ mCT 20 Excel.
We ran simulations on our local cluster to reduce computation time.
The task was split into several jobs that were then triggered simultaneously on the cluster nodes.
The Biograph TM mCT 20 Excel was validated using the NEMA NU 2-2012 protocol.
Our results showed good agreement with experimental data.
The simulated sensitivity, peak true count rate, peak noise equivalent count rate (NECR), and scatter fraction showed agreement within 3.
62%, 5.
77%, 0.
6%, and 2.
69%, respectively.
In addition, the spatial resolution agreed within <0.
51 mm.
The results showed that a decrease in the coincidence time window and the block gap and an increase in the intercrystal gap increase the count rate performance and improve the spatial resolution.
The results also showed that decreasing the coincidence time window increased the NECR by 27.
37%.
Changing the intercrystal gap from 0 to 0.
2 mm and the block gap and from 4 to 0.
4 mm increased the NECR by 5.
53% and improved the spatial resolution at 1 cm by 2.
91% and that at 10 cm by 3.
85%.
The coincidence time window, crystal gap, and block gap are important parameters with respect to improving the spatial resolution.

Related Results

SEMANA DE ENFERMAGEM E SEUS ASPECTOS SOCIAIS NA VALORIZAÇÃO PROFISSIONAL: UM RELATO DE EXPERIÊNCIA DO GRUPO PET-ENFERMAGEM
SEMANA DE ENFERMAGEM E SEUS ASPECTOS SOCIAIS NA VALORIZAÇÃO PROFISSIONAL: UM RELATO DE EXPERIÊNCIA DO GRUPO PET-ENFERMAGEM
A enfermagem é o pilar da assistência pois está na linha de frente do cuidado holístico, todavia esta é estigmatizada e desvalorizada, assim como não possui reconhecimento consider...
Inter-scanner Aβ-amyloid PET harmonization using barrel phantom spatial resolution matching
Inter-scanner Aβ-amyloid PET harmonization using barrel phantom spatial resolution matching
Abstract The standardized uptake value ratio (SUVR) is used to measure Aβ uptake in PET images of the brain. Variations in PET scanner technologies and image recons...
Notched Gate and Graded Gate Oxide Processing for Reduced Capacitance Application in RF MOSFETs
Notched Gate and Graded Gate Oxide Processing for Reduced Capacitance Application in RF MOSFETs
As the demands of RF applications are rising, optimization of internal MOSFETs capacitances is a key issue to improve the cut-off frequency. In this abstract we report the developm...
Comparative study of total-body PET and PET/MR in the diagnosis of liver metastases
Comparative study of total-body PET and PET/MR in the diagnosis of liver metastases
ObjectiveTo compare the diagnostic differences between total-body PET/CT (positron emission tomography/computed tomography) and PET/MR (positron emission tomography/magnetic resona...
Baseline Staging Evaluation in Lymphoma: The Role of FDG PET, CT, and Bone Marrow Biopsy
Baseline Staging Evaluation in Lymphoma: The Role of FDG PET, CT, and Bone Marrow Biopsy
Abstract Abstract 2640 BACKGROUND: The revised response criteria for malignant lymphoma (Cheson et al JCO 25:579 ...
Musical and verbal auditory cognition : child development and neurophysiological markers
Musical and verbal auditory cognition : child development and neurophysiological markers
Cognition auditive verbale et musicale : développement chez l’enfant et marqueurs neurophysiologiques Les processus auditifs centraux jouent un rôle essentiel dans ...

Back to Top