Assessment of different regions of interest-based methods for [99mTc]Tc DAT-SPECT quantification using an anthropomorphic striatal phantom
View through CrossRef
Abstract
Background and aims
Molecular imaging of the dopamine transporters (DAT) provides valuable information about neurodegenerative diseases, such as Parkinson’s. This study assessed the accuracy and precision of DAT-SPECT quantification methods.
Methods
Twenty-three DAT-SPECT images of a striatal phantom were acquired. The specific (caudate and putamen) and the non-specific (background activity) chambers were filled with [
99m
Tc]Tc. Different specific-to-non-specific activity ratios (10, 9, 8, 7, 6, 5, 4, 3 and 2 to 1) and the specific binding ratio (SBR) were calculated. Five methods using ROIs were assessed: (a) Manual ROIs on SPECT images; (b) TwoBox and (c) ThreeBox methods and Volume of Interest (VOI) using structural images; (d) MRI and (e) CT. Accuracy was evaluated by the concordance correlation coefficient (CCC) and precision by Pearson’s coefficient and linear regression.
Results
The SBR quantified in the specific and striatal chambers resulted in a CCC increase with a decrease in the nominal values. For lower SBR, MRI and CT showed higher CCCs when caudate (
$$\hbox {CCC}_{\mathrm{MRI-CA}}$$
CCC
MRI
-
CA
= 0.89 e
$$\hbox {CCC}_{\mathrm{CT-CA}}$$
CCC
CT
-
CA
= 0.84) and putamen (
$$\hbox {CCC}_{\mathrm{MRI-PU}}$$
CCC
MRI
-
PU
= 0.86 e
$$\hbox {CCC}_{\mathrm{CT-PU}}$$
CCC
CT
-
PU
= 0.82) were evaluated. For striatal assessments, the TwoBox method was the most accurate (
$$\hbox {CCC}_{\mathrm{TWOBOX-ST}}$$
CCC
TWOBOX
-
ST
= 0.95). High Pearson’s coefficients were found in the correlations between all methods.
conclusions
All five methods showed high precision even when applied to images with different activities. MRI and CT were the most accurate for assessing the caudate or putamen. To assess the striatal chamber and in the absence of structural information, the TwoBox method is advisable.
Springer Science and Business Media LLC
Title: Assessment of different regions of interest-based methods for [99mTc]Tc DAT-SPECT quantification using an anthropomorphic striatal phantom
Description:
Abstract
Background and aims
Molecular imaging of the dopamine transporters (DAT) provides valuable information about neurodegenerative diseases, such as Parkinson’s.
This study assessed the accuracy and precision of DAT-SPECT quantification methods.
Methods
Twenty-three DAT-SPECT images of a striatal phantom were acquired.
The specific (caudate and putamen) and the non-specific (background activity) chambers were filled with [
99m
Tc]Tc.
Different specific-to-non-specific activity ratios (10, 9, 8, 7, 6, 5, 4, 3 and 2 to 1) and the specific binding ratio (SBR) were calculated.
Five methods using ROIs were assessed: (a) Manual ROIs on SPECT images; (b) TwoBox and (c) ThreeBox methods and Volume of Interest (VOI) using structural images; (d) MRI and (e) CT.
Accuracy was evaluated by the concordance correlation coefficient (CCC) and precision by Pearson’s coefficient and linear regression.
Results
The SBR quantified in the specific and striatal chambers resulted in a CCC increase with a decrease in the nominal values.
For lower SBR, MRI and CT showed higher CCCs when caudate (
$$\hbox {CCC}_{\mathrm{MRI-CA}}$$
CCC
MRI
-
CA
= 0.
89 e
$$\hbox {CCC}_{\mathrm{CT-CA}}$$
CCC
CT
-
CA
= 0.
84) and putamen (
$$\hbox {CCC}_{\mathrm{MRI-PU}}$$
CCC
MRI
-
PU
= 0.
86 e
$$\hbox {CCC}_{\mathrm{CT-PU}}$$
CCC
CT
-
PU
= 0.
82) were evaluated.
For striatal assessments, the TwoBox method was the most accurate (
$$\hbox {CCC}_{\mathrm{TWOBOX-ST}}$$
CCC
TWOBOX
-
ST
= 0.
95).
High Pearson’s coefficients were found in the correlations between all methods.
conclusions
All five methods showed high precision even when applied to images with different activities.
MRI and CT were the most accurate for assessing the caudate or putamen.
To assess the striatal chamber and in the absence of structural information, the TwoBox method is advisable.
Related Results
Absolute quantification in brain SPECT imaging
Absolute quantification in brain SPECT imaging
Certes malalties neurològiques estan associades amb problemes en els sistemes de neurotransmissió. Una aproximació a l'estudi d'aquests sistemes és la tomografia d'emissió SPECT (S...
Non-Invasive Mycobacterium avium Detection Using 99mTc-GSA on Single-Photon Emission Computed Tomography
Non-Invasive Mycobacterium avium Detection Using 99mTc-GSA on Single-Photon Emission Computed Tomography
Background: The prevalence of nontuberculous mycobacteria (NTM) infection is on the rise, surpassing that of pulmonary tuberculosis in Japan. Current standard therapy for NTM infec...
Assessment of different based regions of interest methods for [99mTc]Tc DAT-SPECT quantification using an anthropomorphic striatal phantom.
Assessment of different based regions of interest methods for [99mTc]Tc DAT-SPECT quantification using an anthropomorphic striatal phantom.
Abstract
Background and Aims: Molecular imaging of the dopamine transporters (DAT) provides usefull information about neurodegenerative diseases, such as Parkinson’s. This ...
Comparative Kinetics of Microvascular Inulin and 99mTc-Labelled Diethylenetriaminepenta-Acetic Acid Exchange
Comparative Kinetics of Microvascular Inulin and 99mTc-Labelled Diethylenetriaminepenta-Acetic Acid Exchange
1. After simultaneous intravenous injection as a mixture, 99mTc-labelled diethylenetriaminepentaacetic acid (99mTc-DTPA; molecular mass 492 Da) and inulin (∼6000 Da) gave arterial ...
Accuracy comparison of various quantitative [99mTc]Tc-DPD SPECT/CT reconstruction techniques in patients with symptomatic hip and knee joint prostheses
Accuracy comparison of various quantitative [99mTc]Tc-DPD SPECT/CT reconstruction techniques in patients with symptomatic hip and knee joint prostheses
Abstract
Background
There is a need for better diagnostic tools that identify loose total hip and knee arthroplasties. Here, we present the accuracy...
Assessment of Cardiac Amyloidosis with 99mTc-Pyrophosphate (PYP) Quantitative SPECT
Assessment of Cardiac Amyloidosis with 99mTc-Pyrophosphate (PYP) Quantitative SPECT
Abstract
Background: 99mTc-PYP scintigraphy provides differential diagnosis of ATTR cardiomyopathy (ATTR-CM) from light chain cardiac amyloidosis and other myocardial disor...
Assessment of Cardiac Amyloidosis with 99mTc-Pyrophosphate (PYP) Quantitative SPECT
Assessment of Cardiac Amyloidosis with 99mTc-Pyrophosphate (PYP) Quantitative SPECT
Abstract
Background: 99mTc-PYP scintigraphy provides differential diagnosis of ATTR cardiomyopathy (ATTR-CM) from light chain cardiac amyloidosis and other myocardial disor...
Radiation Dose Reduction Strategy for SPECT/CT Bone Scan
Radiation Dose Reduction Strategy for SPECT/CT Bone Scan
Abstract
Purpose: The aim of this study is to introduce the optimization method of CT parameters to reduce patient radiation exposure in bone SPECT/CT while maintaining ima...